Analysis of biometric parameters of cataract eyes measured with optical biometer Lenstar LS900, IOL Master 700, and OPD-SCAN III

被引:6
作者
Bao, Tong [1 ]
Wang, Lifeng [1 ]
Liu, Caihui [1 ]
Yang, Yajun [2 ]
Pang, Yunlei [3 ]
机构
[1] Chifeng Chaoju Eye Hosp, Dept Ophthalmol Examinat, Chifeng, Peoples R China
[2] Chifeng Chaoju Eye Hosp, Dept Cataract, Chifeng, Peoples R China
[3] Chifeng Chaoju Eye Hosp, Dept Ophthalm Plast Surg, Chifeng, Peoples R China
关键词
Biometry; Cataract; Swept source optical coherence tomography; Cornea indicators; CHAMBER DEPTH MEASUREMENTS; DEVICE;
D O I
10.1016/j.pdpdt.2023.103646
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: To compare the characteristics of biometric measurements among the LS900, IOL Master, and OPD-SCAN III in patients with mild-to-moderate cataract. Methods: We recruited 85 eyes of 78 patients diagnosed with mild-to-moderate cataract at our hospital between January and April 2023. The K (average corneal curvature; K1 and K2), astigmatism (AST), and white to white (WTW) were measured using those three devices, respectively. Differences and correlations between those measures based on the device were explored. Results: The K1, K2, and AST were highly consistent among the groups (P = 0.851, P = 0.626, P = 0.473, respectively). The WTW measured by IOL Master were larger than those measured by the LS900 and OPD-SCAN III (all P < 0.001). However, all three devices were highly correlated in all measures (P < 0.001). Bland-Altman plots indicated a significant agreement between three devices in all measurements (P < 0.001). Conclusions: K1, K2, and AST in patients with mild-to-moderate cataract were similar among all three devices, whereas the WTW were not. WTW measurement differences between three devices should be concerned in cataract cases.
引用
收藏
页数:4
相关论文
共 15 条
  • [1] Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500
    Akman, Ahmet
    Asena, Leyla
    Gungor, Sirel Gur
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (09) : 1201 - 1205
  • [2] Agreement of anterior segment measurements between LenStar LS 900 optical biometer and OPD Scan III wavefront aberrometer devices in eyes with cataract
    Bao, Tong
    Yin, Lihua
    Liu, Caihui
    Pang, Yunlei
    Zhang, Guisen
    Yang, Yajun
    Zhang, Bozhou
    [J]. PHOTODIAGNOSIS AND PHOTODYNAMIC THERAPY, 2023, 41
  • [3] A new optical low coherence reflectometry device for ocular biometry in cataract patients
    Buckhurst, P. J.
    Wolffsohn, J. S.
    Shah, S.
    Naroo, S. A.
    Davies, L. N.
    Berrow, E. J.
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (07) : 949 - 953
  • [4] Haider Muhammad Ali, 2023, J Ayub Med Coll Abbottabad, V35, P76, DOI 10.55519/JAMC-01-11149
  • [5] Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry
    Hoffer, Kenneth J.
    Hoffmann, Peter C.
    Savini, Giacomo
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) : 1165 - 1172
  • [6] Comparison of 2 laser instruments for measuring axial length
    Hoffer, Kenneth J.
    Shammas, H. John
    Savini, Giacomo
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (04) : 644 - 648
  • [7] Agreement of anterior segment measurements between four diagnostic imaging devices in myopic patients
    Placide, John
    Da Silva, Helio V. Neves
    McCabe, Shannon E.
    Ronquillo, Yasmyne C.
    Moshirfar, Majid
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2021, 18 (12) : 1235 - 1243
  • [8] Should Artificial Tears Be Used During the Preoperative Assessment of Toric IOLs Before Age-Related Cataract Surgery? The TORIDE Study
    Rochet, Elsa
    Levron, Antoine
    Agard, Emilie
    El Chehab, Hussam
    Plas, Helene
    Bouvarel, Hugo
    Chirpaz, Nicolas
    Billant, Jeremy
    Dot, Corinne
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2021, 37 (11) : 759 - +
  • [9] Visual and Refractive Outcomes of a Trifocal Toric Intraocular Lens Implanted in Eyes With High Corneal Astigmatism
    Ruiz-Mesa, Ramon
    Tana-Sanz, Pedro
    Tana-Sanz, Santiago
    Orts-Vila, Paz
    Tana-Rivero, Pedro
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2023, 39 (04) : 229 - +
  • [10] A new non-contact optical device for ocular biometry
    Santodomingo-Rubido, J
    Mallen, EAH
    Gilmartin, B
    Wolffsohn, JS
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2002, 86 (04) : 458 - 462