A label-free dual immunosensor for the simultaneous electrochemical determination of CA125 and HE4 biomarkers for the early diagnosis of ovarian cancer

被引:23
作者
Bilgi Kamac, Melike [1 ]
Altun, Muhammed [1 ]
Yilmaz, Merve [1 ]
Sezginturk, Mustafa Kemal [2 ]
机构
[1] Cankiri Karatekin Univ, Fac Sci, Chem Dept, TR-18100 Cankiri, Turkiye
[2] Canakkale Onsekiz Mart Univ, Fac Engn, Bioengn Dept, Canakkale, Turkiye
关键词
CA125; HE4; Simultaneous determination; Dual electrochemical immunosensors; Ovarian cancer; ANTIGEN; 125; BIOSENSOR; DISEASE; PROTEIN; POINT; PREDICTION; MULTIPLE; MARKER; SIGNAL; RISK;
D O I
10.1007/s00216-023-04569-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The blood levels of cancer antigen 125 (CA125) and human epididymal secretory protein 4 (HE4) are measured in the diagnosis and progression monitoring of ovarian cancer (OC), and the Risk of Ovarian Malignancy Algorithm (ROMA) score% values are calculated for cancer risk assessment. For the first time, disposable dual screen-printed carbon electrodes modified with reduced graphene oxide, polythionine, and gold nanoparticles were used to fabricate label-free electrochemical dual CA125-HE4 immunosensors for the sensitive, fast, and practical simultaneous determination of CA125 and HE4. DPV and SWV methods were used to simultaneously determine antigens in two different linear ranges (1-100 pg mL(-1) and 1-50 ng mL(-1)). High sensitivity, low LOD, and LOQ were obtained for two linear ranges with a correlation coefficient above 0.99. The application stability of the dual CA125-HE4 immunosensors was determined as 60 days, and the storage stability was determined as 16 weeks. The dual immunosensors exhibited high selectivity in eight different antigen mixtures. The reusability of the dual immunosensors has been tested up to 9 cycles. ROMA score% values for pre-menopausal and post-menopausal status were calculated using the concentration of CA125 and HE4 in the blood serum and assessing OC risk. The disposable dual immunosensors can be used in point-of-care tests for rapid and practical simultaneous determination of CA125 and HE4 with high selectivity, sensitivity, and repeatability.
引用
收藏
页码:1709 / 1718
页数:10
相关论文
共 49 条
[11]   Comparison of Serum Human Epididymis Protein 4 with Cancer Antigen 125 as a Tumor Marker in Patients with Malignant and Nonmalignant Diseases [J].
Escudero, Jose M. ;
Auge, Jose M. ;
Filella, Xavier ;
Torne, Aureli ;
Pahisa, Jaume ;
Molina, Rafael .
CLINICAL CHEMISTRY, 2011, 57 (11) :1534-1544
[12]   Point of Care Diagnostics: Status and Future [J].
Gubala, Vladimir ;
Harris, Leanne F. ;
Ricco, Antonio J. ;
Tan, Ming X. ;
Williams, David E. .
ANALYTICAL CHEMISTRY, 2012, 84 (02) :487-515
[13]   Evaluation of biomarker panels for early stage ovarian cancer detection and monitoring for disease recurrence [J].
Havrilesky, Laura J. ;
Whitehead, Clark M. ;
Rubatt, Jennifer M. ;
Cheek, Robert L. ;
Groelke, John ;
He, Qin ;
Malinowski, Douglas P. ;
Fischer, Timothy J. ;
Berchuck, Andrew .
GYNECOLOGIC ONCOLOGY, 2008, 110 (03) :374-382
[14]  
Hellström I, 2003, CANCER RES, V63, P3695
[15]   Label-free electrochemical immunosensor for detection of oncomarker CA125 in serum [J].
Honorato de Castro, Ana Cristina ;
Alves, Livia Maria ;
Silva Siquieroli, Ana Carolina ;
Madurro, Joao Marcos ;
Brito-Madurro, Ana Graci .
MICROCHEMICAL JOURNAL, 2020, 155
[16]   Serum HE4 concentration differentiates malignant ovarian tumours from ovarian endometriotic cysts [J].
Huhtinen, K. ;
Suvitie, P. ;
Hiissa, J. ;
Junnila, J. ;
Huvila, J. ;
Kujari, H. ;
Setala, M. ;
Harkki, P. ;
Jalkanen, J. ;
Fraser, J. ;
Makinen, J. ;
Auranen, A. ;
Poutanen, M. ;
Perheentupa, A. .
BRITISH JOURNAL OF CANCER, 2009, 100 (08) :1315-1319
[17]   Determination of carcinoembryonic antigen as a tumor marker using a novel graphene-based label-free electrochemical immunosensor [J].
Jozghorbani, Maryam ;
Fathi, Mojtaba ;
Kazemi, Sayed Habib ;
Alinejadian, Navid .
ANALYTICAL BIOCHEMISTRY, 2021, 613
[18]   High preoperative blood levels of HE4 predicts poor prognosis in patients with ovarian cancer [J].
Kalapotharakos, Grigorios ;
Asciutto, Christine ;
Henic, Emir ;
Casslen, Bertil ;
Borgfeldt, Christer .
JOURNAL OF OVARIAN RESEARCH, 2012, 5
[19]   A new disposable amperometric NADH sensor based on screen-printed electrode modified with reduced graphene oxide/polyneutral red/gold nanoparticle [J].
Kamac, Melike Bilgi ;
Onat, Eda Kiymaz ;
Yilmaz, Merve .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2020, 100 (04) :419-431
[20]   Cerebrospinal fluid levels of alpha-synuclein measured using a poly-glutamic acid-modified gold nanoparticle-doped disposable neuro-biosensor system [J].
Karaboga, Munteha Nur Sonuc ;
Sezginturk, Mustafa Kemal .
ANALYST, 2019, 144 (02) :611-621