Determination and prospective validation of cut-off values for the diagnosis of liver cirrhosis for point shear-wave elastography/acoustic radiation force impulse imaging using the ACUSON Sequoia ultrasound system

被引:0
|
作者
Meitner-Schellhaas, B. [1 ]
Schueler, S. [1 ]
Vogl, T. [1 ]
Jesper, D. [1 ]
Vetter, M. [1 ]
Waldner, M. [1 ]
Strobel, D. [1 ]
机构
[1] FAU Univ Erlangen Nurnberg, Erlangen Univ Hosp, Dept Internal Med 1, Ulmenweg 18, D-91054 Erlangen, Germany
关键词
acoustic radiation force impulse; cut-off values; elastography; fibrosis; liver cirrhosis; liver; point shear-wave elastography; ultrasound; CLINICAL-USE; GUIDELINES; RECOMMENDATIONS; FIBROSIS;
D O I
10.1097/MEG.0000000000002683
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Objectives Point shear-wave elastography (pSWE) alias acoustic radiation force impulse (ARFI) imaging is a well-established ultrasound-based technique for the non-invasive assessment of liver tissue stiffness. As cut-off values for liver cirrhosis cannot be transferred from one ultrasound system to another, this study aimed at determining cut-off values for the newly developed Siemens ACUSON Sequoia ultrasound system.Methods In a pilot study phase, two independent examiners conducted 10 pSWE measurements in an elasticity phantom and 32 healthy individuals for the determination of inter-examiner agreement. Afterwards, 22 cirrhotic patients and 57 patients with chronic liver disease undergoing liver biopsy underwent pSWE. Patient characteristics and stiffness values were compared for individuals with and without liver cirrhosis. Diagnostic accuracies of cut-off values for the diagnosis of liver cirrhosis were calculated using areas under the receiver operating characteristics analysis and Youden's index. In a subsequent validation study phase, these cut-off values were validated prospectively in 107 cirrhotic and 68 non-cirrhotic patients.Results Inter-examiner agreement was excellent for measurements in the elasticity phantom (intra-class correlation coefficient [ICC] = 0.998; P < 0.001), and good for measurements in patients (ICC = 0.844; P < 0.001). The best cut-off value for the diagnosis of liver cirrhosis was 1.405 m/s with an AUC of 0.872, a sensitivity of 88.2% and a specificity of 88.2% (P < 0.001).Conclusion ARFI elastography using the Siemens ACUSON Sequoia showed a good inter-examiner agreement. The optimal cut-off value was lower than the cut-off values described for former generations of ultrasound devices. These preliminary results should be confirmed in larger patient collectives with histology as the reference standard.
引用
收藏
页码:135 / 140
页数:6
相关论文
共 7 条
  • [1] Point Shear-Wave Elastography Using Acoustic Radiation Force Impulse Imaging for the Prediction of Liver-Related Events in Patients With Chronic Viral Hepatitis
    Hernandez Sampere, Lisa
    Vermehren, Johannes
    Muecke, Victoria T.
    Graf, Christiana
    Peiffer, Kai-Henrik
    Dultz, Georg
    Zeuzem, Stefan
    Waidmann, Oliver
    Filmann, Natalie
    Bojunga, Joerg
    Sarrazin, Christoph
    Friedrich-Rust, Mireen
    Muecke, Marcus M.
    HEPATOLOGY COMMUNICATIONS, 2021, 5 (01) : 112 - 121
  • [2] The diagnostic performances of conventional strain elastography (SE), acoustic radiation force impulse (ARFI) imaging and point shear-wave speed (pSWS) measurement for non-calcified thyroid nodules
    Chen, Bao-Ding
    Xu, Hui-Xiong
    Zhang, Yi-Feng
    Liu, Bo-Ji
    Guo, Le-Hang
    Li, Dan-Dan
    Zhao, Chong-Ke
    Li, Xiao-Long
    Wang, Dan
    Zhao, Shuang-Shuang
    CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2017, 65 (03) : 259 - 273
  • [3] Comparison of strain elastography, point shear wave elastography using acoustic radiation force impulse imaging and 2D-shear wave elastography for the differentiation of thyroid nodules
    Kyriakidou, Georgia
    Friedrich-Rust, Mireen
    Bon, Dimitra
    Sircar, Ishani
    Schrecker, Christopher
    Bogdanou, Dimitra
    Herrmann, Eva
    Bojunga, Joerg
    PLOS ONE, 2018, 13 (09):
  • [4] Point shear wave elastography (pSWE) using Acoustic Radiation Force Impulse (ARFI) imaging: a feasibility study and norm values for renal parenchymal stiffness in healthy children and adolescents
    Grass, Lilian
    Szekely, Nora
    Alrajab, Abdulsattar
    Thi Thanh Tam Bui-Ta
    Hoffmann, Georg F.
    Wuehl, Elke
    Schenk, Jens Peter
    MEDICAL ULTRASONOGRAPHY, 2017, 19 (04) : 366 - 373
  • [5] NORMAL VALUES OF LIVER SHEAR WAVE VELOCITY IN HEALTHY CHILDREN ASSESSED BY ACOUSTIC RADIATION FORCE IMPULSE IMAGING USING A CONVEX PROBE AND A LINEAR PROBE
    Fontanilla, Teresa
    Canas, Teresa
    Macia, Araceli
    Alfageme, Marta
    Gutierrez Junquera, Carolina
    Malalana, Ana
    Luz Cilleruelo, Maria
    Roman, Enriqueta
    Miralles, Maria
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (03) : 470 - 477
  • [6] Prospective Evaluation of Acoustic Radiation Force Impulse (ARFI) Elastography and High-Frequency B-Mode Ultrasound in Compensated Patients for the Diagnosis of Liver Fibrosis/Cirrhosis in Comparison to Mini-Laparoscopic Biopsy
    Pfeifer, L.
    Zopf, S.
    Siebler, J.
    Schwitulla, J.
    Wildner, D.
    Wachter, D.
    Neurath, M. F.
    Strobel, D.
    ULTRASCHALL IN DER MEDIZIN, 2015, 36 (06): : 581 - 589
  • [7] Evaluation of Point Shear Wave Elastography Using Acoustic Radiation Force Impulse Imaging for Longitudinal Fibrosis Assessment in Patients with HBeAg-Negative HBV Infection
    Graf, Christiana
    Mondorf, Antonia
    Knop, Viola
    Peiffer, Kai-Henrik
    Dietz, Julia
    Friess, Julia
    Wedemeyer, Heiner
    Buggisch, Peter
    Mauss, Stefan
    Berg, Thomas
    Rausch, Michael
    Sprinzl, Martin
    Klinker, Hartwig
    Hinrichsen, Holger
    Bronowicki, Jean-Pierre
    Haag, Sebastian
    Hueppe, Dietrich
    Lutz, Thomas
    Poynard, Thierry
    Zeuzem, Stefan
    Friedrich-Rust, Mireen
    Sarrazin, Christoph
    Vermehren, Johannes
    JOURNAL OF CLINICAL MEDICINE, 2019, 8 (12)