HPLC DETERMINATION OF BIPERIDEN IN SOLID DOSAGE FORMS

被引:0
作者
Angelov, Trifon Milchev [1 ]
Krastanova, Biljana Slaveva [2 ]
Pashev, Alexandar Svilenov [1 ]
Jotova, Maya Chavdarova [1 ]
机构
[1] Med Univ Pleven, Fac Pharm, 1 St Kliment Ohridski St, Pleven 5800, Bulgaria
[2] Tchaikapharma High Qual Med Inc, 1 GM Dimitrov Blvd, Sofia 1172, Bulgaria
来源
COMPTES RENDUS DE L ACADEMIE BULGARE DES SCIENCES | 2023年 / 76卷 / 02期
关键词
biperiden; HPLC-UV; assay; method development;
D O I
10.7546/CRABS.2023.02.02
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
A rapid, new analytical method, based on liquid chromatography with UV detection, has been developed and utilised for the determination of biperi-den hydrochloride in solid dosage forms. The chromatographic separation was performed on a conventional C18 reversed phase column with a mobile phase composed of acetonitrile-buffer mixture. Ultraviolet detection was carried out at 205 nm. A rapid extraction procedure using ultrasonic bath obtained good extraction yield values for the analyte (>= 98.4%). The method was developed specifically for routine analysis of biperiden in solid dosage forms in the appli-cable concentration range. The linearity and range of the method investigated for the assay of biperiden is between 2-6 mu g mL-1. The method is selective and sufficiently precise. Thus, the method developed is suitable for routine analysis of biperiden.
引用
收藏
页码:184 / 191
页数:8
相关论文
共 9 条
[1]  
Capacio BR, 2002, BIOMED CHROMATOGR, V16, P47, DOI 10.1002/bmc.113.abs
[2]  
Drugbank online, US
[3]  
Hadidi Kamal A, 2004, Leg Med (Tokyo), V6, P233, DOI 10.1016/j.legalmed.2004.06.001
[4]  
ISSA F. A. A., 2012, J PHARMACEUT BIOMED, V11, P49
[5]   Anti-Parkinson Drug Biperiden Inhibits Enzyme Acetylcholinesterase [J].
Kostelnik, Adam ;
Cegan, Alexander ;
Pohanka, Miroslav .
BIOMED RESEARCH INTERNATIONAL, 2017, 2017
[6]  
Lotfy H. M., 2014, Anal. Chem. Indian J, V14, P106
[7]   Enantio separation and quality control of biperiden in pharmaceutical formulations by capillary electrophoresis [J].
Mandrioli, R ;
Musenga, A ;
Lasaponara, SS ;
Saracino, MA ;
Fanali, S ;
Raggi, MA .
ANALYTICA CHIMICA ACTA, 2006, 560 (1-2) :57-63
[8]   Development and validation of a stability-indicating high performance liquid chromatographic (HPLC) assay for biperiden in bulk form and pharmaceutical dosage forms [J].
Mohammadi, A. ;
Mehramizi, A. ;
Aghaee Moghaddam, F. ;
Erfani jabarian, L. ;
Pourfarzib, M. ;
Kashani, H. N. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 854 (1-2) :152-157
[9]  
United States Pharmacopeial Convention, 2011, USP35NF30, P2363