Wear estimation of hip implants with varying chamfer geometry at the trunnion junction: a finite element analysis

被引:27
作者
Gutmann, Celine [1 ]
Shaikh, Numa [2 ]
Shenoy, B. Satish [3 ]
Bhat, N. Shaymasunder [4 ]
Keni, Laxmikant G. [3 ]
Chethan, K. N. [3 ]
机构
[1] Univ Appl Sci Northwestern Switzerland, Inst Med Engn & Med Informat, Sch Life Sci HLS, FHNW, Windisch, Switzerland
[2] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Mech & Ind Engn, Manipal, Karnataka, India
[3] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Aeronaut & Automobile Engn, Manipal, Karnataka, India
[4] Manipal Acad Higher Educ, Kasturba Med Coll, Dept Orthopaed, Manipal, Karnataka, India
关键词
hip implant; finite element analysis; ansys; linear wear; pressure; TAPER JUNCTION; FRETTING WEAR; ARTHROPLASTY; STEM; REPLACEMENTS; FAILURE; RATES;
D O I
10.1088/2057-1976/acb710
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The hip joint helps the upper body to transfer its weight to lower body. Along with age, there are various reasons for the degeneration of the hip joint. The artificial hip implant replaces the degenerated hip. Wear between the joints is the primary cause of the hip implant becoming loose. The wear can occur due to various reasons. Due to this revision surgery are most common in young and active patients. In the design phase of the implant if this is taken care then life expectancy of the implant can be improved. Small design changes can significantly enhance the implant's life. In this work, elliptical-shaped hip implant stem is designed, and linear wear is estimated at trunnion junction. In this work, a 28 mm diameter femoral head with a 4 mm thick acetabular cup and a 2 mm thick backing cup is used. The top surface taper radiuses are changed. Solid works was used to create the models. Ansys was used to perform the analysis. It was found that as the radius of the TTR decreased, the wear rate decreased. The least wear rate was found in 12/14 mm taper with a value of 1.15E(-02) mm year(-1) for the first material combination and with a value of 1.23E (-02) mm year(-1) for the second material combination. In the comparison between the models with 1 mm chamfer and no chamfer, it was found that the wear rate was lower for the models with 1 mm chamfer. When the chamfer was increased (more than 1 mm), the linear wear increased. Wear is the main reason for the loosening of hip implants, which leads to a revision of an implant. It was found that with a decrease in TTR, there was a small increase in the linear wear rate. Overall, the implant with TTR 6 mm and a chamfer of 1 mm was found to have the least wear rate. To validate these results, the implant can be 3D printed and tested on a hip simulator.
引用
收藏
页数:11
相关论文
共 32 条
[1]  
[Anonymous], 2020, F2996 standard test method for finite element analysis (FEA) of nonmodular metallic orthopaedic hip femoral stems
[2]   Does a micro-grooved trunnion stem surface finish improve fixation and reduce fretting wear at the taper junction of total hip replacements? A finite element evaluation [J].
Ashkanfar, Ariyan ;
Langton, David J. ;
Joyce, Thomas J. .
JOURNAL OF BIOMECHANICS, 2017, 63 :47-54
[3]   A large taper mismatch is one of the key factors behind high wear rates and failure at the taper junction of total hip replacements: A finite element wear analysis [J].
Ashkanfar, Ariyan ;
Langton, David J. ;
Joyce, Thomas J. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 69 :257-266
[4]   Finite element simulation of early creep and wear in total hip arthroplasty [J].
Bevill, SL ;
Bevill, GR ;
Penmetsa, JR ;
Petrella, AJ ;
Rullkoetter, PJ .
JOURNAL OF BIOMECHANICS, 2005, 38 (12) :2365-2374
[5]   Static structural analysis of the effect of change in femoral head sizes used in Total Hip Arthroplasty using finite element method [J].
Bhawe, Aniket Kunal ;
Shah, Kavish Maulik ;
Somani, Samridhi ;
Shenoy, Satish B. ;
Bhat, Shyamasunder N. ;
Zuber, Mohammad ;
Chethan, K. N. .
COGENT ENGINEERING, 2022, 9 (01)
[6]   Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system [J].
Brock, Timothy M. ;
Sidaginamale, Raghavendra ;
Rushton, Steven ;
Nargol, Antoni V. F. ;
Bowsher, John G. ;
Savisaar, Christina ;
Joyce, Tom J. ;
Deehan, David J. ;
Lord, James K. ;
Langton, David J. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (12) :1868-1874
[7]   Finite element analysis of hip implant with varying in taper neck lengths under static loading conditions [J].
Chethan, K. N. ;
Bhat, N. Shyamasunder ;
Zuber, Mohammad ;
Shenoy, B. Satish .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2021, 208
[8]   Wear estimation of trapezoidal and circular shaped hip implants along with varying taper trunnion radiuses using finite element method [J].
Chethan, K. N. ;
Ogulcan, Guldeniz ;
Bhat, Shyamasunder N. ;
Zuber, Mohammad ;
Shenoy, Satish B. .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 196
[9]   Optimized trapezoidal-shaped hip implant for total hip arthroplasty using finite element analysis [J].
Chethan, K. N. ;
Zuber, Mohammad ;
Bhat, Shyamasunder N. ;
Shenoy, Satish B. .
COGENT ENGINEERING, 2020, 7 (01)
[10]   Static structural analysis of different stem designs used in total hip arthroplasty using finite element method [J].
Chethan, K. N. ;
Zuber, Mohammad ;
Bhat, Shyamasunder N. ;
Shenoy, Satish B. ;
Kini, Chandrakant R. .
HELIYON, 2019, 5 (06)