Catalytic Surgical Smoke Filtration Unit Reduces Formaldehyde Levels in a Simulated Operating Room Environment

被引:3
作者
Carroll, Gregory T. [1 ]
Kirschman, David L. [1 ]
机构
[1] Aerobiotix Inc, Sci Affairs, Miamisburg, OH 45342 USA
关键词
formaldehyde; catalysis; operating room; medical device; air quality; electrosurgery; VOC; ADSORPTION; OXIDATION;
D O I
10.1021/acs.chas.2c00071
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Electrosurgery and procedures that utilize formalin expose healthcare workers to formaldehyde. Incorporating catalytic materials into air-purifying devices has the potential to reduce formaldehyde levels in nosocomial settings. A catalytic transition metal oxide was interfaced with a portable negative pressure (PNP) unit in a simulated operating room (OR) polluted with formaldehyde by exposure to formalin and by applying electrosurgical cuts to porcine muscle. The formaldehyde levels decayed biexponentially when the PNP unit was equipped with a catalyst. When the PNP unit was off, the time-weighted average concentration of formaldehyde detected in the perimeter of the OR during simulated electrosurgery was 0.45 (0.21) ppm. When the PNP unit was on, the time-weighted average concentration reduced to 0.15 (0.08) ppm. The device has the potential to considerably reduce formaldehyde levels in OR environments.
引用
收藏
页码:21 / 28
页数:8
相关论文
共 32 条
  • [1] [Anonymous], 1997, Health Devices, V26, P132
  • [2] [Anonymous], 2022, OSHA FACT SHEET FORM
  • [3] [Anonymous], 2022, MED MAN GUID FORM
  • [4] [Anonymous], 1702017 ANSIASHRAEAS
  • [5] Towards a formalin-free hospital. Levels of 15-F2t-isoprostane and malondialdehyde to monitor exposure to formaldehyde in nurses from operating theatres
    Bellisario, Valeria
    Mengozzi, Giulio
    Grignani, Elena
    Bugiani, Massimiliano
    Sapino, Anna
    Bussolati, Gianni
    Bono, Roberto
    [J]. TOXICOLOGY RESEARCH, 2016, 5 (04) : 1122 - 1129
  • [6] Formaldehyde toxicity reports fromin vitroandin vivostudies: a review and updated data
    Bernardini, Leticia
    Barbosa, Eduardo
    Charao, Mariele Feiffer
    Brucker, Natalia
    [J]. DRUG AND CHEMICAL TOXICOLOGY, 2022, 45 (03) : 972 - 984
  • [7] Carroll G. T., 2022, CHEMRXIV, DOI [10.26434/chemrxiv-2022-plfcp, DOI 10.26434/CHEMRXIV-2022-PLFCP]
  • [8] Discrete room pressure drops predict door openings and contamination levels in the operating room setting
    Carroll, Gregory T.
    Kirschman, David L.
    [J]. PERIOPERATIVE CARE AND OPERATING ROOM MANAGEMENT, 2022, 29
  • [9] A Peripherally Located Air Recirculation Device Containing an Activated Carbon Filter Reduces VOC Levels in a Simulated Operating Room
    Carroll, Gregory T.
    Kirschman, David L.
    [J]. ACS OMEGA, 2022, 7 (50): : 46640 - 46645
  • [10] Increased CO2 levels in the operating room correlate with the number of healthcare workers present: an imperative for intentional crowd control
    Carroll, Gregory T.
    Kirschman, David L.
    Mammana, Angela
    [J]. PATIENT SAFETY IN SURGERY, 2022, 16 (01)