A bench study of balloon-expandable valves for the treatment of self-expanding valve failure

被引:9
作者
Akodad, Mariama [1 ,2 ,3 ,4 ]
Meier, David [1 ,2 ,3 ]
Sellers, Stephanie [1 ,2 ,3 ]
De Backer, Ole [5 ]
Mylotte, Darren [6 ,7 ]
Landes, Uri [8 ,9 ]
Frawley, Chris [10 ]
Lynch, Lisa [10 ]
Tang, Gilbert H. L. [11 ]
Sondergaard, Lars [5 ]
Wood, David A. [1 ,2 ,3 ]
Webb, John G. [1 ,2 ,3 ]
Sathananthan, Janarthanan [1 ,2 ,3 ]
机构
[1] Univ British Columbia, St Pauls Hosp, Ctr Heart Valve Innovat, Vancouver, BC, Canada
[2] Univ British Columbia, Ctr Cardiovasc Innovat, Vancouver, BC, Canada
[3] Ctr Heart Lung Innovat & Providence Res, Cardiovasc Translat Lab, Vancouver, BC, Canada
[4] Inst Cardiovasc Paris Sud, Hop Prive Jacques Cartier, Ramsay Sante, Massy, France
[5] Copenhagen Univ Hosp, Dept Cardiol, Heart Ctr, Rigshosp, Copenhagen, Denmark
[6] Natl Univ Ireland, Dept Cardiol, Galway, Ireland
[7] NUIG, Galway, Ireland
[8] Edith Wolfson Med Ctr, Holon, Israel
[9] Tel Aviv Univ, Tel Aviv, Israel
[10] Boston Sci Corp, Marlborough, MA USA
[11] Mt Sinai Hosp, Dept Cardiovasc Surg, New York, NY USA
关键词
aortic stenosis; other; TAVI; TRANSCATHETER; REPLACEMENT; OUTCOMES;
D O I
10.4244/EIJ-D-22-00769
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Coronary obstruction and access are concerns in patients undergoing redo transcatheter aortic valve implantation (TAVI). Aims: We sought to assess the neoskirt height, leaflet overhang, leaflet deflection,and transcatheter heart valve (THV) expansion and performance, at 2 different implant depths, of the SAPIEN 3 Ultra (S3U) within the ACURATE neo2 (ACn2) THV. Methods: An in vitro study was performed with a 23 mm S3U deployed within a small (S) ACn2 and a 26 mm S3U deployed within a medium (M) and a large (L) ACn2. The S3U outflow was positioned at the top of the ACn2 crown (low implant) and at the base of the commissural post of the ACn2 (high implant). Testing was performed under physiological conditions as per ISO-5840-3 standard. Results: The neoskirt height was shorter when the S3U outflow was positioned at a low implantation depth (S: 9.6 mm, M: 12.2 mm, L: 13.8 mm vs S: 15.2 mm, M: 15.1 mm, L: 17.8 mm ACn2 for high implants). Hydrodynamic performance was acceptable for all configurations. Leaflet overhang was <50% for all configurations except the low implant of the 26 mm S3U in the L ACn2 (77.6%). There was a gap from the side of the neoskirt to the outer border of the THV frame which was >2 mm for all configurations. The S3U was underexpanded for all configurations, and the achieved area was 77.9%-92.9% of the expected nominal area. Conclusions: Redo TAVI with an S3U within an ACn2 has favourable hydrodynamics and moderate leaflet overhang. Importantly, the design of the ACn2 results in a neoskirt that is not deflected all the way to the outer dimensions of the THV, hence creating a space that facilitates coronary flow and access.
引用
收藏
页码:93 / +
页数:12
相关论文
共 18 条
  • [1] Late Balloon Valvuloplasty for Transcatheter Heart Valve Dysfunction
    Akodad, Mariama
    Blanke, Philipp
    Chuang, Ming-Yu A.
    Duchscherer, Jade
    Sellers, Stephanie L.
    Chatfield, Andrew G.
    Gulsin, Gaurav G.
    Lauck, Sandra
    Leipsic, Jonathon A.
    Meier, David
    Moss, Rob R.
    Cheung, Anson
    Sathananthan, Janarthanan
    Wood, David A.
    Ye, Jian
    Webb, John G.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2022, 79 (14) : 1340 - 1351
  • [2] Balloon-Expandable Valve for Treatment of Evolut Valve Failure Implications on Neoskirt Height and Leaflet Overhang
    Akodad, Mariama
    Sellers, Stephanie
    Landes, Uri
    Meier, David
    Tang, Gilbert H. L.
    Gada, Hemal
    Rogers, Toby
    Caskey, Michael
    Rutkin, Bruce
    Puri, Rishi
    Rovin, Joshua
    Leipsic, Jonathon
    Sondergaard, Lars
    Grubb, Kendra J.
    Gleason, Patrick
    Garde, Kshitija
    Tadros, Hatem
    Teodoru, Sebastian
    Wood, David A.
    Webb, John G.
    Sathananthan, Janarthanan
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2022, 15 (04) : 368 - 377
  • [3] Leaflet and Neoskirt Height in Transcatheter Heart Valves Implications for Repeat Procedures and Coronary Access
    Akodad, Mariama
    Sellers, Stephanie
    Gulsin, Gaurav S.
    Tzimas, Georgios
    Landes, Uri
    Chatfield, Andrew G.
    Chuang, Anthony
    Meier, David
    Leipsic, Jonathon
    Blanke, Philippe
    Ye, Jian
    Cheung, Anson
    Wood, David A.
    Khan, Jaffar M.
    Webb, John G.
    Sathananthan, Janarthanan
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (20) : 2298 - 2300
  • [4] Next-generation balloon-expandable transcatheter heart valve: the SAPIEN 3 Ultra valve
    Chatfield, Andrew
    Sathananthan, Janarthanan
    Wood, David A.
    Webb, John G.
    [J]. FUTURE CARDIOLOGY, 2021, 17 (05) : 811 - 816
  • [5] Coronary Access After TAVR-in-TAVR as Evaluated by Multidetector Computed Tomography
    De Backer, Ole
    Landes, Uri
    Fuchs, Andreas
    Yoon, Sung-Han
    Mathiassen, Ole Norling
    Sedaghat, Alexander
    Kim, Won-Keun
    Pilgrim, Thomas
    Buzzatti, Nicola
    Ruile, Philipp
    El Sabbagh, Abdallah
    Barbanti, Marco
    Fiorina, Claudia
    Nombela-Franco, Luis
    Steinvil, Arie
    Finkelstein, Ariel
    Montorfano, Matteo
    Maurovich-Horvat, Pal
    Kofoed, Klaus Fuglsang
    Blanke, Philipp
    Bunc, Matjaz
    Neumann, Franz-Josef
    Latib, Azeem
    Windecker, Stephan
    Sinning, Jan-Malte
    Norgaard, Bjarne Linde
    Makkar, Raj
    Webb, John G.
    Sondergaard, Lars
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (21) : 2528 - 2538
  • [6] Surgical explantation of transcatheter aortic bioprostheses: Results and clinical implications
    Fukuhara, Shinichi
    Brescia, Alexander A.
    Shiomi, Suzuna
    Rosati, Carlo M.
    Yang, Bo
    Kim, Karen M.
    Deeb, G. Michael
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2021, 162 (02) : 539 - +
  • [7] Incidence, Characteristics, Predictors, and Outcomes of Surgical Explantation After Transcatheter Aortic Valve Replacement
    Hirji, Sameer A.
    Percy, Edward D.
    McGurk, Siobhan
    Malarczyk, Alexandra
    Harloff, Morgan T.
    Yazdchi, Farhang
    Sabe, Ashraf A.
    Bapat, Vinayak N.
    Tang, Gilbert H. L.
    Bhatt, Deepak L.
    Thourani, Vinod H.
    Leon, Martin B.
    O'Gara, Patrick
    Shah, Pinak B.
    Kaneko, Tsuyoshi
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 76 (16) : 1848 - 1859
  • [8] International Organization for Standardization, 2021, ISO 5840-3:2021
  • [9] Incomplete expansion of transcatheter aortic valves is associated with propensity for valve thrombosis
    Khodaee, Farhan
    Barakat, Mohammed
    Abbasi, Mostafa
    Dvir, Danny
    Azadani, Ali N.
    [J]. INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2020, 30 (01) : 39 - 46
  • [10] Transcatheter Replacement of Transcatheter Versus Surgically Implanted Aortic Valve Bioprostheses
    Landes, Uri
    Sathananthan, Janarthanan
    Witberg, Guy
    De Backer, Ole
    Sondergaard, Lars
    Abdel-Wahab, Mohamed
    Holzhey, David
    Kim, Won-Keun
    Hamm, Christian
    Buzzatti, Nicola
    Montorfano, Matteo
    Ludwig, Sebastian
    Conradi, Lenard
    Seiffert, Moritz
    Guerrero, Mayra
    El Sabbagh, Abdallah
    Rodes-Cabau, Josep
    Guimaraes, Leonardo
    Codner, Pablo
    Okuno, Taishi
    Pilgrim, Thomas
    Fiorina, Claudia
    Colombo, Antonio
    Mangieri, Antonio
    Eltchaninoff, Helene
    Nombela-Franco, Luis
    Van Wiechen, Maarten P. H.
    Van Mieghem, Nicolas M.
    Tchetche, Didier
    Schoels, Wolfgang H.
    Kullmer, Matthias
    Tamburino, Corrado
    Sinning, Jan-Malte
    Al-Kassou, Baravan
    Perlman, Gidon Y.
    Danenberg, Haim
    Ielasi, Alfonso
    Fraccaro, Chiara
    Tarantini, Giuseppe
    De Marco, Federico
    Redwood, Simon R.
    Lisko, John C.
    Babaliaros, Vasilis C.
    Laine, Mika
    Nerla, Roberto
    Castriota, Fausto
    Finkelstein, Ariel
    Loewenstein, Itamar
    Eitan, Amnon
    Jaffe, Ronen
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (01) : 1 - 14