Simulations of X-ray spectra, half value layer, and mean energy from mammography using EGSnrc Monte Carlo and SpekPy

被引:1
作者
Huy, B. Ngoc [1 ]
Van Dung, P. [1 ]
机构
[1] Dalat Nucl Res Inst DNRI, 01 Nguyen Tu Luc, Da Lat, Vietnam
关键词
Monte Carlo; EGSnrc simulation; SpekPy simulation; Mammography; Half value layer;
D O I
10.1016/j.radi.2022.09.009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: This study presents the simulation results of X-ray spectra, half value layers (HVLs), and mean energies (Emean) of two mammography units using EGSnrc Monte Carlo (MC) and SpekPy computer codes.Methods: The spectra caused by different combinations of targets/filters at various tube voltages (kVps) of two mammography units were simulated using two different computer codes. The EGSnrc MC simulated data of spectra and Emean were compared with those obtained from SpekPy. The simulated values of two units' HVLs obtained from two computer codes were compared with those from physical measurements from mammography machines used in clinical practice.Results: The maximum discrepancies in Emean simulated from two codes were less than 4.1% and 1.5% for the target/filter combination of W/Rh and Mo/Mo, respectively. The HVLs of the SpekPy were well matched to the physical measurements. The percentage differences were within 3.3% and 6.8% for two units, respectively. The EGSnrc MC simulated values of HVLs show the percentage differences within 8.9% and 7.0% with those from physical measurements for two units, respectively. All methods of HVLs determination comply with the requirements of IAEA Human Health Series No.17.Conclusions: The HVLs, Emean, spectra varied depending on the target/filter combinations and composition of the mammography tubes. The simulation results verify that the HVLs evaluation using the EGSnrc MC and SpekPy can be validated for mammography standard beam qualities and provide prediction almost immediately compared with physical experiments. Implications for practice: EGSnrc MC and SpekPy have been considered powerful toolkits to simulate typical X-ray tubes used in mammography due to the good agreement between the calculation of Emean, physical measurements and simulated HVLs.(c) 2022 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:28 / 37
页数:10
相关论文
共 23 条
[1]  
[Anonymous], 2022, PRIMARY STANDARDS PT
[2]  
[Anonymous], 2022, WOLFRAM MATH MODERN
[3]   Monte Carlo simulation of x-ray spectra in diagnostic radiology and mammography using MCNP4C [J].
Ay, MR ;
Shahriari, M ;
Sarkar, S ;
Adib, M ;
Zaidi, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (21) :4897-4917
[4]   Off-axis x-ray spectra: A comparison of Monte Carlo simulated and computed x-ray spectra with measured spectra [J].
Bhat, M ;
Pattison, J ;
Bibbo, G ;
Caon, M .
MEDICAL PHYSICS, 1999, 26 (02) :303-309
[5]   COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR [J].
BIRCH, R ;
MARSHALL, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :505-517
[6]  
Briesmeister JF, 2000, LA13709M LOS AL NAT
[7]   A validation of SpekPy: A software toolkit for modelling X-ray tube spectra [J].
Bujila, Robert ;
Omar, Artur ;
Poludniowski, Gavin .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 75 :44-54
[8]  
Cranley K., 1997, The Institute of Physics and Engineering in Medicine Report
[9]   Experimental and Monte Carlo-simulated spectra of standard mammography-quality beams [J].
David, M. G. ;
Pires, E. J. ;
Bernal, M. A. ;
Peixoto, J. G. ;
Dealmeida, C. E. .
BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1013) :629-635
[10]  
Engen R, 2006, EUROPEAN PROTOCOL QU, P57