Real-time assessment of high-intensity focused ultrasound heating and cavitation with hybrid optoacoustic ultrasound imaging
被引:14
作者:
Ozsoy, Cagla
[1
,2
,3
]
Lafci, Berkan
[1
,2
,3
]
Reiss, Michael
[1
,2
,3
]
Dean-Bena, Xose Luis
[1
,2
,3
,4
,5
,6
]
Razansky, Daniel
[1
,2
,3
,4
,5
,6
]
机构:
[1] Univ Zurich, Inst Biomed Engn, Fac Med, Zurich, Switzerland
[2] Univ Zurich, Inst Pharmacol & Toxicol, Fac Med, Zurich, Switzerland
[3] Swiss Fed Inst Technol, Inst Biomed Engn, Dept Informat Technol & Elect Engn, Zurich, Switzerland
[4] Univ Zurich, Inst Biomed Engn, Zurich, Switzerland
[5] Univ Zurich, Inst Pharmacol & Toxicol, Zurich, Switzerland
[6] Swiss Fed Inst Technol, Zurich, Switzerland
来源:
PHOTOACOUSTICS
|
2023年
/
31卷
基金:
欧洲研究理事会;
关键词:
Optoacoustic imaging;
Photoacoustic imaging;
Ultrasound imaging;
High-intensity focused ultrasound treatment;
monitoring;
TEMPERATURE-DEPENDENCE;
PHYSICAL-MECHANISMS;
GRUNEISEN-PARAMETER;
ACOUSTIC CAVITATION;
ABLATION;
FREQUENCY;
THERAPY;
TOMOGRAPHY;
CATHETER;
TISSUE;
D O I:
10.1016/j.pacs.2023.100508
中图分类号:
O42 [声学];
学科分类号:
070206 ;
082403 ;
摘要:
High-intensity focused ultrasound (HIFU) enables localized ablation of biological tissues by capitalizing on the synergistic effects of heating and cavitation. Monitoring of those effects is essential for improving the efficacy and safety of HIFU interventions. Herein, we suggest a hybrid optoacoustic-ultrasound (OPUS) approach for real-time assessment of heating and cavitation processes while providing an essential anatomical reference for ac-curate localization of the HIFU-induced lesion. Both effects could clearly be observed by exploiting the tem-perature dependence of optoacoustic (OA) signals and the strong contrast of gas bubbles in pulse-echo ultrasound (US) images. The differences in temperature increase and its rate, as recorded with a thermal camera for different HIFU pressures, evinced the onset of cavitation at the expected pressure threshold. The estimated temperatures based on OA signal variations were also within 10-20 % agreement with the camera readings for temperatures below the coagulation threshold (similar to 50 degrees C). Experiments performed in excised tissues as well as in a post-mortem mouse demonstrate that both heating and cavitation effects can be effectively visualized and tracked using the OPUS approach. The good sensitivity of the suggested method for HIFU monitoring purposes was manifested by a significant increase in contrast-to-noise ratio within the ablated region by > 10 dB and > 5 dB for the OA and US images, respectively. The hybrid OPUS-based monitoring approach offers the ease of handheld operation thus can readily be implemented in a bedside setting to benefit several types of HIFU treatments used in the clinics.