Is μCT irradiation nondestructive? A noble gas study on matrix samples from the CV3 chondrite Allende

被引:7
作者
Ghaznavi, Parastoo [1 ]
Kadlag, Yogita [1 ]
Haberthuer, David [2 ]
Hlushchuk, Ruslan [2 ]
Leya, Ingo [1 ,3 ]
机构
[1] Univ Bern, Space Sci & Planetol, Bern, Switzerland
[2] Univ Bern, Inst Anat, Bern, Switzerland
[3] Univ Bern, Space Res & Planetol, Sidlerstr 5, CH-3012 Bern, Switzerland
基金
瑞士国家科学基金会;
关键词
RAY COMPUTED-TOMOGRAPHY; SPECTRA; TUBE;
D O I
10.1111/maps.13996
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
Micro-computed tomography (mu CT) is a fast and powerful technology for studying textural, physical, and chemical properties of solid objects in three dimensions. While regularly used for sample documentation and curation, it is often assumed that mu CT techniques are essentially nondestructive or at least very little destructive. However, there are very few studies proving or rejecting the assumption of nondestructiveness. Here we study whether X-ray tomographic imaging affects the noble gas budget of matrix samples from the CV3 carbonaceous chondrite Allende. We irradiated powdered and homogenized matrix samples in the Bruker SkyScan 1272 mu CT instrument at three different X-ray tube acceleration voltages of 30, 70, and 100 keV. By comparing the noble gas concentrations and especially the elemental and isotopic ratios of the irradiated samples with data for two nonirradiated aliquots, we found no significant differences. Our study therefore demonstrates that X-ray tomographic imaging has no measurable effect on the noble gas budget and can therefore safely be used for sample characterization prior to noble gas studies.
引用
收藏
页码:897 / 900
页数:4
相关论文
共 14 条
  • [1] Sectioning effects of porphyritic chondrules: Implications for the PP/POP/PO classification and correcting modal abundances of mineralogically zoned chondrules
    Barosch, Jens
    Hezel, Dominik C.
    Sawatzki, Lena
    Halbauer, Lucia
    Marrocchi, Yves
    [J]. METEORITICS & PLANETARY SCIENCE, 2020, 55 (05) : 993 - 999
  • [2] Experimental investigation of the nebular formation of chondrule rims and the formation of chondrite parent bodies
    Beitz, E.
    Blum, J.
    Mathieu, R.
    Pack, A.
    Hezel, D. C.
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 2013, 116 : 41 - 51
  • [3] Three-dimensional textural and compositional analysis of particle tracks and fragmentation history in aerogel
    Ebel, Denton S.
    Greenberg, Michael
    Rivers, Mark L.
    Newville, Matthew
    [J]. METEORITICS & PLANETARY SCIENCE, 2009, 44 (10) : 1445 - 1463
  • [4] Effect of a synchrotron X-ray microtomography imaging experiment on the amino acid content of a CM chondrite
    Friedrich, Jon M.
    Glavin, Daniel P.
    Rivers, Mark L.
    Dworkin, Jason P.
    [J]. METEORITICS & PLANETARY SCIENCE, 2016, 51 (02) : 429 - 437
  • [5] Improved segmentation of meteorite micro-CT images using local histograms
    Griffin, L. D.
    Elangovan, P.
    Mundell, A.
    Hezel, D. C.
    [J]. COMPUTERS & GEOSCIENCES, 2012, 39 : 129 - 134
  • [6] X-ray computed tomography of planetary materials: A primer and review of recent studies
    Hanna, Romy D.
    Ketcham, Richard A.
    [J]. CHEMIE DER ERDE-GEOCHEMISTRY, 2017, 77 (04) : 547 - 572
  • [7] Visualisation and quantification of CV chondrite petrography using micro-tomography
    Hezel, Dominik C.
    Elangovan, Premkumar
    Viehmann, Sebastian
    Howard, Lauren
    Abel, Richard L.
    Armstrong, Robin
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 2013, 116 : 33 - 40
  • [8] SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes
    Poludniowski, G.
    Landry, G.
    DeBlois, F.
    Evans, P. M.
    Verhaegen, F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (19) : N433 - N438
  • [9] Calculation of x-ray spectra emerging from an x-ray tube. Part I. Electron penetration characteristics in x-ray targets
    Poludniowski, Gavin G.
    Evans, Philip M.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2164 - 2174
  • [10] Calculation of x-ray spectra emerging from an x-ray tube. Part II. X-ray production and filtration in x-ray targets
    Poludniowski, Gavin G.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2175 - 2186