Dosimetric evaluation of LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery with more than 20 targets: comparing MME, HyperArc, and RapidArc

被引:3
|
作者
Jung, Hyunuk [1 ]
Yoon, Jihyung [1 ]
Lemus, Olga Dona [1 ]
Tanny, Sean [1 ]
Zhou, Yuwei [1 ]
Milano, Michael [1 ]
Usuki, Kenneth [1 ]
Hardy, Sara [1 ]
Zheng, Dandan [1 ]
机构
[1] Univ Rochester, Dept Radiat Oncol, Rochester, NY 14627 USA
关键词
Brain metastases; Stereotactic radiosurgery; LINAC-based SRS; HyperArc; RapidArc; MME; BRAIN METASTASES; RADIATION-THERAPY; CONFORMITY;
D O I
10.1186/s13014-024-02416-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
BackgroundTo compare the dosimetric quality of three widely used techniques for LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery (fSRS) with more than 20 targets: dynamic conformal arc (DCA) in BrainLAB Multiple Metastases Elements (MME) module and volumetric modulated arc therapy (VMAT) using RapidArc (RA) and HyperArc (HA) in Varian Eclipse.MethodsTen patients who received single-isocenter fSRS with 20-37 targets were retrospectively replanned using MME, RA, and HA. Various dosimetric parameters, such as conformity index (CI), Paddick CI, gradient index (GI), normal brain dose exposures, maximum organ-at-risk (OAR) doses, and beam-on times were extracted and compared among the three techniques. Wilcoxon signed-rank test was used for statistical analysis.ResultsAll plans achieved the prescribed dose coverage goal of at least 95% of the planning target volume (PTV). HA plans showed superior conformity compared to RA and MME plans. MME plans showed superior GI compared to RA and HA plans. RA plans resulted in significantly higher low and intermediate dose exposure to normal brain compared to HA and MME plans, especially for lower doses of >= 8Gy and >= 5Gy. No significant differences were observed in the maximum dose to OARs among the three techniques. The beam-on time of MME plans was about two times longer than RA and HA plans.ConclusionsHA plans achieved the best conformity, while MME plans achieved the best dose fall-off for LINAC-based single-isocenter multi-target multi-fraction SRS with more than 20 targets. The choice of the optimal technique should consider the trade-offs between dosimetric quality, beam-on time, and planning effort.
引用
收藏
页数:9
相关论文
共 4 条
  • [1] Dosimetric evaluation of LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery with more than 20 targets: comparing MME, HyperArc, and RapidArc
    Hyunuk Jung
    Jihyung Yoon
    Olga Dona Lemus
    Sean Tanny
    Yuwei Zhou
    Michael Milano
    Kenneth Usuki
    Sara Hardy
    Dandan Zheng
    Radiation Oncology, 19
  • [2] Innovative margin design and optimized isocenter to minimize the normal tissue in target volumes for single-isocenter multi-target stereotactic radiosurgery
    Yock, Adam D.
    Grees, Beshoi
    Luo, Guozhen
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (19)
  • [3] Dosimetric Performance and Planning/Delivery Efficiency of a Dual-Layer Stacked and Staggered MLC on Treating Multiple Small Targets: A Planning Study Based on Single-Isocenter Multi-Target Stereotactic Radiosurgery (SRS) to Brain Metastases
    Li, Taoran
    Irmen, Peyton
    Liu, Haisong
    Shi, Wenyin
    Alonso-Basanta, Michelle
    Zou, Wei
    Teo, Boon-Keng Kevin
    Metz, James M.
    Dong, Lei
    FRONTIERS IN ONCOLOGY, 2019, 9
  • [4] Dosimetric commissioning of a high-resolution CMOS 2D detector array for patient-specific QA of single-isocenter multi-target VMAT stereotactic radiosurgery
    Teng, Ching-Ling
    Lin, Shih-Chi
    Lovelock, Dale Michael
    Lim, Seng Boh