Quantitative Analysis of Ethanol Based on Laser Raman Spectroscopy Normalization Method

被引:0
|
作者
Liu Hao-dong [1 ,2 ]
Jiang Xi-quan [1 ,2 ]
Niu Hao [1 ,2 ]
Liu Yu-bo [1 ]
Li Hui [2 ]
Liu Yuan [2 ]
Wei Zhang
Li Lu-yan [1 ]
Chen Ting [1 ]
Zhao Yan-jie [1 ]
Ni Jia-sheng [2 ]
机构
[1] Shandong Jianzhu Univ, Coll Sci, Jinan 250101, Peoples R China
[2] Qilu Univ Technol, Laser Inst, Shandong Acad Sci, Jinan 250104, Peoples R China
关键词
Raman spectroscopy; Quantitative analysis; Peak ratio method; Internal standard method; MIXTURES;
D O I
10.3964/j.issn.1000-0593(2023)12-3820-06
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Raman spectroscopy has the advantages of high resolution, fast analysis, simple sample preparation, nondestructive, online measurement, etc. It is widely used to analyze the composition and molecular structure information of a wide range of organic and inorganic substances for qualitative and quantitative analytical measurement. At present, it is mainly used for quantitative analysis. When used for quantitative analysis, the poor reproducibility and the imperfect analysis theory of Raman spectroscopy are two main factors that limit its application. In the quantitative analysis based on intensity ratios, the emergence of Raman intensity normalization theory has provided a theoretical basis for its application. The choice of the reference peak/ internal standard peak and the fitting method have a great influence on the measurement accuracy and stability. In this paper, the relative intensities of the characteristic peaks (C - C-0 symmetrical stretching, 874 cm) of the Raman spectra of ethanol and other reference/internal standard peaks with different ethanol concentrations were investigated using a laser Raman system. A peak ratio method based on ethanol intrinsic peak and an internal standard method based on the characteristic peak position of CCl4 were developed, and both methods can effectively eliminate the effects of mutation noise and strong fluorescence background in the system through normalization. The reference peak/internal standard peak with the best accuracy and stability of the two methods was determined by statistical methods such as joint hypothesis testing for data differences within and between different groups. The F-test and t-test showed that the standard curve established with the characteristic peak at 1 446 cm(-1) (CH3-asymmetric deformation) as the reference peak could more accurately invert the concentration of ethanol when the calibration was performed by the self-peak ratio method, while the standard curve with the Raman characteristic peak at 446 cm as the internal standard peak had higher stability and accuracy when CCl4 was used as the internal standard. The retest within 30 days eliminates the need to measure and plot the standard curve again. The linear regression model established according to the two calibration methods can provide an experimental basis for the quantitative analysis of ethanol solution concentration. The ethanol concentration can be inverted more accurately in real-time, through the application of the model in the ethanol solution concentration detection system, to achieve accurate, rapid and real-time quantitative analysis of ethanol solutions in the high concentration range with strong fluorescence background interference.
引用
收藏
页码:3820 / 3825
页数:6
相关论文
共 14 条
  • [1] [白雪 Bai Xue], 2015, [核化学与放射化学, Journal of Nuclear and Radiochemistry], V37, P143
  • [2] Quantitative Analysis of Anatase-Rutile Mixtures by Raman Spectroscopy
    Cao, Guoqiang
    Yi, Nan
    [J]. CHEMISTRYSELECT, 2020, 5 (37): : 11530 - 11533
  • [3] A fast and novel internal calibration method for quantitative Raman measurements on aqueous solutions
    Duraipandian, S.
    Knopp, M. M.
    Pollard, M. R.
    Kerdoncuff, H.
    Petersen, J. C.
    Mullertz, A.
    [J]. ANALYTICAL METHODS, 2018, 10 (29) : 3589 - 3593
  • [4] Automated material identification with a Raman spectrometer based on the contribution enhancement of small differences and the adaptive target Raman peak subtraction
    Huang, Fei
    Xue, Meng
    Yang, Zhaoqing
    Guo, Hanming
    [J]. APPLIED OPTICS, 2021, 60 (19) : 5682 - 5690
  • [5] Raman spectra of tridymite modifications: MC, MX-1, and PO-10
    Kanzaki, Masami
    [J]. JOURNAL OF MINERALOGICAL AND PETROLOGICAL SCIENCES, 2019, 114 (04) : 214 - 218
  • [6] Liu WH, 2007, CHINESE J ANAL CHEM, V35, P1503
  • [7] Comparison of conventional and shifted excitation Raman difference spectroscopy for bacterial identification
    Lorenz, Bjoern
    Guo, Shuxia
    Raab, Christoph
    Leisching, Patrick
    Bocklitz, Thomas
    Roesch, Petra
    Popp, Juergen
    [J]. JOURNAL OF RAMAN SPECTROSCOPY, 2022, 53 (07) : 1285 - 1292
  • [8] Quantitative Raman determination of hydrogen peroxide using the solvent as internal standard: Online application in the direct synthesis of hydrogen peroxide
    Moreno, T.
    Moran Lopez, M. A.
    Huerta Illera, I.
    Piqueras, C. M.
    Sanz Arranz, A.
    Garcia Serna, J.
    Cocero, M. J.
    [J]. CHEMICAL ENGINEERING JOURNAL, 2011, 166 (03) : 1061 - 1065
  • [9] Quantitative analysis of alcohol-water binary solutions using Raman spectroscopy
    Numata, Yasushi
    Iida, Yoshiyuki
    Tanaka, Hiroyuki
    [J]. JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2011, 112 (06) : 1043 - 1049
  • [10] Quantitative Analysis of Diabetic Blood Raman Spectroscopy Based on XGBoost
    Wang Ming-xuan
    Wang Qiao-yun
    Pian Fei-fei
    Shan Peng
    Li Zhi-gang
    Ma Zhen-he
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42 (06) : 1721 - 1727