Simultaneous Evaluation of Tibiofemoral and Patellofemoral Mechanics in Total Knee Arthroplasty: A Combined Experimental and Computational Approach

被引:5
作者
Behnam, Yashar A. [1 ]
Anantha Krishnan, Ahilan [1 ]
Wilson, Hayden [1 ]
Clary, Chadd W. [1 ]
机构
[1] Univ Denver, Dept Mech & Mat Engn, 2155 East Wesley Ave, Denver, CO 80208 USA
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2024年 / 146卷 / 01期
关键词
PATELLA ALTA; REPLACEMENT; CONTACT; SENSITIVITY; SIMULATION; KINEMATICS; COMPONENT; OUTCOMES; SYSTEM; MUSCLE;
D O I
10.1115/1.4063950
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Contemporary total knee arthroplasty (TKA) has not fully restored natural patellofemoral (P-F) mechanics across the patient population. Previous experimental simulations have been limited in their ability to create dynamic, unconstrained, muscle-driven P-F articulation while simultaneously controlling tibiofemoral (T-F) contact mechanics. The purpose of this study was to develop a novel experimental simulation and validate a corresponding finite element model to evaluate T-F and P-F mechanics. A commercially available wear simulator was retrofitted with custom fixturing to evaluate whole-knee TKA mechanics with varying patella heights during a simulated deep knee bend. A corresponding dynamic finite element model was developed to validate kinematic and kinetic predictions against experimental measurements. Patella alta reduced P-F reaction forces in early and midflexion, corresponding with an increase in T-F forces that indicated an increase in extensor mechanism efficiency. Due to reduced wrapping of the extensor mechanism in deeper flexion for the alta condition, peak P-F forces in flexion increased from 101% to 135% of the applied quadriceps load for the baja and alta conditions, respectively. Strong agreement was observed between the experiment and model predictions with root-mean-square errors (RMSE) for P-F kinematics ranging from 0.8 deg to 3.3 deg and 0.7 mm to 1.4 mm. RMSE for P-F forces ranged from 7.4 N to 53.6 N. By simultaneously controlling dynamic, physiological loading of the T-F and P-F joint, this novel experimental simulation and validated model will be a valuable tool for investigation of future TKA designs and surgical techniques.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Contact Analyses of the Patellofemoral Components in Total Knee Arthroplasty
    Liu, Chien-Wei
    Wang, Ching-Sung
    Ho, Chao-Chung
    Lo, Chia-Chi
    Yu, Chen-Tung
    INFORMATION TECHNOLOGY FOR MANUFACTURING SYSTEMS II, PTS 1-3, 2011, 58-60 : 584 - +
  • [22] Patellofemoral Imbalance in a Balanced Total Knee Arthroplasty: How Does it Occur?
    Evangelista, Perry Jaymes
    Lenz, Nathan
    Laster, Scott
    Sheth, Neil P.
    Schwarzkopf, Ran
    JOURNAL OF KNEE SURGERY, 2019, 32 (02) : 192 - 195
  • [23] Outcomes of Total Knee Replacement after Patellofemoral Arthroplasty
    Hutt, Jonathan
    Dodd, Matthew
    Bourke, Henry
    Bell, Jonathan
    JOURNAL OF KNEE SURGERY, 2013, 26 (04) : 219 - 223
  • [24] Cost effectiveness of patellofemoral versus total knee arthroplasty in younger patients
    Chawla, H.
    Nwachukwu, B. U.
    Van der List, J. P.
    Eggman, A. A.
    Pearle, A. D.
    Ghomrawi, H. M.
    BONE & JOINT JOURNAL, 2017, 99B (08) : 1028 - 1036
  • [25] A new patellofemoral scoring system for total knee arthroplasty
    Baldini, Andrea
    Anderson, John A.
    Zampetti, Piergiuseppe
    Pavlov, Helene
    Sculco, Thomas P.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (452) : 150 - 154
  • [26] The biomechanical effect of tibiofemoral conformity design for patient-specific cruciate retainging total knee arthroplasty using computational simulation
    Koh, Yong-Gon
    Park, Kyoung-Mi
    Kang, Kyoung-Tak
    JOURNAL OF EXPERIMENTAL ORTHOPAEDICS, 2019, 6 (01)
  • [27] Conversion of Patellofemoral Arthroplasty to Robotic-Assisted Total Knee Arthroplasty
    Lachance, Andrew D.
    Steika, Roman
    Lutton, Jeffrey
    Austin, David
    ARTHROPLASTY TODAY, 2023, 23
  • [28] How does asymmetric tibial insert affect tibiofemoral kinematics and contact stresses in total knee Arthroplasty?
    Song, Young Dong
    Nakamura, Shinichiro
    Kuriyama, Shinichi
    Nishitani, Kohei
    Ito, Hiromu
    Tanaka, Yoshihisa
    Morita, Yugo
    Matsuda, Shuichi
    KNEE, 2022, 39 : 185 - 196
  • [29] Evaluation of the Painful Total Knee Arthroplasty
    Cercek, Robert
    Bassett, Richard
    Myerthall, Steve
    JOURNAL OF KNEE SURGERY, 2015, 28 (02) : 113 - 117
  • [30] Evaluation of the Painful Total Knee Arthroplasty
    Flierl, Michael A.
    Sobh, Ali H.
    Culp, Brian M.
    Baker, Erin A.
    Sporer, Scott M.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2019, 27 (20) : 743 - 751