Industry perspectives on the global use of validated blood pressure measuring devices

被引:7
作者
Li, Jim [1 ]
Frick, Gerhard [2 ]
Herberigs, Karel [1 ]
Matsumura, Paul [3 ]
Sarkis, Josh [4 ]
Verberk, Willem J. [2 ,5 ]
Wojcik, Caryl [6 ]
机构
[1] Omron Healthcare Co Ltd, Kyoto, Japan
[2] Microlife AG, Widnau, Switzerland
[3] SunTech Med, Morrisville, NC USA
[4] PharmaSmart, Rochester, NY USA
[5] Maastricht Univ, CARIM Sch Cardiovasc Dis, Dept Biomed Engn, Maastricht, Netherlands
[6] A&D Co Ltd, Tokyo, Japan
关键词
MEDICAL INSTRUMENTATION/EUROPEAN SOCIETY; HYPERTENSION/INTERNATIONAL ORGANIZATION; UNIVERSAL STANDARD; EUROPEAN-SOCIETY; HYPERTENSION; ASSOCIATION; ADVANCEMENT; MONITORS; CALL;
D O I
10.1038/s41371-022-00717-6
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
A group of experts from reputable blood pressure measuring device (BPMD) manufacturers was invited to provide industry perspectives on the global use of validated BPMD. The authors support the recommendations of (1) using the consolidated universal ISO 81060-2:2018 in all future validation studies to ensure consistent and trustworthy quality standards; (2) validation studies to be led by investigators independent from the manufacturer; (3) validation study results to be published in peer-reviewed journals with an independent investigator as the corresponding author; and (4) validated BPMDs to be listed on validated device registries such as STRIDE BP, Validated Device Listing (VDL), and others that are backed by acknowledged scientific associations. The authors call for public awareness of the existence of legally marketed consumer BPMDs that lack sufficient evidence of clinical accuracy. Other important issues and future considerations were discussed, including the need: for awareness building and promoting the use of validated BPMDs among practitioners; to identify a non-mercury sphygmomanometer based reference device to validate BPMDs; to include all cuffs available for use with each BPMD in the validation study; for the promotion of validation studies for special patient populations; for validated wrist BPMDs as an alternative for some patients; for technical innovations to help reduce limitations related to the human aspect of validation studies; for validation of cuffless BPMDs; for validation through equivalency to validated base models; and to use validated BPMD in remote patient monitoring programs. A future collaborative to find solutions to support the use of validated BPMD is envisaged.
引用
收藏
页码:130 / 133
页数:4
相关论文
共 50 条
[11]   Blood pressure measurements with the OptiBP smartphone app validated against reference auscultatory measurements [J].
Schoettker, Patrick ;
Degott, Jean ;
Hofmann, Gregory ;
Proenca, Martin ;
Bonnier, Guillaume ;
Lemkaddem, Alia ;
Lemay, Mathieu ;
Schorer, Raoul ;
Christen, Urvan ;
Knebel, Jean-Francois ;
Wuerzner, Arlene ;
Burnier, Michel ;
Wuerzner, Gregoire .
SCIENTIFIC REPORTS, 2020, 10 (01)
[12]   European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices: European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability [J].
Stergiou, George S. ;
Avolio, Alberto P. ;
Palatini, Paolo ;
Kyriakoulis, Konstantinos G. ;
Schutte, Aletta E. ;
Mieke, Stephan ;
Kollias, Anastasios ;
Parati, Gianfranco ;
Asmar, Roland ;
Pantazis, Nikos ;
Stamoulopoulos, Achilleas ;
Asayama, Kei ;
Castiglioni, Paolo ;
De la Sierra, Alejandro ;
Hahn, Jin-Oh ;
Kario, Kazuomi ;
McManus, Richard J. ;
Myers, Martin ;
Ohkubo, Takayoshi ;
Shroff, Sanjeev G. ;
Tan, Isabella ;
Wang, Jiguang ;
Zhang, Yuanting ;
Kreutz, Reinhold ;
O'Brien, Eoin ;
Mukkamala, Ramakrishna .
JOURNAL OF HYPERTENSION, 2023, 41 (12) :2074-2087
[13]   Improving the accuracy of blood pressure measurement: the influence of the European Society of Hypertension International Protocol (ESH-IP) for the validation of blood pressure measuring devices and future perspectives [J].
Stergiou, George S. ;
Asmar, Roland ;
Myers, Martin ;
Palatini, Paolo ;
Parati, Gianfranco ;
Shennan, Andrew ;
Wang, Jiguang ;
O'Brien, Eoin .
JOURNAL OF HYPERTENSION, 2018, 36 (03) :479-487
[14]   Requirements for design and function of blood pressure measuring devices used for the management of hypertension: Consensus Statement by the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability and STRIDE BP [J].
Stergiou, George S. ;
Parati, Gianfranco ;
Kollias, Anastasios ;
Schutte, Aletta E. ;
Asayama, Kei ;
Asmar, Roland ;
Bilo, Grzegorz ;
de la Sierra, Alejandro ;
Dolan, Eamon ;
Filipovsky, Jan ;
Head, Geoffrey ;
Kario, Kazuomi ;
Kyriakoulis, Konstantinos G. ;
Mancia, Giuseppe ;
Manios, Efstathios ;
Menti, Ariadni ;
McManus, Richard J. ;
Mihailidou, Anastasia S. ;
Muntner, Paul ;
Niiranen, Teemu ;
Ohkubo, Takayoshi ;
Omboni, Stefano ;
Protogerou, Athanasios ;
Saladini, Francesca ;
Sharman, James ;
Shennan, Andrew ;
Shimbo, Daichi ;
Topouchian, Jirar ;
Wang, Jiguang ;
O'Brien, Eoin ;
Palatini, Paolo .
JOURNAL OF HYPERTENSION, 2023, 41 (12) :2088-2094
[15]   Evaluation scale to assess the accuracy of cuff-less blood pressure measuring devices [J].
Yan, Iris Renfei ;
Poon, Carmen Chung Yan ;
Zhang, Yuan-Ting .
BLOOD PRESSURE MONITORING, 2009, 14 (06) :257-267
[16]   Use of oscillometric devices in atrial fibrillation: a comparison of three devices and invasive blood pressure measurement [J].
Halfon, Matthieu ;
Wuerzner, Gregoire ;
Marques-Vidal, Pedro ;
Taffe, Patrick ;
Vaucher, Julien ;
Waeber, Bernard ;
Liaudet, Lucas ;
Ltaief, Zied ;
Popov, Milen ;
Waeber, Gerard .
BLOOD PRESSURE, 2018, 27 (01) :48-55
[17]   Validation protocols for blood pressure measuring devices: the impact of the European Society of Hypertension International Protocol and the development of a Universal Standard [J].
O'Brien, Eoin ;
Stergiou, George ;
Palatini, Paolo ;
Asmar, Roland ;
Ioannidis, John P. ;
Kollias, Anastasios ;
Lacy, Peter ;
McManus, Richard J. ;
Myers, Martin G. ;
Shennan, Andrew ;
Wang, Jiguang ;
Parati, Gianfranco .
BLOOD PRESSURE MONITORING, 2019, 24 (04) :163-166
[18]   A model for early failure prediction of blood pressure measurement devices in a stepped validation approach [J].
Vischer, Annina S. ;
Dutilh, Gilles ;
Socrates, Thenral ;
Burkard, Thilo .
JOURNAL OF CLINICAL HYPERTENSION, 2022, 24 (05) :582-590
[19]   Accuracy of community based blood pressure devices versus validated self muse devices [J].
Karr, Samantha ;
Gurney, Mary K. ;
Early, Nicole K. .
JOURNAL OF THE AMERICAN PHARMACISTS ASSOCIATION, 2015, 55 (04) :419-423
[20]   Accuracy of the KOROT P3 Accurate automated auscultatory blood pressure measuring device for professional use in people with extra-large arms [J].
Fania, Claudio ;
Giletto, Antonella ;
Niero, Alice ;
Palatini, Paolo .
BLOOD PRESSURE MONITORING, 2024, 29 (03) :144-148