Wobbling nature of gamma passing rate as a function of calibration field sizes in patient-specific quality assurance

被引:0
|
作者
Raj, Sathiya [1 ]
Sundaram, Venugopal [2 ]
Godson, Henry Finlay [3 ]
John, Retna [3 ]
机构
[1] Kidwai Mem Inst Oncol, Dept Radiat Phys, Dr MH Marigowda Rd, Bengaluru, India
[2] Meherbai Tata Mem Hosp, R53J 3HJ Stocking Rd, Jamshedpur, Jharkhand, India
[3] Christian Med Coll Vellore, Dept Radiat Oncol, Ranipet Campus, Kilminnal, TN, India
关键词
Gamma analysis; patient-specific QA; calibration; IMRT;
D O I
10.1017/S1460396923000444
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose:This study aimed to investigate the influence of calibration field size on the gamma passing rate (GPR) in patient-specific quality assurance (PSQA).Methods:Two independent detectors, PTW OCTAVIUS 4D (4DOCT) and Arc Check, were utilised in volumetric modulated arc therapy plans for 26 patients (14 with Arc Check and 12 with 4DOCT). Plans were administered using Varian Unique machine (with 4DOCT) and Varian TrueBeam (with Arc Check), each employing different calibration factors (CFs): 4 x 4, 6 x 6, 8 x 8, 10 x 10, 12 x 12 and 15 x 15 cm2 field sizes. Gamma analysis was conducted with 2%2mm, 2%3mm and 3%3mm gamma criteria.Results:GPR exhibited variations across different CFs. GPR demonstrated an increasing trend below 10 x 10 cm(2) CFs, while it displayed a decreasing trend above 10 x 10 cm(2). Both detectors exhibited similar GPR patterns. The correlation between 4DOCT and Arc Check was strong in tighter criteria (2%2mm) with an R-2 value of 0 center dot 9957, moderate criteria (2%3mm) with an R-2 value of 0 center dot 9868, but reduced in liberal criteria (3%3mm) with an R-2 value of 0 center dot 4226.Conclusion:This study demonstrates that calibration field sizes significantly influence GPR in PSQA. This study recommends the plan specific calibration field must obtain to calibrate the QA devices for modulated plans.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Guaranteed performance of individual control chart used in gamma passing rate-based patient-specific quality assurance
    Li, Guangjun
    Xiao, Qing
    Dai, Guyu
    Wang, Qiang
    Bai, Long
    Zhang, Xiangbin
    Zhang, Xiangyu
    Duan, Lian
    Zhong, Renming
    Bai, Sen
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 109
  • [2] Deep Learning for Patient-Specific Quality Assurance : Predicting Gamma Passing Rates for IMRT Based Log File
    Huang, Y.
    Huang, Y.
    Xu, Z.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [3] A robust approach to establish tolerance limits for the gamma passing rate-based patient-specific quality assurance using the heuristic control charts
    Xiao, Qing
    Bai, Long
    Li, Guangjun
    Zhang, Xiangbin
    Li, Zhibin
    Duan, Lian
    Peng, Ruilin
    Zhong, Renming
    Wang, Qiang
    Wang, Xuetao
    Bai, Sen
    MEDICAL PHYSICS, 2022, 49 (02) : 1312 - 1330
  • [4] Statistical analysis of correlation of gamma passing results for two quality assurance phantoms used for patient-specific quality assurance in volumetric modulated arc radiotherapy
    Kunii, Yuki
    Tanabe, Yoshinori
    Nakamoto, Akira
    Nishioka, Kunio
    MEDICAL DOSIMETRY, 2022, 47 (04) : 329 - 333
  • [5] Patient-Specific Independent GammaPlan Quality Assurance for Gamma Knife Perfexion
    Mamalui-Hunter, M.
    Zhao, T.
    Drzymala, R.
    MEDICAL PHYSICS, 2010, 37 (06) : 3305 - +
  • [6] Tornotherapy patient-specific quality assurance
    Chaudhari, S.
    Goddu, S.
    Mutic, S.
    Grigsby, J.
    Low, D.
    MEDICAL PHYSICS, 2007, 34 (06) : 2498 - 2499
  • [7] Assessment of using a gamma index analysis for patient-specific quality assurance in Japan
    Anetai, Yusuke
    Sumida, Iori
    Kumazaki, Yu
    Kito, Satoshi
    Kurooka, Masahiko
    Ueda, Yoshihiro
    Otani, Yuki
    Narita, Yuichiro
    Kawamorita, Ryu
    Akita, Kazuhiko
    Kato, Takahiro
    Nakamura, Mitsuhiro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (10):
  • [8] The development of patient-specific quality assurance (PSQA) for a proton wobbling nozzle using PTSim Monte Carlo code
    Oanh, Luu Dang Hoang
    Aso, Tsukasa
    Wang, Chun-Chieh
    Chao, Tsi-Chian
    Cho, I-Chun
    RADIATION PHYSICS AND CHEMISTRY, 2024, 223
  • [9] Gamma Passing Rate - Correlation with Patient Specific Quality Assurance Devices using Two Photon energies in Pelvic IMRT
    Nigam, Jitendra
    Kumar, Piyush
    Singh, P. P.
    Silambarasan, N. S.
    Navitha, S.
    INDIAN JOURNAL OF PURE & APPLIED PHYSICS, 2022, 60 (09) : 773 - 782
  • [10] Deep Learning for Patient-Specific Quality Assurance: Predicting Gamma Passing Rates for IMRT Based on Delivery Fluence Informed by log Files
    Huang, Ying
    Pi, Yifei
    Ma, Kui
    Miao, Xiaojuan
    Fu, Sichao
    Zhu, Zhen
    Cheng, Yifan
    Zhang, Zhepei
    Chen, Hua
    Wang, Hao
    Gu, Hengle
    Shao, Yan
    Duan, Yanhua
    Feng, Aihui
    Zhuo, Weihai
    Xu, Zhiyong
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2022, 21