Numerical study on the performance of mixed flow blood pump with superhydrophobic surface

被引:3
|
作者
Li, Chengcheng [1 ]
Qiu, Huihe [2 ]
Ma, Jianying [3 ]
Wang, Ying [1 ,2 ]
机构
[1] Univ Shanghai Sci & Technol, Sch Energy & Power Engn, Shanghai Key Lab Multiphase Flow & Heat Transfer, Shanghai 200093, Peoples R China
[2] Hong Kong Univ Sci & Technol, Dept Mech & Aerosp Engn, Clear Water Bay, Hong Kong, Peoples R China
[3] Fudan Univ, Shanghai Inst Cardiovasc Dis, Zhongshan Hosp, Dept Cardiol, Shanghai 200032, Peoples R China
关键词
Mixed flow blood pump; Superhydrophobic surface; Hemolysis index; Blade wrap angles; Numerical simulation; SLIP; REDUCTION; MODELS; DAMAGE;
D O I
10.1007/s11517-023-02880-5
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
To meet the clinical status of the wide application of percutaneous mechanical circulatory support, this paper selects the mixed flow blood pump applied with superhydrophobic surface as the research object. The Navier slip model was used to simulate the slip characteristics of superhydrophobic surface, and the effects of the blade wrap angle and the superhydrophobic surface on the performance of the mixed flow blood pump are studied by numerical simulation. The results show that (1) considering the head, hydraulic efficiency, and hemolysis index of the blood pump, the optimal value of the blade wrap angle of the mixed flow blood pump in this paper is 60 degrees. (2) The hydraulic efficiency of the blood pump with superhydrophobic surface is improved, and the maximum growth rate is about 13.9%; superhydrophobic surface can reduce the hemolysis index of blood pump under various working conditions, and the maximum reduction rate of hemolysis index of blood pump is 22.9%. (3) The variation trends of blood pump head, hydraulic efficiency, and hemolysis index with the increased rotating speed before and after setting superhydrophobic slip boundary conditions are the same as their original variation trends.
引用
收藏
页码:3103 / 3121
页数:19
相关论文
共 50 条
  • [1] Numerical study on the performance of mixed flow blood pump with superhydrophobic surface
    Chengcheng Li
    Huihe Qiu
    Jianying Ma
    Ying Wang
    Medical & Biological Engineering & Computing, 2023, 61 : 3103 - 3121
  • [2] Numerical study on the performance of centrifugal blood pump with superhydrophobic surface
    Li, Chengcheng
    Qiu, Huihe
    Ma, Jianying
    Wang, Ying
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2022, 45 (12) : 1028 - 1036
  • [3] Numerical simulation and performance analysis of mixed flow blood pump
    Luo J.
    Huang D.
    Xu B.
    Huang, Diangui (dghuang@usst.edu.cn), 1600, West China Hospital, Sichuan Institute of Biomedical Engineering (37): : 296 - 303
  • [4] Numerical study on the performance of an interventional microaxial blood pump with superhydrophobic surface
    Geng, Xuesong
    Li, Chengcheng
    Ma, Jianying
    Qiu, Huihe
    Cui, Guomin
    Wang, Ying
    PHYSICS OF FLUIDS, 2025, 37 (01)
  • [5] Numerical Investigation of the Influence of Blade Radial Gap Flow on Axial Blood Pump Performance
    Liu, Guang-Mao
    Jin, Dong-Hai
    Zhou, Jian-Ye
    Zhang, Yan
    Chen, Hai-bo
    Sun, Han-Song
    Hu, Sheng-shou
    Gui, Xing-Min
    ASAIO JOURNAL, 2019, 65 (01) : 59 - 69
  • [6] NUMERICAL AND EXPERIMENTAL STUDY OF INNER FLOW AND HYDRAULIC PERFORMANCE OF THE MIXED-FLOW PUMP WITH LOW SPECIFIC SPEED
    Yuan, Jianping
    Fan, Meng
    Li, Yanjun
    Fu, Yanxia
    Lu, Rong
    PROCEEDINGS OF THE ASME FLUIDS ENGINEERING DIVISION SUMMER MEETING, 2018, VOL 3, 2018,
  • [7] Numerical study flow and heat transfer characteristics of superhydrophobic single tube bundle surface and wake flow
    Zhou, S. D.
    Shi, X. M.
    Yu, Q. X.
    Lin, Feng
    Wang, Z. C.
    MEASUREMENT & CONTROL, 2025, 58 (02) : 196 - 207
  • [8] A numerical study of the effects of superhydrophobic surface on skin-friction drag in turbulent channel flow
    Park, Hyunwook
    Park, Hyungmin
    Kim, John
    PHYSICS OF FLUIDS, 2013, 25 (11)
  • [9] Hemodynamic investigation and in vitro evaluation of a novel mixed-flow blood pump
    Qu, Yifei
    Guo, Ziyu
    Zhang, Jing
    Li, Guiling
    Zhang, Song
    Li, Donghai
    ARTIFICIAL ORGANS, 2022, 46 (08) : 1533 - 1543
  • [10] Numerical Study and Structural Improvement on Mixed-Flow Pump Impeller
    Xie, Rong
    Shan, Yu-jiao
    Wang, Xiao-fang
    Li, Liang
    2010 ASIA-PACIFIC POWER AND ENERGY ENGINEERING CONFERENCE (APPEEC), 2010,