Efficiency and safety of high-power ablation guided by Lesion size index: An ex vivo porcine heart study

被引:1
|
作者
Huo, Shengqi [1 ]
Wang, Qian [1 ]
Jiang, Yue [1 ]
Shi, Wei [1 ]
Luo, Pengcheng [2 ]
Guo, Junyi [1 ]
Peng, Dewei [1 ]
Zhu, Mengying [1 ]
Men, Lintong [1 ]
Jiang, Tao [2 ]
Wang, Moran [1 ]
Peng, Lulu [1 ]
Huang, Bingyu [1 ]
Shi, Meijing [4 ]
Zhu, Bo [5 ]
Bai, Rong [6 ]
Day, John D. D. [3 ]
Lv, Jiagao [1 ]
Lin, Li [1 ]
机构
[1] Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Div Cardiol,Dept Internal Med, Jiefang Ave 1095,POB 430030, Wuhan, Peoples R China
[2] Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Dept Geriatr, Wuhan, Peoples R China
[3] Intermt Med Ctr, Intermt Med Ctr Heart Inst, Salt Lake City, UT USA
[4] Gen Hosp Yangtze River Shipping, Dept Emergence, Wuhan, Peoples R China
[5] Jianghan Univ, Dept Internal Med, Div Cardiol, Hubei 3 Peoples Hosp, Wuhan, Peoples R China
[6] Capital Med Univ, Beijing Anzhen Hosp, Dept Cardiol, Beijing, Peoples R China
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2023年 / 46卷 / 06期
关键词
atrial fibrillation; high power; lesion size index; radiofrequency catheter ablation; SHORT-DURATION ABLATION; ATRIAL-FIBRILLATION; CONTACT FORCE; CATHETER ABLATION; PREDICTORS;
D O I
10.1111/pace.14659
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundAlthough Lesion size index (LSI) has been reported to highly predict radiofrequency lesion size in vitro, its accuracy in lesion size and steam pop estimation has not been well investigated for every possible scenario. MethodsInitially, radiofrequency ablations were performed on porcine myocardial slabs at various power, CF, and time settings with blinded LSI. Subsequently, radiofrequency power at 20, 30, 40, 50, and 60 W was applied at CF values of 5, 10, 20, and 30 g to reach target LSIs of 4, 5, 6, and 7. Lesion size and steam pops were recorded for each ablation. ResultsLesion size was positively correlated with LSI regardless of power settings (p < 0.001). The linear correlation coefficients of lesion size and LSI decreased at higher power settings. At high power combined with high CF settings (50 W/20 g), lesion depth and LSI showed an irrelevant correlation (p = 0.7855). High-power ablation shortened ablation time and increased the effect of resistive heating. LSI could predict the risk of steam pops at high-power settings with the optimal threshold of 5.65 (sensitivity, 94.1%; specificity, 46.1%). The ablation depth of the heavy heart was shallower than that of the light heart under similar ablation settings. ConclusionsLSI could predict radiofrequency lesion size and steam pops at high power settings in vitro, while synchronous high power and high CF should be avoided. Lighter hearts require relatively lower ablation settings to create appropriate ablation depth.
引用
收藏
页码:487 / 497
页数:11
相关论文
共 42 条
  • [41] Acute oesophageal safety and long-term follow-up of AI-guided high-power short-duration with 50 W for atrial fibrillation ablation
    Mueller, Julian
    Nentwich, Karin
    Berkovitz, Artur
    Ene, Elena
    Sonne, Kai
    Zhuravlev, Vitaly
    Chakarov, Ivaylo
    Barth, Sebastian
    Waechter, Christian
    Behnes, Michael
    Halbfass, Philipp
    Deneke, Thomas
    EUROPACE, 2023, 25 (04): : 1379 - 1391
  • [42] Pulmonary vein isolation using pulsed field ablation vs. high-power short-duration radiofrequency ablation in paroxysmal atrial fibrillation: efficacy, safety, and long-term follow-up (PRIORI study)
    Reinsch, Nico
    Fueting, Anna
    Hartl, Stefan
    Hoewel, Dennis
    Rausch, Eva
    Lin, Yali
    Kasparian, Karampet
    Neven, Kars
    EUROPACE, 2024, 26 (07):