AAPM Task Group Report 306: Quality control and assurance for tomotherapy: An update to Task Group Report 148

被引:23
作者
Chen, Quan [1 ]
Rong, Yi [2 ]
Burmeister, Jay W. W. [3 ,4 ]
Chao, Edward H. H. [5 ]
Corradini, Nathan A. A. [6 ]
Followill, David S. S. [7 ]
Li, X. Allen [8 ]
Liu, An [9 ]
Qi, X. Sharon [10 ]
Shi, Hairong [11 ]
Smilowitz, Jennifer B. B. [12 ]
机构
[1] City Hope Med Ctr, Radiat Oncol, Duarte, CA 91010 USA
[2] Mayo Clin Hosp, Dept Radiat Oncol, Phoenix, AZ USA
[3] Gershenson ROC, Karmanos Canc Ctr, Detroit, MI USA
[4] Wayne State Univ, Dept Oncol, Sch Med, Detroit, MI USA
[5] Accuray Inc, Madison, WI USA
[6] Clin Luganese Moncucco, Radiat Oncol, Lugano, Switzerland
[7] Univ Texas MD Anderson Canc Ctr, Radiat Phys, Houston, TX USA
[8] Med Coll Wisconsin, Radiat Oncol, Milwaukee, WI USA
[9] City Hope Med Ctr, Radiat Oncol, Duarte, CA USA
[10] UCLA Sch Med, Radiat Oncol, Los Angeles, CA USA
[11] Oklahoma Canc Specialists & Res Inst, Radiat Oncol, Tulsa, OK USA
[12] Univ Wisconsin, Human Oncol & Med Phys, Madison, WI USA
关键词
quality assurance; tomotherapy; DIODE-ARRAY DOSIMETER; PORTAL DOSE IMAGES; ROTATIONAL OUTPUT VARIATION; GUIDED RADIATION-THERAPY; HIGH-ENERGY PHOTON; STEP-AND-SHOOT; HELICAL TOMOTHERAPY; FAILURE MODE; DYNAMIC JAW; ANGULAR-DEPENDENCE;
D O I
10.1002/mp.16150
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Since the publication of AAPM Task Group (TG) 148 on quality assurance (QA) for helical tomotherapy, there have been many new developments on the tomotherapy platform involving treatment delivery, on-board imaging options, motion management, and treatment planning systems (TPSs). In response to a need for guidance on quality control (QC) and QA for these technologies, the AAPM Therapy Physics Committee commissioned TG 306 to review these changes and make recommendations related to these technology updates. The specific objectives of this TG were (1) to update, as needed, recommendations on tolerance limits, frequencies and QC/QA testing methodology in TG 148, (2) address the commissioning and necessary QA checks, as a supplement to Medical Physics Practice Guidelines (MPPG) with respect to tomotherapy TPS and (3) to provide risk-based recommendations on the new technology implemented clinically and treatment delivery workflow. Detailed recommendations on QA tests and their tolerance levels are provided for dynamic jaws, binary multileaf collimators, and Synchrony motion management. A subset of TPS commissioning and QA checks in MPPG 5.a. applicable to tomotherapy are recommended. In addition, failure mode and effects analysis has been conducted among TG members to obtain multi-institutional analysis on tomotherapy-related failure modes and their effect ranking.
引用
收藏
页码:E25 / E52
页数:28
相关论文
共 139 条
[91]  
Papanikolaou N., 2004, 85 AAPM TASK GROUP
[92]  
Papanikolaou N., 2010, J CANC SCI THER, V2, P145, DOI DOI 10.4172/1948-5956.1000040
[93]   An efficient procedure for tomotherapy treatment plan verification using the on-board detector [J].
Pisaturo, O. ;
Mieville, F. ;
Tercier, P-A ;
Allal, A. S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (04) :1625-1639
[94]   TransitQA - A new method for transit dosimetry of Tomotherapy patients [J].
Pisaturo, Olivier ;
Mieville, Frederic ;
Tercier, Pierre-Alain ;
Allal, Abdelkarim Said .
MEDICAL PHYSICS, 2018, 45 (01) :438-447
[95]   Compensation of intrafractional motion for lung stereotactic body radiotherapy (SBRT) on helical TomoTherapy [J].
Price, Alex ;
Chen, Jiawei ;
Chao, Edward ;
Schnarr, Eric ;
Schreiber, Eric ;
Lu, Lan ;
Cox, Andrea ;
Chang, Sha ;
Lian, Jun .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (02)
[96]   The effect of temporal HU variations on the uncertainty of dose recalculations performed on MVCT images [J].
Pukala, J. ;
Meeks, S. L. ;
Bova, F. J. ;
Langen, K. M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (24) :7829-7841
[97]   Helical tomotherapy with dynamic running-start-stop delivery compared to conventional tomotherapy delivery [J].
Rong, Yi ;
Chen, Yu ;
Shang, Lu ;
Zuo, Li ;
Lu, Weiguo ;
Chen, Quan .
MEDICAL PHYSICS, 2014, 41 (05)
[98]   A planning study for palliative spine treatment using StatRT and megavoltage CT simulation [J].
Rong, Yi ;
Yadav, Poonam ;
Paliwal, Bhudatt ;
Shang, Lu ;
Welsh, James S. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (01) :97-107
[99]   Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers [J].
Rowshanfarzad, Pejman ;
Sabet, Mahsheed ;
O'Connor, Daryl J. ;
McCowan, Peter M. ;
McCurdy, Boyd M. C. ;
Greer, Peter B. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (06) :203-214
[100]   Lung and liver SBRT using helical tomotherapy - a dosimetric comparison of fixed jaw and dynamic jaw delivery [J].
Rudofsky, Leonie ;
Aynsley, Eleanor ;
Beck, Sebastian ;
Schubert, Kai ;
Habl, Gregor ;
Krause, Sonja ;
Debus, Juergen ;
Sterzing, Florian .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (03) :114-121