Validation of cardiac index measured by four-dimensional cardiac magnetic resonance flow against the invasively measured cardiac index in patients with pulmonary hypertension
被引:1
作者:
Romeih, Soha
[1
,2
,8
]
Elkafrawy, Fatma
[1
,3
]
Shaaban, Mahmoud
[1
,2
]
Hassan, Ahmed
[4
,5
]
Samuel, Irini
[4
,5
]
Gergis, Micheal
[5
]
El Mozy, Wesam
[1
,6
]
Yacoub, Magdi
[1
,7
]
机构:
[1] Aswan Heart Ctr, Radiol Dept, Aswan, Egypt
[2] Tanta Univ, Cardiol Dept, Tanta, Egypt
[3] Alexandria Univ, Radiol Dept, Alexandria, Egypt
[4] Cairo Univ, Cardiol Dept, Giza, Egypt
[5] Aswan Heart Ctr, Cardiol Dept, Aswan, Egypt
[6] Cairo Univ, Radiol Dept, Giza, Egypt
[7] Imperial Collage London, London, England
[8] Aswan Heart Ctr, Aswan, Egypt
关键词:
CMR flow;
Fick method;
noninvasive CI measurement;
QUANTIFICATION;
ARTERY;
HEART;
THERMODILUTION;
HEMODYNAMICS;
DIAGNOSIS;
ACCURATE;
OUTPUT;
VOLUME;
VALVE;
D O I:
10.1002/pul2.12203
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
The purpose of this study was to validate cardiac index (CI) measured noninvasively by four-dimensional (4D) and two-dimensional (2D) cardiovascular magnetic resonance (CMR) flows against the invasively measured CI by right heart catheterization (RHC) in patients with pulmonary hypertension (PH). Thirty patients with PH (mean age: 32 +/- 10 years) were included. 4D and 2D flow measured CI within 24 h from RHC measured CI invasively. Qualitative analysis of 4D pulmonary flow (vortex presence and eccentricity of flow) was performed.All patients had helical right-sided flow with vortex formation; the mean vortex diameter was 29 +/- 7 mm, occupying 69% of the main pulmonary artery (MPA) lumen. MPA was dilated (42 +/- 9 mm). Mean CI measured by 4D flow CMR was closer to mean CI measured invasively (indirect Fick method CI = 2.1 +/- 0.8 L/min/m(2) vs. PA 4D flow = 2.3 +/- 0.7 L/min/m(2) "bias 0.22 +/- 0.25 L/min, p = 0.001," and Ao 4D flow = 2.3 +/- 0.7 L/min/m(2) "bias 0.2 +/- 0.28 L/min, p = 0.001"), while 2D flow had a higher mean CI (PA 2D flow = 2.5 +/- 0.7 L/min/m(2) "bias 0.45 +/- 0.7 L/min, p = 0.001" and Ao 2D flow = 2.5 +/- 0.8 L/min/m(2) "bias 0.45 +/- 0.67 L/min, p = 0.001"). The correlation coefficients among the different comparisons of CI showed: a low correlation between the 2D flow-indirect Fick method (Ao r(2) = 0.37, PA r(2) = 0.32) and a high correlation between the 4D flow-indirect Fick method (Ao r(2) = 0.86, PA r(2) = 0.89).There is an excellent agreement between CI measured by 4D flow and CI measured invasively. 4D flow, a noninvasive imaging technique, could accurately measure CI better than the conventional 2D flow in patients with PH.