Localization of reference points in electromagnetic tracking data and their application for treatment error detection in interstitial breast brachytherapy

被引:3
|
作者
Duerrbeck, Christopher [1 ,2 ,3 ]
Schuster, Sabrina [1 ,2 ]
Sauer, Birte Christina [1 ,2 ]
Abu-Hossin, Nadin [1 ,2 ]
Strnad, Vratislav [1 ,2 ]
Fietkau, Rainer [1 ,2 ]
Bert, Christoph [1 ,2 ]
机构
[1] Friedrich Alexander Univ Erlangen Nurnberg FAU, Dept Radiat Oncol, Univ Klinikum Erlangen, Erlangen, Germany
[2] Comprehens Canc Ctr Erlangen EMN CCC ER EMN, Erlangen, Germany
[3] Univ Klinikum Erlangen, Dept Radiat Oncol, Univ Str 27, D-91054 Erlangen, Germany
关键词
brachytherapy; electromagnetic tracking; error detection; quality assurance; DOSE-RATE BRACHYTHERAPY; TREATMENT VERIFICATION; CATHETER RECONSTRUCTION; DOSIMETRY; ACCURACY; DELIVERY; SYSTEM;
D O I
10.1002/mp.16629
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundElectromagnetic tracking (EMT) is a promising technology that holds great potential to advance patient-specific pre-treatment verification in interstitial brachytherapy (iBT). It allows easy determination of the implant geometry without line-of-sight restrictions and without dose exposure to the patient. What it cannot provide, however, is a link to anatomical landmarks, such as the exit points of catheters or needles on the skin surface. These landmarks are required for the registration of EMT data with other imaging modalities and for the detection of treatment errors such as incorrect indexer lengths, and catheter or needle shifts. PurposeTo develop an easily applicable method to detect reference points in the positional data of the trajectory of an EMT sensor, specifically the exit points of catheters in breast iBT, and to apply the approach to pre-treatment error detection. MethodsSmall metal objects were attached to catheter fixation buttons that rest against the breast surface to intentionally induce a local, spatially limited perturbation of the magnetic field on which the working principle of EMT relies. This perturbation can be sensed by the EMT sensor as it passes by, allowing it to localize the metal object and thus the catheter exit point. For the proof-of-concept, different small metal objects (magnets, washers, and bushes) and EMT sensor drive speeds were used to find the optimal parameters. The approach was then applied to treatment error detection and validated in-vitro on a phantom. Lastly, the in-vivo feasibility of the approach was tested on a patient cohort of four patients to assess the impact on the clinical workflow. ResultsAll investigated metal objects were able to measurably perturb the magnetic field, which resulted in missing sensor readings, that is two data gaps, one for the sensor moving towards the tip end and one when retracting from there. The size of the resulting data gaps varied depending on the choice of gap points used for calculation of the gap size; it was found that the start points of the gaps in both directions showed the smallest variability. The median size of data gaps was & LE;8 mm for all tested materials and sensor drive speeds. The variability of the determined object position was & LE;0.5 mm at a speed of 1.0 cm/s and & LE;0.7 mm at 2.5 cm/s, with an increase up to 2.3 mm at 5.0 cm/s. The in-vitro validation of the error detection yielded a 100% detection rate for catheter shifts of & GE;2.2 mm. All simulated wrong indexer lengths were correctly identified. The in-vivo feasibility assessment showed that the metal objects did not interfere with the routine clinical workflow. ConclusionsThe developed approach was able to successfully detect reference points in EMT data, which can be used for registration to other imaging modalities, but also for treatment error detection. It can thus advance the automation of patient-specific, pre-treatment quality assurance in iBT.
引用
收藏
页码:5772 / 5783
页数:12
相关论文
共 9 条
  • [1] Electromagnetic tracking for treatment verification in interstitial brachytherapy
    Bert, Christoph
    Kellermeier, Markus
    Tanderup, Kari
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2016, 8 (05) : 448 - 453
  • [2] Error detection using an electromagnetic tracking system in multi-catheter breast interstitial brachytherapy
    Masitho, Siti
    Kallis, Karoline
    Strnad, Vratislav
    Fietkau, Rainer
    Bert, Christoph
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (20)
  • [3] Assessment of the implant geometry in fractionated interstitial HDR breast brachytherapy using an electromagnetic tracking system
    Kellermeier, Markus
    Fietkau, Rainer
    Strnad, Vratislav
    Bert, Christoph
    BRACHYTHERAPY, 2018, 17 (01) : 94 - 102
  • [4] Electromagnetic tracking (EMT) technology for improved treatment quality assurance in interstitial brachytherapy
    Kellermeier, Markus
    Herbolzheimer, Jens
    Kreppner, Stephan
    Lotter, Michael
    Strnad, Vratislav
    Bert, Christoph
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (01): : 211 - 222
  • [5] Influence and compensation of patient motion in electromagnetic tracking based quality assurance in interstitial brachytherapy of the breast
    Duerrbeck, Christopher
    Gulde, Sarah
    Abu-Hossin, Nadin
    Fietkau, Rainer
    Strnad, Vratislav
    Bert, Christoph
    MEDICAL PHYSICS, 2022, 49 (04) : 2652 - 2662
  • [6] Automating implant reconstruction in interstitial brachytherapy of the breast: A hybrid approach combining electromagnetic tracking and image segmentation
    Durrbeck, Christopher
    Sauer, Birte Christina
    Lotter, Michael
    Kreppner, Stephan
    Strnad, Vratislav
    Fietkau, Rainer
    Bert, Christoph
    RADIOTHERAPY AND ONCOLOGY, 2022, 176 : 172 - 178
  • [7] Accuracy of dwell position detection with a combined electromagnetic tracking brachytherapy system for treatment verification in pelvic brachytherapy
    van Heerden, Laura
    Schiphof-Godart, Jeremy
    Christianen, Miranda
    Mens, Jan-Willem
    Franckena, Martine
    Maenhout, Metha
    Hoogeman, Mischa
    Kolkman-Deurloo, Inger-Karine
    RADIOTHERAPY AND ONCOLOGY, 2021, 154 : 249 - 254
  • [8] Time-resolved clinical dose volume metrics, calculations and predictions based on source tracking measurements and uncertainties to aid treatment verification and error detection for HDR brachytherapy-a proof-of-principle study
    van Wagenberg, Teun
    Voncken, Robert
    van Beveren, Celine
    Berbee, Maaike
    van Limbergen, Evert
    Verhaegen, Frank
    Fonseca, Gabriel Paiva
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (13)
  • [9] Estimating follow-up CTs from geometric deformations of catheter implants in interstitial breast brachytherapy: A feasibility study using electromagnetic tracking
    Duerrbeck, Christopher
    Schulz, Moritz
    Pflaum, Leonie
    Kallis, Karoline
    Geimer, Tobias
    Abu-Hossin, Nadin
    Strnad, Vratislav
    Maier, Andreas
    Fietkau, Rainer
    Bert, Christoph
    MEDICAL PHYSICS, 2023, 50 (09) : 5793 - 5805