Probing the Influence of Hybrid Thread Design on Biomechanical Response of Dental Implants: Finite Element Study and Experimental Validation

被引:7
|
作者
Chakraborty, Arindam [1 ]
Sahare, Kalash Darshan [1 ]
Datta, Pallab [2 ]
Majumder, Santanu [1 ]
Roychowdhury, Amit [1 ]
Basu, Bikramjit [3 ]
机构
[1] Indian Inst Engn Sci & Technol, Dept Aerosp Engn & Appl Mech, PO Bot Garden, Howrah 711103, W Bengal, India
[2] Natl Inst Pharmaceut Educ & Res, Dept Pharmaceut, Kolkata 700054, W Bengal, India
[3] Indian Inst Sci, Mat Res Ctr, Bengaluru 560012, Karnataka, India
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2023年 / 145卷 / 01期
关键词
dental implant design; thread design; finite element analysis; bone/implant contact; initial stability; additive manufacturing; STRESS-DISTRIBUTION; BONE TISSUE; OSSEOINTEGRATION; GEOMETRY; STABILITY; TORQUE;
D O I
10.1115/1.4054984
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This study aimed to perform quantitative biomechanical analysis for probing the effect of varying thread shapes in an implant for improved primary stability in prosthodontics surgery. Dental implants were designed with square (SQR), buttress (BUT), and triangular (TRI) thread shapes or their combinations. Cone-beam computed tomography images of mandible molar zones in human subjects belonging to three age groups were used for virtual implantation of the designed implants, to quantify patient-specific peri-implant bone microstrain, using finite element analyses. The in silico analyses were carried out considering frictional contact to simulate immediate loading with a static masticatory force of 200 N. To validate computational biomechanics results, compression tests were performed on three-dimensional printed implants having the investigated thread architectures. Bone/implant contact areas were also quantitatively assessed. It was observed that, bone/implant contact was maximum for SQR implants followed by BUT and TRI implants. For all the cases, peak microstrain was recorded in the cervical cortical bone. The combination of different thread shapes in the middle or in the apical part (or both) was demonstrated to improve peri-implant microstrain, particularly for BUT and TRI. Considering 1500-2000 microstrain generates in the peri-implant bone during regular physiological functioning, BUT-SQR, BUT-TRI-SQR, TRI-SQR-BUT, SQR, and SQR-BUT-TRI design concepts were suitable for younger; BUT-TRI-SQR, BUT-SQR-TRI, TRI-SQR-BUT, SQR-BUT, SQR-TRI for middle-aged, and BUT-TRI-SQR, BUT-SQR-TRI, TRI-BUT-SQR, SQR, and SQR-TRI for the older group of human patients.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] The influence of implant body and thread design of mini dental implants on the loading of surrounding bone: a finite element analysis
    Toth, Arpad
    Hasan, Istabrak
    Bourauel, Christoph
    Mundt, Torsten
    Biffar, Reiner
    Heinemann, Friedhelm
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2017, 62 (04): : 393 - 405
  • [2] UNDERSTANDING THE BIOMECHANICAL RESPONSE OF PROGRESSIVE THREAD DENTAL IMPLANTS USING MULTI-SCALE FINITE ELEMENT ANALYSIS
    Chakraborty, Arindam
    Sahare, Kalash Darshan
    Majumder, Santanu
    Chowdhury, Amit Roy
    INTERNATIONAL JOURNAL FOR MULTISCALE COMPUTATIONAL ENGINEERING, 2024, 22 (04) : 31 - 44
  • [3] Biomechanical Design Application on the Effect of Different Occlusion Conditions on Dental Implants with Different Positions-A Finite Element Analysis
    Lin, Pei-Ju
    Su, Kuo-Chih
    APPLIED SCIENCES-BASEL, 2020, 10 (17):
  • [4] Biomechanical Investigation of Patient-Specific Porous Dental Implants: A Finite Element Study
    Mehboob, Hassan
    Ouldyerou, Abdelhak
    Ijaz, Muhammad Farzik
    APPLIED SCIENCES-BASEL, 2023, 13 (12):
  • [5] Biomechanical Analysis of Axial Gradient Porous Dental Implants: A Finite Element Analysis
    Zhang, Chunyu
    Wang, Yuehong
    JOURNAL OF FUNCTIONAL BIOMATERIALS, 2023, 14 (12)
  • [6] Biomechanical analysis of different fixed dental restorations on short implants: a finite element study
    Wagner, Christian
    Herberg, Samira
    Bourauel, Christoph
    Stark, Helmut
    Doersam, Istabrak
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2023, 68 (03): : 241 - 250
  • [7] Finite-element study of biomechanical explanations for bone loss around dental implants
    Merdji A.
    Taharou B.
    Hillstrom R.
    Benaissa A.
    Roy S.
    Chong P.L.
    Mukdadi O.
    Della N.
    Bouiadjra B.-A.B.
    Journal of Long-Term Effects of Medical Implants, 2020, 30 (01) : 21 - 30
  • [8] Biomechanical analysis of peri-prosthetic bone response to hybrid threaded zirconia dental implants: An in silico model
    Mishra, Deepa
    Basu, Bikramjit
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2024, 150
  • [9] Probing combinational influence of design variables on bone biomechanical response around dental implant-supported fixed prosthesis
    Chakraborty, Arindam
    Datta, Pallab
    Kumar, Cheruvu Siva
    Majumder, Santanu
    Roychowdhury, Amit
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2022, 110 (10) : 2338 - 2352
  • [10] Design of customized coated dental implants using finite element analysis
    Dommeti, Vamsi Krishna
    Pramanik, Sumit