Blood-pH Optical Measurement: A Model to Compensate for the Effects of Temperature

被引:4
作者
Goldoni, Daniele [1 ,2 ]
Ferrari, Alberto [1 ,2 ]
Piccini, Mattia [1 ,2 ]
Cattini, Stefano [1 ]
Molinari, Riccardo [3 ]
Rovati, Luigi [1 ,2 ]
机构
[1] Univ Modena & Reggio Emilia, Dept Engn Enzo Ferrari, I-41125 Modena, Italy
[2] Democtr Fdn Mirandola, TPM, Sci & Technol Pk Med, I-41037 Modena, Italy
[3] Med Grp, Medolla, Italy
关键词
8-hydroxypyrene-1,3,6-trisulfonic acid trisodium salt (HPTS); biomedical sensor; blood; extracorporeal circulation (ECC); fluorescence; phosphate buffer saline (PBS); pH sensor; temperature; temperature-compensated model; EXTRACORPOREAL MEMBRANE-OXYGENATION; INSTRUMENT; SENSOR; FLUORESCENCE; SYSTEM;
D O I
10.1109/TIM.2023.3250282
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Under physiological conditions, the human body maintains blood pH within [7.36, 7.44] pH. Small deviations from this range can reveal the onset of pathological states and worsen the patient's condition. This article reports the performance analysis of a real-time, noninvasive pH-measuring system for extracorporeal circulation (ECC). In particular, this study focuses on the analysis of the effects that the measurand temperature may have on the error in estimating blood pH. Even if the blood temperature in ECC is often thermostated at 37 degrees C, there are treatments in which the blood temperature is varied by a few Celsius degrees, and the exploited measurement principlefluorescence-is known to be affected by temperature. First, we verified that the temperature-induced error could exceed the maximum permissible measurement error of +/- 0.04 pH. Hence, a linear-correction factor for temperature compensation was proposed. The results obtained showed how the simple addition to the measuring system of a temperature sensor and the use of a linear-correction factor can effectively allow maintaining the measurement error within the +/- 0.04-pH range, even when the fluid-phosphate buffer saline (PBS) and blood-temperature is varied in the range [30 degrees C, 39 degrees C].
引用
收藏
页数:8
相关论文
共 50 条
  • [31] EFFECTS OF TEMPERATURE AND PH LEVEL ON NEUTROPHIL FILTERABILITY
    AOSHIBA, K
    NAGAI, A
    KONNO, K
    CLINICAL HEMORHEOLOGY, 1993, 13 (01): : 23 - 34
  • [32] Fluorescent properties of organic carbon in cave dripwaters: Effects of filtration, temperature and pH
    Hartland, A.
    Fairchild, I. J.
    Lead, J. R.
    Baker, A.
    SCIENCE OF THE TOTAL ENVIRONMENT, 2010, 408 (23) : 5940 - 5950
  • [33] Optical properties of human blood as a function of temperature
    Randeberg, LL
    Hagen, AJD
    Svaasand, LO
    LASERS IN SURGERY: ADVANCED CHARACTERIZATION, THERAPEUTICS, AND SYSTEMS XII, 2002, 4609 : 20 - 28
  • [34] Combined effects of temperature, pH and water activity on predictive ability of microbial kinetic inactivation model
    Gil, Maria M.
    Miller, Fatima A.
    Brandao, Teresa R. S.
    Silva, Cristina L. M.
    9TH INTERNATIONAL CONFERENCE ON PREDICTIVE MODELLING IN FOOD, 2016, 7 : 67 - 70
  • [35] Effects of temperature, pH and sugar binding on the structures of lectins ebulin f and SELfd
    Carrillo, Celia
    Cordoba-Diaz, Damian
    Cordoba-Diaz, Manuel
    Girbes, Tomas
    Jimenez, Pilar
    FOOD CHEMISTRY, 2017, 220 : 324 - 330
  • [36] Influence and correction of temperature on optical measurement for fat and protein contents in a complex food model system
    Zhang, Xuedian
    Chang, Min
    Xing, Liuji
    Hu, Jiahui
    INFRARED PHYSICS & TECHNOLOGY, 2010, 53 (03) : 177 - 181
  • [37] Optical Probing of Local pH and Temperature in Complex Fluids with Covalently Functionalized, Semiconducting Carbon Nanotubes
    Kwon, Hyejin
    Kim, Mijin
    Meany, Brendan
    Piao, Yanmei
    Powell, Lyndsey R.
    Wang, YuHuang
    JOURNAL OF PHYSICAL CHEMISTRY C, 2015, 119 (07) : 3733 - 3739
  • [38] Combined optical measurement of dissolved oxygen tension (DOT), pH value, biomass and viscosity in shake flasks
    Dinter, Carl
    Vonester, David
    Flitsch, David
    Mertens, Moritz
    Tueschenboenner, Marc
    Hoffmann, Maximilian
    Buechs, Jochen
    Magnus, Jorgen
    BIOCHEMICAL ENGINEERING JOURNAL, 2024, 212
  • [39] A Review of Measurement Calibration and Interpretation for Seepage Monitoring by Optical Fiber Distributed Temperature Sensors
    Ghafoori, Yaser
    Vidmar, Andrej
    Riha, Jaromir
    Kryzanowski, Andrej
    SENSORS, 2020, 20 (19) : 1 - 23
  • [40] Optical temperature measurement method of premixed flames using a multi-laser system
    Jeon, Min-Gyu
    Doh, Deog-Hee
    Deguchi, Yoshihiro
    JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2021, 35 (06) : 2535 - 2542