Blood-pH Optical Measurement: A Model to Compensate for the Effects of Temperature

被引:4
作者
Goldoni, Daniele [1 ,2 ]
Ferrari, Alberto [1 ,2 ]
Piccini, Mattia [1 ,2 ]
Cattini, Stefano [1 ]
Molinari, Riccardo [3 ]
Rovati, Luigi [1 ,2 ]
机构
[1] Univ Modena & Reggio Emilia, Dept Engn Enzo Ferrari, I-41125 Modena, Italy
[2] Democtr Fdn Mirandola, TPM, Sci & Technol Pk Med, I-41037 Modena, Italy
[3] Med Grp, Medolla, Italy
关键词
8-hydroxypyrene-1,3,6-trisulfonic acid trisodium salt (HPTS); biomedical sensor; blood; extracorporeal circulation (ECC); fluorescence; phosphate buffer saline (PBS); pH sensor; temperature; temperature-compensated model; EXTRACORPOREAL MEMBRANE-OXYGENATION; INSTRUMENT; SENSOR; FLUORESCENCE; SYSTEM;
D O I
10.1109/TIM.2023.3250282
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Under physiological conditions, the human body maintains blood pH within [7.36, 7.44] pH. Small deviations from this range can reveal the onset of pathological states and worsen the patient's condition. This article reports the performance analysis of a real-time, noninvasive pH-measuring system for extracorporeal circulation (ECC). In particular, this study focuses on the analysis of the effects that the measurand temperature may have on the error in estimating blood pH. Even if the blood temperature in ECC is often thermostated at 37 degrees C, there are treatments in which the blood temperature is varied by a few Celsius degrees, and the exploited measurement principlefluorescence-is known to be affected by temperature. First, we verified that the temperature-induced error could exceed the maximum permissible measurement error of +/- 0.04 pH. Hence, a linear-correction factor for temperature compensation was proposed. The results obtained showed how the simple addition to the measuring system of a temperature sensor and the use of a linear-correction factor can effectively allow maintaining the measurement error within the +/- 0.04-pH range, even when the fluid-phosphate buffer saline (PBS) and blood-temperature is varied in the range [30 degrees C, 39 degrees C].
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Towards a Temperature Compensated Model for a Blood-pH Sensor in Extracorporeal Circulation
    Goldoni, Daniele
    Ferrari, Alberto
    Piccini, Mattia
    Cattini, Stefano
    Rovati, Luigi
    2022 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC 2022), 2022,
  • [2] On the Development of an Instrument for In-Line and Real-Time Monitoring of Blood-pH in Extracorporeal Circulation
    Cattini, Stefano
    Accorsi, Luca
    Truzzi, Stefano
    Rovati, Luigi
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2020, 69 (08) : 5640 - 5648
  • [3] An Integrated System to Compensate for Temperature Drift and Ageing in Non-invasive Blood Pressure Measurement
    Abderahman, Huthaifa N.
    Dajani, Hilmi R.
    Bolic, Miodrag
    Groza, Voicu Z.
    2016 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2016, : 301 - 306
  • [4] Temperature and pH effects on myoglobin optical absorption spectra
    Ciesielski, WA
    Arakaki, LSL
    Schenkman, KA
    Optical Diagnostics and Sensing V, 2005, 5702 : 151 - 159
  • [5] A distributed optical fibre fluorosensor for pH measurement
    Wallace, PA
    Campbell, M
    Yang, Y
    HolmesSmith, AS
    Uttamlal, M
    JOURNAL OF LUMINESCENCE, 1997, 72-4 : 1017 - 1019
  • [6] Temperature Measurement Using Optical Fiber Methods: Overview and Evaluation
    Mikolajek, Martin
    Martinek, Radek
    Koziorek, Jiri
    Hejduk, Stanislav
    Vitasek, Jan
    Vanderka, Ales
    Poboril, Radek
    Vasinek, Vladimir
    Hercik, Radim
    JOURNAL OF SENSORS, 2020, 2020 (2020)
  • [7] Optical sensing techniques for temperature measurement
    Reddy, B. Rami
    Kamma, Indumathi
    Kommidi, Praveena
    APPLIED OPTICS, 2013, 52 (04) : B33 - B39
  • [8] Effects of pH and temperature on metmyoglobin solubility in a model system
    Zhu, LG
    Brewer, MS
    MEAT SCIENCE, 2002, 61 (04) : 419 - 424
  • [9] Fluorescent molecular probe based optical fiber sensor dedicated to pH measurement of concrete
    Tariq, Ayedah
    Baydoun, Jalal
    Remy, Charlotte
    Ghasemi, Rasta
    Lefevre, Jean Pierre
    Mongin, Cedric
    Dauzeres, Alexandre
    Leray, Isabelle
    SENSORS AND ACTUATORS B-CHEMICAL, 2021, 327
  • [10] A novel method of optical fiber temperature measurement - Measuring the inner temperature of the cage motor in starting
    Jia, DP
    Lin, W
    Lin, YW
    ADVANCED PHOTONIC SENSORS: TECHNOLOGY AND APPLICATIONS, 2000, 4220 : 210 - 213