A case of supraventricular tachycardia misdiagnosis after dual-chamber implantable cardioverter defibrillator implantation with His bundle pacing

被引:0
|
作者
Sun, Xianchao [1 ]
Huang, Yue [1 ]
机构
[1] First Hosp Jiaxing, Jiaxing, Zhejiang, Peoples R China
关键词
atrial fibrillation; His bundle packing; implantable cardioverter defibrillator; supraventricular tachycardia; ventricular tachycardia; PREVENTION; RECIPIENTS; SURVIVAL;
D O I
10.1111/anec.13042
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The combined use of an implantable cardioverter defibrillator (ICD) and His bundle pacing (HBP) is believed to benefit patients with dilated cardiomyopathy (DCM) and atrial fibrillation but needs more guidance built on clinical experience. A patient who had an implanted dual-chamber ICD with HBP presented to our hospital with DCM and atrial fibrillation. In the present case, the criteria for differential diagnosis were set as "if all," which requires matching the morphological template, sudden onset, and interval stability. Ventricular tachycardia (VT) was misdiagnosed as supraventricular tachycardia due to morphological matching to sinus tachycardia. The misdiagnosis and mistaken VT treatment(ICD adaptive treatment) highlight the importance of personalized settings for differential diagnosis to avoid such a misdiagnosis or omission.
引用
收藏
页数:7
相关论文
共 41 条
  • [1] Implantable dual-chamber cardioverter-defibrillator-pacemaker
    Pfeiffer D.
    Mende M.
    Hagendorff A.
    Current Cardiology Reports, 2000, 2 (4) : 335 - 340
  • [2] Failure to deliver a shock in a dual-chamber implantable cardioverter-defibrillator: A case report
    Sousa, Pedro A.
    Candeias, Rui
    Jesus, Ilidio
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2015, 34 (7-8) : 495.e1 - 495.e5
  • [3] Performance of a dual-chamber implantable defibrillator algorithm for discrimination of ventricular from supraventricular tachycardia
    Kouakam, C
    Kacet, S
    Hazard, JR
    Ferraci, A
    Mansour, H
    Defaye, P
    Davy, JM
    Lambiez, M
    EUROPACE, 2004, 6 (01): : 32 - 42
  • [4] Initiation of Ventricular Tachycardia by Interruption of Pacemaker-Mediated Tachycardia in a Patient with a Dual-Chamber Implantable Cardioverter Defibrillator
    Van Gelder, Berry M.
    Meijer, Albert
    Dekker, Lukas R.
    Bracke, Frank A.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (09): : 1227 - 1230
  • [5] Delayed detection of ventricular tachycardia in a dual chamber rate adaptive pacing implantable cardioverter defibrillator: A case of intradevice interaction
    Shalaby, AA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (08): : 1164 - 1166
  • [6] Anesthetic management of cardiac tamponade after dual-chamber implantable cardioverter defibrillator implantation in a patient with dilated cardiomyopathy
    Sato, Y
    Horiguchi, T
    Nishikawa, T
    JOURNAL OF CLINICAL ANESTHESIA, 2004, 16 (07) : 554 - 556
  • [7] His-bundle pacing using dual-chamber cardioverter-defibrillator in patients with chronic atrial fibrillation: different pacing and sensing features
    Slawuta, Agnieszka
    Bialy, Dariusz
    Zysko, Dorota
    Mazur, Grzegorz
    Gajek, Jacek
    ACTA CARDIOLOGICA, 2017, 72 (03) : 343 - 344
  • [8] Clinical experience with a dual-chamber implantable cardioverter defibrillator to treat atrial tachyarrhythmias
    Gold, MR
    Sulke, N
    Schwartzman, DS
    Mehra, R
    Euler, DE
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (11) : 1247 - 1253
  • [9] Worldwide clinical experience with a new dual-chamber implantable cardioverter defibrillator system
    Schoels, W
    Swerdlow, CD
    Jung, W
    Stein, KM
    Seidl, K
    Haffajee, CJ
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (05) : 521 - 528
  • [10] A new classification algorithm for discrimination of ventricular from supraventricular tachycardia in a dual chamber implantable cardioverter defibrillator
    Korte, T
    Jung, W
    Wolpert, C
    Spehl, S
    Schumacher, B
    Esmailzadeh, B
    Lüderitz, B
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (01) : 70 - 73