Quantitative study on dose distribution of Freiburg flap for keloid high-dose-rate brachytherapy based on MatriXX

被引:0
作者
Ni, Jie [1 ]
Gan, Guanghui [1 ]
Xu, Xiaoting [1 ]
机构
[1] Soochow Univ, Affiliated Hosp 1, Radiat Therapy Ctr, 899 Pinghai St, Suzhou 215006, Jiangsu, Peoples R China
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2023年 / 24卷 / 09期
关键词
dose verification; HDR brachytherapy; MatriXX; AMERICAN BRACHYTHERAPY; TG-43; FORMALISM; DETECTOR ARRAY; DOSIMETRY; VERIFICATION;
D O I
10.1002/acm2.14118
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo quantify the dose distribution effect of insufficient scattering conditions in keloid HDR brachytherapy with Freiburg fFlap (FF) applicator. Materials and MethodsA phantom composed of FF applicator, MatriXX and solid water slices was designed and scanned for treatment planning. Bolus with different thicknesses were covered to offer different scatter conditions. Planar dose distributions were measured by MatriXX. The maximum value (Max), average value (Avg) and & gamma; passing rate (3 mm/3%) were evaluated by the software MyQA Platform. ResultsThe maximum and average doses measured by MatriXX were lower than the calculated values. The difference increased as field size decreased. The Max value, found at 0.86 cm level in the two tube case, was -20.0%, and the avg value was -11.9%. All the & gamma; values were less than 95%. This difference gradually decreased with increasing bolus thickness and the & gamma; values were significantly improved. ConclusionMatriXX could be used for dose verification of HDR brachytherapy with an FF applicator. When the FF applicator was applied for keloid, insufficient scattering conditions would cause an insufficient target dose. This difference could be reduced by covering the bolus with different thicknesses on the applicator. The smaller the field, the thicker the bolus required.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Sensitivity of dose-volume indices to computation settings in high-dose-rate prostate brachytherapy treatment plan evaluation
    van der Meer, Marjolein C.
    Bosman, Peter A. N.
    Pieters, Bradley R.
    Niatsetski, Yury
    van Wieringen, Niek
    Alderliesten, Tanja
    Bel, Arjan
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (04): : 66 - 74
  • [42] Monte Carlo dose calculations for high-dose-rate brachytherapy using GPU-accelerated processing
    Tian, Z.
    Zhang, M.
    Hrycushko, B.
    Albuquerque, K.
    Jiang, S. B.
    Jia, X.
    BRACHYTHERAPY, 2016, 15 (03) : 387 - 398
  • [43] An independent dose-to-point calculation program for the verification of high-dose-rate brachytherapy treatment planning
    Cohen, GN
    Amols, HI
    Zaider, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (04): : 1251 - 1258
  • [44] Ultrasound-planned high-dose-rate prostate brachytherapy: Dose painting to the dominant intraprostatic lesion
    Crook, Juanita
    Ots, Ana
    Gaztanaga, Miren
    Schmid, Matt
    Araujo, Cynthia
    Hilts, Michelle
    Batchelar, Deidre
    Parker, Brent
    Bachand, Francois
    Milette, Marie-Pierre
    BRACHYTHERAPY, 2014, 13 (05) : 433 - 441
  • [45] High-dose-rate prostate brachytherapy inverse planning on dose-volume criteria by simulated annealing
    Deist, T. M.
    Gorissen, B. L.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (03) : 1155 - 1170
  • [46] Image-guided high-dose-rate brachytherapy of head and neck - a case series study
    Cisek, Pawel
    Kieszko, Dariusz
    Brzozowska, Anna
    Kordzinska-Cisek, Izabela
    Mazurkiewicz, Maria
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2016, 8 (06) : 544 - 553
  • [47] End-to-end test for computed tomography-based high-dose-rate brachytherapy
    Krause, Fabian
    Risske, Franziska
    Bohn, Susann
    Delaperriere, Marc
    Dunst, Juergen
    Siebert, Frank-Andre
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (06) : 551 - 558
  • [48] ACR-ABS-ASTRO practice parameter for the performance of radionuclide-based high-dose-rate brachytherapy
    Harkenrider, Matthew M.
    Albuquerque, Kevin
    Brown, Derek
    Kamrava, Mitchell
    King, Martin
    Mourtada, Firas
    Orio, Peter
    Patel, Rakesh
    Price, Michael
    Rassiah, Prema
    Solanki, Abhishek A.
    Small, William, Jr.
    Schechter, Naomi R.
    BRACHYTHERAPY, 2021, 20 (06) : 1071 - 1082
  • [49] Dosimetry of local failure with single dose 19 Gy high-dose-rate brachytherapy for prostate cancer
    Armstrong, Shreya
    Tsang, Yatman
    Lowe, Gerry
    Tharmalingam, Hannah
    Alonzi, Roberto
    Ostler, Peter
    Hughes, Robert
    Hoskin, Peter
    RADIOTHERAPY AND ONCOLOGY, 2021, 157 : 93 - 98
  • [50] Dosimetric effect of external beam planning preceding combined high-dose-rate brachytherapy of the prostate
    Martin, Jarad M.
    Brett, Richard
    Blyth, Jemma
    Morrison, Stewart
    Bryant, Daniel
    Plank, Ashley
    Cheuk, Robyn
    Fay, Michael
    Dickie, Graeme
    Yaxley, John
    BRACHYTHERAPY, 2011, 10 (06) : 474 - 478