摘要
PurposeTo quantify the dose distribution effect of insufficient scattering conditions in keloid HDR brachytherapy with Freiburg fFlap (FF) applicator. Materials and MethodsA phantom composed of FF applicator, MatriXX and solid water slices was designed and scanned for treatment planning. Bolus with different thicknesses were covered to offer different scatter conditions. Planar dose distributions were measured by MatriXX. The maximum value (Max), average value (Avg) and & gamma; passing rate (3 mm/3%) were evaluated by the software MyQA Platform. ResultsThe maximum and average doses measured by MatriXX were lower than the calculated values. The difference increased as field size decreased. The Max value, found at 0.86 cm level in the two tube case, was -20.0%, and the avg value was -11.9%. All the & gamma; values were less than 95%. This difference gradually decreased with increasing bolus thickness and the & gamma; values were significantly improved. ConclusionMatriXX could be used for dose verification of HDR brachytherapy with an FF applicator. When the FF applicator was applied for keloid, insufficient scattering conditions would cause an insufficient target dose. This difference could be reduced by covering the bolus with different thicknesses on the applicator. The smaller the field, the thicker the bolus required.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Comparison of the IPSA and HIPO algorithms for interstitial tongue high-dose-rate brachytherapy
    Choi, Chang Heon
    Park, So-Yeon
    Park, Jong Min
    Wu, Hong-Gyun
    Kim, Jin-Ho
    Kim, Jung-in
    PLOS ONE, 2018, 13 (10):
  • [32] Improved rectal dosimetry with the use of SpaceOAR during high-dose-rate brachytherapy
    Wu, Susan Y.
    Boreta, Lauren
    Wu, Ashley
    Cheung, Joey P.
    Cunha, J. Adam M.
    Shinohara, Katsuto
    Chang, Albert J.
    BRACHYTHERAPY, 2018, 17 (02) : 259 - 264
  • [33] Dosimetric effect of interfractional needle displacement in prostate high-dose-rate brachytherapy
    Kovalchuk, Nataliya
    Furutani, Keith M.
    MacDonald, O. Kenneth
    Pisansky, Thomas M.
    BRACHYTHERAPY, 2012, 11 (02) : 111 - 118
  • [34] High-dose-rate and pulsed-dose-rate brachytherapy in palliative treatment of head and neck cancers
    Bartochowska, Anna
    Wierzbicka, Malgorzata
    Skowronek, Janusz
    Leszczynska, Malgorzata
    Szyfter, Witold
    BRACHYTHERAPY, 2012, 11 (02) : 137 - 143
  • [35] Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy
    Yogo, Katsunori
    Matsushita, Akihiro
    Tatsuno, Yuya
    Shimo, Takahiro
    Hirota, Seiko
    Nozawa, Marika
    Ozawa, Shuichi
    Ishiyama, Hiromichi
    Yasuda, Hiroshi
    Nagata, Yasushi
    Hayakawa, Kazushige
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [36] Validation of MRI to TRUS registration for high-dose-rate prostate brachytherapy
    Poulin, Eric
    Boudam, Karim
    Pinter, Csaba
    Kadoury, Samuel
    Lasso, Andras
    Fichtinger, Gabor
    Menard, Cynthia
    BRACHYTHERAPY, 2018, 17 (02) : 283 - 290
  • [37] Limitations of the TG-43 formalism for skin high-dose-rate brachytherapy dose calculations
    Granero, Domingo
    Perez-Calatayud, Jose
    Vijande, Javier
    Ballester, Facundo
    Rivard, Mark J.
    MEDICAL PHYSICS, 2014, 41 (02)
  • [38] High-Dose-Rate Brachytherapy as an Organ-Sparing Treatment for Early Penile Cancer
    Pohankova, Denisa
    Sirak, Igor
    Vosmik, Milan
    Kasaova, Linda
    Grepl, Jakub
    Paluska, Petr
    Holub, Lukas
    Spacek, Jiri
    Hodek, Miroslav
    Kopecek, Martin
    Petera, Jiri
    CANCERS, 2022, 14 (24)
  • [39] A surrogate urethra for real-time planning of high-dose-rate prostate brachytherapy
    Halperin, Heather
    Hilts, Michelle
    Crook, Juanita
    Batchelar, Deidre
    Tisseverasinghe, Steven
    Tetreault-LaFlamme, Audrey
    Bachand, Francois
    BRACHYTHERAPY, 2019, 18 (05) : 675 - 682
  • [40] Second salvage high-dose-rate brachytherapy for radiorecurrent prostate cancer
    Maenhout, Metha
    van Vulpen, Marco
    Moerland, Marinus
    Peters, Max
    Meijer, Richard
    van den Bosch, Maurice
    Nguyen, Paul
    Frank, Steven
    van Zyp, Jochem van der Voort
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2017, 9 (02) : 161 - 166