Evaluation of an Automated System for the Counting of Microbial Colonies

被引:2
作者
Heuser, Elisa [1 ]
Becker, Karsten [1 ]
Idelevich, Evgeny A. [1 ,2 ]
机构
[1] Univ Med Greifswald, Friedrich Loeffler Inst Med Microbiol, Greifswald, Germany
[2] Univ Hosp Munster, Inst Med Microbiol, Munster, Germany
来源
MICROBIOLOGY SPECTRUM | 2023年 / 11卷 / 04期
关键词
colony counting; automatic counts; bacteria; yeast; agar; ENUMERATION;
D O I
10.1128/spectrum.00673-23
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Counting of microbial colonies is a common technique employed in research and diagnostics. To simplify this tedious and time-consuming process, automated systems have been proposed. This study aimed to elucidate the reliability of automated colony counting. We evaluated a commercially available instrument (UVP ColonyDoc-It Imaging Station) in regard to its accuracy and potential time savings. Suspensions of Staphylococcus aureus, Escherichia coli, Pseudomonas aeruginosa, Klebsiella pneumoniae, Enterococcus faecium, and Candida albicans (n=20 each) were adjusted to achieve growth of approximately 1,000, 100, 10, and 1 colony per plate, respectively, after overnight incubation on different solid media. Compared with manual counting, each plate was automatically counted by the UVP ColonyDoc-It with and without visual adjustment on a computer display. For all bacterial species and concentrations automatically counted without visual correction, an overall mean difference from manual counts of 59.7%, a proportion of isolates with overestimation/underestimation of colony numbers of 29%/45%, respectively, and only a moderate relationship (R-2 = 0.77) with the manual counting were shown. Applying visual correction, the overall mean difference from manual counts was 1.8%, the proportion of isolates with overestimation/underestimation of colony numbers amounted to 2%/42%, respectively, and a strong relationship (R-2 = 0.99) with the manual counting was observed. The mean time needed for manual counting compared with automated counting without and with visual correction was 70s, 30s, and 104s, respectively, for bacterial colonies through all concentrations tested. Generally, similar performance regarding accuracy and counting time was observed with C. albicans. In conclusion, fully automatic counting showed low accuracy, especially for plates with very high or very low colony numbers. After visual correction of the automatically generated results, the concordance with manual counts was high; however, there was no advantage in reading time. Colony counting is a widely utilized technique in the field of microbiology. The accuracy and convenience of automated colony counters are essential for research and diagnostics. However, there is only sparse evidence on performance and usefulness of such instruments. This study examined the current state of reliability and practicality of the automated colony counting with an advanced modern system. For this, we thoroughly evaluated a commercially available instrument in terms of its accuracy and counting time required. Our findings indicate that fully automatic counting resulted in low accuracy, particularly for plates with very high or very low colony numbers. Visual correction of the automated results on a computer screen improved concordance with manual counts, but there was no benefit in counting time.
引用
收藏
页数:11
相关论文
共 24 条
  • [1] [Anonymous], 1999, Methods for determining bactericidal activity of antimicrobial agents: approved guideline M26-A
  • [2] A smart device application for the automated determination of E-coli colonies on agar plates
    Austerjost, Jonas
    Marquard, Daniel
    Raddatz, Lukas
    Geier, Dominik
    Becker, Thomas
    Scheper, Thomas
    Lindner, Patrick
    Beutel, Sascha
    [J]. ENGINEERING IN LIFE SCIENCES, 2017, 17 (08): : 959 - 966
  • [3] CRITICAL-EVALUATION OF AUTOMATIC BACTERIAL COLONY COUNTERS
    BRODSKY, MH
    CIEBIN, BW
    SCHIEMANN, DA
    [J]. JOURNAL OF FOOD PROTECTION, 1979, 42 (02) : 138 - 143
  • [4] Automated Counting of Bacterial Colony Forming Units on Agar Plates
    Brugger, Silvio D.
    Baumberger, Christian
    Jost, Marcel
    Jenni, Werner
    Brugger, Urs
    Muehlemann, Kathrin
    [J]. PLOS ONE, 2012, 7 (03):
  • [5] Automated counting of bacterial colonies by image analysis
    Chiang, Pei-Ju
    Tseng, Min-Jen
    He, Zong-Sian
    Li, Chia-Hsun
    [J]. JOURNAL OF MICROBIOLOGICAL METHODS, 2015, 108 : 74 - 82
  • [6] Low-Cost, High-Throughput, Automated Counting of Bacterial Colonies
    Clarke, Matthew L.
    Burton, Robert L.
    Hill, A. Nayo
    Litorja, Maritoni
    Nahm, Moon H.
    Hwang, Jeeseong
    [J]. CYTOMETRY PART A, 2010, 77A (08) : 790 - 797
  • [7] Clinical and Laboratory Standards Institute, 2018, METHODS DILUTION ANT, V10th ed
  • [8] ENUMERATING CHROMOGENIC AGAR PLATES USING THE COLOR QCOUNT AUTOMATED COLONY COUNTER
    Garry, Eileen
    Ouattara, Grace
    Williams, Patrick
    Pesta, Meredith
    [J]. JOURNAL OF RAPID METHODS AND AUTOMATION IN MICROBIOLOGY, 2009, 17 (01) : 46 - 54
  • [9] OpenCFU, a New Free and Open-Source Software to Count Cell Colonies and Other Circular Objects
    Geissmann, Quentin
    [J]. PLOS ONE, 2013, 8 (02):
  • [10] EVALUATION OF AN AUTOMATED COLONY COUNTER
    GOSS, WA
    MICHAUD, RN
    MCGRATH, MB
    [J]. APPLIED MICROBIOLOGY, 1974, 27 (01) : 264 - 267