Multilevel Venous Obstruction in Patients with Cardiac Implantable Electronic Devices

被引:1
作者
Czajkowski, Marek [1 ]
Polewczyk, Anna [2 ,3 ]
Jachec, Wojciech [4 ]
Kosior, Jaroslaw [5 ]
Nowosielecka, Dorota [6 ]
Tulecki, Lukasz [7 ]
Stefanczyk, Pawel [6 ]
Kutarski, Andrzej [8 ]
机构
[1] Med Univ Lublin, Dept Cardiac Surg, PL-20059 Lublin, Poland
[2] Jan Kochanowski Univ, Inst Med Sci, PL-25317 Kielce, Poland
[3] Swietokrzyskie Ctr Cardiol, Dept Cardiac Surg, PL-25736 Kielce, Poland
[4] Med Univ Silesia, Fac Med Sci Zabrze, Dept Cardiol 2, PL-41800 Zabrze, Poland
[5] Masovian Specialist Hosp Radom, Dept Cardiol, PL-26617 Radom, Poland
[6] Pope John Paul II Prov Hosp Zamosc, Dept Cardiol, PL-22400 Zamosc, Poland
[7] Pope John Paul II Prov Hosp Zamosc, Dept Cardiac Surg, PL-22400 Zamosc, Poland
[8] Med Univ Lublin, Dept Cardiol, Lublin, Poland
来源
MEDICINA-LITHUANIA | 2024年 / 60卷 / 02期
关键词
multilevel lead-related venous obstruction; risk factors; transvenous lead extraction; complications; long-term survival; EXPERT CONSENSUS STATEMENT; RISK-FACTORS; LEAD EXTRACTION; CARDIOVERTER-DEFIBRILLATOR; TERM THROMBOSIS; SUBCLAVIAN VEIN; PACEMAKER; OCCLUSION; PREVALENCE; REVISION;
D O I
10.3390/medicina60020336
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives: The nature of multilevel lead-related venous stenosis/occlusion (MLVSO) and its influence on transvenous lead extraction (TLE) as well as long-term survival remains poorly understood. Materials and Methods: A total of 3002 venograms obtained before a TLE were analyzed to identify the risk factors for MLVSO, as well as the procedure effectiveness and long-term survival. Results: An older patient age at the first system implantation (OR = 1.015; p < 0.001), the number of leads in the heart (OR = 1.556; p < 0.001), the placement of the coronary sinus (CS) lead (OR = 1.270; p = 0.027), leads on both sides of the chest (OR = 7.203; p < 0.001), and a previous device upgrade or downgrade with lead abandonment (OR = 2.298; p < 0.001) were the strongest predictors of MLVSO. Conclusions: The presence of MLVSO predisposes patients with cardiac implantable electronic devices (CIED) to the development of infectious complications. Patients with multiple narrowed veins are likely to undergo longer and more complex procedures with complications, and the rates of clinical and procedural success are lower in this group. Long-term survival after a TLE is similar in patients with MLVSO and those without venous obstruction. MLVSO probably better depicts the severity of global venous obstruction than the degree of vein narrowing at only one point.
引用
收藏
页数:15
相关论文
共 29 条
[21]   The incidence and risk factors for venous obstruction after implantation of transvenous pacing leads [J].
Oginosawa, Y ;
Abe, H ;
Nakashima, Y .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (11) :1605-1611
[22]   Venous thrombosis and stenosis after implantation of pacemakers and defibrillators [J].
Rozmus, G ;
Daubert, JP ;
Huang, DT ;
Rosero, S ;
Hall, B ;
Francis, C .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2005, 13 (01) :9-19
[23]  
Safi M, 2017, ROM J INTERN MED, V55, P139, DOI 10.1515/rjim-2017-0018
[24]   Laser lead extraction to facilitate cardiac implantable electronic device upgrade and revision in the presence of central venous obstruction [J].
Sohal, Manav ;
Williams, Steven ;
Akhtar, Majid ;
Shah, Amit ;
Chen, Zhong ;
Wright, Matthew ;
O'Neill, Mark ;
Patel, Nik ;
Hamid, Shoaib ;
Cooklin, Michael ;
Bucknall, Cliff ;
Bostock, Julian ;
Gill, Jaswinder ;
Rinaldi, Christopher Aldo .
EUROPACE, 2014, 16 (01) :81-87
[25]   Prevalence of central venous occlusion in patients with chronic defibrillator leads [J].
Sticherling, C ;
Chough, SP ;
Baker, RL ;
Wasmer, K ;
Oral, H ;
Tada, H ;
Horwood, L ;
Kim, MH ;
Pelosi, F ;
Michaud, GF ;
Strickberger, SA ;
Morady, F ;
Knight, BP .
AMERICAN HEART JOURNAL, 2001, 141 (05) :813-816
[26]   INCIDENCE OF VENOUS THROMBOSIS FOLLOWING LONG-TERM TRANSVENOUS PACING [J].
STONEY, WS ;
ADDLESTONE, RB ;
ALFORD, WC ;
BURRUS, GR ;
FRIST, RA ;
THOMAS, CS .
ANNALS OF THORACIC SURGERY, 1976, 22 (02) :166-170
[27]   Incidence and risk factors of early venous thrombosis associated with permanent pacemaker leads [J].
Van Rooden, CJ ;
Molhoek, SG ;
Rosendaal, FR ;
Schalij, MJ ;
Meinders, AE ;
Huisman, MV .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (11) :1258-1262
[28]   Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications, and Patient Management [J].
Wilkoff, Bruce L. ;
Love, Charles J. ;
Byrd, Charles L. ;
Bongiorni, Maria Grazia ;
Carrillo, Roger G. ;
Crossley, George H., III ;
Epstein, Laurence M. ;
Friedman, Richard A. ;
Kennergren, Charles E. H. ;
Mitkowski, Przemyslaw ;
Schaerf, Raymond H. M. ;
Wazni, Oussama M. .
HEART RHYTHM, 2009, 6 (07) :1085-1104
[29]   Assessment of the subclavian vein in patients with transvenous pacemaker leads [J].
Zuber, M ;
Huber, P ;
Fricker, U ;
Buser, P ;
Jäger, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (12) :2621-2630