Mechanical Resistance of a 2.9-mm-Diameter Dental Implant With a Morse-Taper Implant-Abutment Connection

被引:1
作者
Alberti, Alice [1 ,2 ]
Corbella, Stefano [1 ,2 ,3 ]
Francetti, Luca [1 ,2 ]
机构
[1] Univ Milan, Dept Biomed Surg & Dent Sci, Milan, Italy
[2] IRCCS Ist Ortoped Galeazzi, Milan, Italy
[3] IM Sechenov First Moscow State Med Univ, Inst Dent, Dept Oral Surg, Moscow, Russia
关键词
dental implants; finite element analysis; fatigue testing; Morse-taper dental implant-abutment connection; narrow diameter implants; DIAMETER; REHABILITATIONS; COMPLICATIONS; SURVIVAL; RATES;
D O I
10.1563/aaid-joi-D-21-00258
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Among the complications that can occur at dental implants, the fracture of any implant component is a relatively infrequent but clinically relevant problem. Because of their mechanical characteristics, small diameter implants are at higher risk of such complication. The aim of this laboratory and finite element method (FEM) study was to compare the mechanical behavior of a 2.9-and 3.3-mm-diameter implant with a conical connection under standard static and dynamic conditions, following the International Organization for Standardization (ISO) 14801:2017. Finite element analysis was performed to compare the stress distribution on the tested implant systems under a 300-N, 30 & DEG; inclined force. Static tests were performed with a load cell of 2 kN; the force was applied on the experimental samples at 30 & DEG; with respect to the implant-abutment axis, with an arm of 5.5 mm. Fatigue tests were performed with decreasing loads, at 2-Hz frequency, until 3 specimens survived without any damage after 2 million cycles. The emergence profile of the abutment resulted the most stressed area in finite element analysis, with a maximum stress of 5829 and 5480 MPa for 2.9-and 3.3-mm-diameter implant complex, respectively. The mean maximum load resulted in 360 N for 2.9-mm-diameter and 370 N for 3.3-mm-diameter implants. The fatigue limit was recorded to be 220 and 240 N, respectively. Despite the more favorable results of 3.3-mm-diameter implants, the difference between the tested implants could be considered clinically negligible. This is probably due to the conical design of the implant-abutment connection, which has been reported to present low stress values in the implant neck region, thus increasing the fracture resistance.
引用
收藏
页码:323 / 329
页数:7
相关论文
共 30 条
[1]   The impact of loads on standard diameter, small diameter and mini implants: a comparative laboratory study [J].
Allum, Simon Rupert ;
Tomlinson, Rachel Anne ;
Joshi, Rajendra .
CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (06) :553-559
[2]  
Balshi T J, 1996, Int J Oral Maxillofac Implants, V11, P660
[3]  
Chee W, 1999, INT J ORAL MAX IMPL, V14, P137
[4]   Survival rates and prosthetic complications of implant fixed complete dental prostheses: An up to 5-year retrospective study [J].
Chochlidakis, Konstantinos ;
Einarsdottir, Erna ;
Tsigarida, Alexandra ;
Papaspyridakos, Panos ;
Romeo, Davide ;
Barmak, Abdul Basir ;
Ercoli, Carlo .
JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (05) :539-546
[5]   Factors influencing the fracture of dental implants [J].
Chrcanovic, Bruno Ramos ;
Kisch, Jeno ;
Albrektsson, Tomas ;
Wennerberg, Ann .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (01) :58-67
[6]   Fatigue lifetime prediction of a reduced-diameter dental implant system: Numerical and experimental study [J].
Duan, Yuanyuan ;
Gonzalez, Jorge A. ;
Kulkarni, Pratim A. ;
Nagy, William W. ;
Griggs, Jason A. .
DENTAL MATERIALS, 2018, 34 (09) :1299-1309
[7]   Performance evaluation of survival regression models in analysing Swedish dental implant complication data with frailty [J].
Fagbamigbe, Adeniyi Francis ;
Karlsson, Karolina ;
Derks, Jan ;
Petzold, Max .
PLOS ONE, 2021, 16 (01)
[8]   Medium- and Long-Term Complications in Full-Arch Rehabilitations Supported by Upright and Tilted Implants [J].
Francetti, Luca ;
Corbella, Stefano ;
Taschieri, Silvio ;
Cavalli, Nicolo ;
Del Fabbro, Massimo .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2015, 17 (04) :758-764
[9]   Expanded Clinical Applications of Narrow-Diameter Implants for Permanent Use [J].
Froum, Stuart J. ;
Natour, Mazen ;
Cho, Sang-Choon ;
Yu, Paul Y. C. ;
Leung, Martin .
INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2020, 40 (04) :529-538
[10]   OCCLUSAL FORCES DURING CHEWING AND SWALLOWING AS MEASURED BY SOUND-TRANSMISSION [J].
GIBBS, CH ;
MAHAN, PE ;
LUNDEEN, HC ;
BREHNAN, K ;
WALSH, EK ;
HOLBROOK, WB .
JOURNAL OF PROSTHETIC DENTISTRY, 1981, 46 (04) :443-449