Development of Wall-Coated Open Tubular Columns and Their Application to Nano Liquid Chromatography Coupled to Tandem Mass Spectrometry

被引:2
|
作者
Santos, Natalia G. P. [1 ]
Medina, Deyber A. V. [1 ]
Lancas, Fernando M. [1 ]
机构
[1] Univ Sao Paulo, Sao Carlos Inst Chem, BR-13566590 Sao Carlos, Brazil
来源
MOLECULES | 2023年 / 28卷 / 13期
基金
巴西圣保罗研究基金会;
关键词
column development; open tubular columns; nano liquid chromatography; mass spectrometry; wall coated open tubular column (WCOT); CHEMICALLY BONDED OCTADECYLSILANE; STATIONARY PHASES; CAPILLARY; OPTIMIZATION; SWOLLEN;
D O I
10.3390/molecules28135103
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
This work presents a study on the application of wall open tubular column (WCOT) in liquid chromatography coupled with tandem mass spectrometry. Each process step reports the column preparation method in detail, subdivided into column pretreatment, silanization, stationary phase coating, and immobilization. Then, an evaluation of the parameters that can affect the efficiency of these columns was made. Atrazine, clomazone, and metolachlor were used as probes during this step. Factors such as stationary phase composition, length, internal diameter, stationary phase mass employed, and injection volume were investigated. In addition, with the help of Knox and Poppe graphs, the columns' performance was evaluated to determine the optimal flow rate and the speed-efficiency relationship, respectively. Based on the results, the best configurations for the WCOT column application to the LC system were defined: length-8 m; inner diameter-25 & mu;m; mass of OV-210-2.5% m/v; and, injection volume-100 nL. Finally, the optimized WCOT column developed in this work was coupled with a commercially-packed trapping column in the nano liquid chromatography system (nanoLC). In this configuration, more significant results were obtained regarding separation resolution, with Rs = 5.9 achieved for the most retained pair of analytes (clomazone and metolachlor).
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Gas chromatography on wall-coated open-tubular columns with ionic liquid stationary phases
    Poole, Colin F.
    Lenca, Nicole
    JOURNAL OF CHROMATOGRAPHY A, 2014, 1357 : 87 - 109
  • [2] Separation characteristics of wall-coated open-tubular columns for gas chromatography
    Poole, Colin F.
    Poole, Salwa K.
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1184 (1-2) : 254 - 280
  • [3] Porous layer open tubular nano liquid chromatography directly coupled to electron ionization mass spectrometry
    Medina, Deyber Arley Vargas
    Burato, Juliana Soares da Silva
    Borsatto, Joao Victor Basolli
    Lancas, Fernando Mauro
    JOURNAL OF CHROMATOGRAPHY A, 2022, 1674
  • [4] Selection of calibration compounds for selectivity evaluation of wall-coated, open-tubular columns for gas chromatography by the solvation parameter model
    Poole, Colin F.
    JOURNAL OF CHROMATOGRAPHY A, 2020, 1629 (1629)
  • [5] Wall-coated open-tubular column chromatography on an organo-clay stationary phase
    Bruno, TJ
    Lewansowska, A
    Tsvetkov, F
    Miller, KE
    Hanley, HJM
    JOURNAL OF CHROMATOGRAPHY A, 2002, 973 (1-2) : 143 - 149
  • [6] Determination of sulfonamides in beeswax by liquid chromatography coupled to tandem mass spectrometry
    Mitrowska, Kamila
    Antczak, Maja
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2015, 1006 : 179 - 186
  • [7] Preparation and Application of Porous Layer Open Tubular Capillary Columns with Narrow Bore in Liquid Chromatography
    Li Ruo-Nan
    Wang Ya-Nan
    Peng Man-Hua
    Wang Xia-Yan
    Guo Guang-Sheng
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2017, 45 (12) : 1865 - 1872
  • [8] An overview of open tubular liquid chromatography with a focus on the coupling with mass spectrometry for the analysis of small molecules
    Vargas Medina, Deyber Arley
    Pereira dos Santos, Natalia Gabrielly
    da Silva Burato, Juliana Soares
    Basolli Borsatto, Joao Victor
    Lancas, Fernando Mauro
    JOURNAL OF CHROMATOGRAPHY A, 2021, 1641 (1641)
  • [9] Determination of Abamectin in Soybean Roots by Liquid Chromatography Coupled to Tandem Mass Spectrometry
    de Oliveira, Eliezer
    da Silva, Bianca Ferreira
    Bottoli, Carla Beatriz Grespan
    BRAZILIAN JOURNAL OF ANALYTICAL CHEMISTRY, 2024, 11 (44): : 76 - 84
  • [10] Miniaturization of liquid chromatography coupled to mass spectrometry 1. Current trends on miniaturized LC columns
    Mejia-Carmona, Karen
    da Silva Burato, Juliana Soares
    Basolli Borsatto, Joao Victor
    de Toffoli, Ana Lucia
    Lancas, Fernando Mauro
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2020, 122