Deep learning reconstruction for brain diffusion-weighted imaging: efficacy for image quality improvement, apparent diffusion coefficient assessment, and intravoxel incoherent motion evaluation in in vitro and in vivo studies

被引:5
作者
Hanamatsu, Satomu [1 ]
Murayama, Kazuhiro [1 ]
Ohno, Yoshiharu [1 ,2 ]
Yamamoto, Kaori [3 ]
Yui, Masao [3 ]
Toyama, Hiroshi [1 ]
机构
[1] Fujita Hlth Univ, Sch Med, Dept Radiol, Toyoake, Aichi, Japan
[2] Fujita Hlth Univ, Sch Med, Joint Res Lab Adv Med Imaging, Toyoake, Aichi, Japan
[3] Canon Med Syst Corp, Otawara, Japan
关键词
Brain; magnetic resonance imaging; diffusion; intravoxel incoherent motion; deep learning reconstruction; HIGH B-VALUE; NOISE-REDUCTION; PERFUSION; MRI;
D O I
10.4274/dir.2023.232149
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE Deep learning reconstruction (DLR) to improve imaging quality has already been introduced, but no studies have evaluated the effect of DLR on diffusion-weighted imaging (DWI) or intravoxel incoherent motion (IVIM) in in vitro or in vivo studies. The purpose of this study was to determine the effect of DLR for magnetic resonance imaging (MRI) in terms of image quality improvement, apparent diffusion coefficient (ADC) assessment, and IVIM index evaluation on DWI through in vitro and in vivo studies. METHODS For the in vitro study, a phantom recommended by the Quantitative Imaging Biomarkers Alliance was scanned and reconstructed with and without DLR, and 15 patients with brain tumors with normal-appearing gray and white matter examined using IVIM and reconstructed with and without DLR were included in the in vivo study. The ADCs of all phantoms for DWI with and without DLR, as well as the coefficient of variation percentage (CV%), and ADCs and IVIM indexes for each participant, were evaluated based on DWI with and without DLR by means of region-of-interest measurements. For the in vitro study, using the mean ADCs for all phantoms, a t-test was adopted to compare DWI with and without DLR. For the in vivo study, a Wilcoxon signed-rank test was used to compare the CV% between the two types of DWI. In addition, the Wilcoxon signed-rank test was used to compare the ADC, true diffusion coefficient (D), pseudodiffusion coefficient (D*), and percentage of water molecules in micro perfusion within 1 voxel (f) with and without DLR; the limits of agreement of each parameter were determined through a Bland-Altman analysis. RESULTS The in vitro study identified no significant differences between the ADC values for DWI with and without DLR (P > 0.05), and the CV% was significantly different for DWI with and without DLR (P < 0.05) when b values >= 250 s/mm(2) were used. The in vivo study revealed that D* and f with and without DLR were significantly different (P < 0.001). The limits of agreement of the ADC, D, and D* values for DWI with and without DLR were determined as 0.00 +/- 0.51 x 10(-3), 0.00 +/- 0.06 x 10(-3), and 1.13 +/- 4.04 x 10(-3) mm(2)/s, respectively. The limits of agreement of the f values for DWI with and without DLR were determined as -0.01 +/- 0.07. CONCLUSION Deep learning reconstruction for MRI has the potential to significantly improve DWI quality at higher b values. It has some effect on D* and f values in the IVIM index evaluation, but ADC and D values are less affected by DLR.
引用
收藏
页码:664 / 673
页数:10
相关论文
共 41 条
[1]   Acquisition time reduction of diffusion-weighted liver imaging using deep learning image reconstruction [J].
Afat, Saif ;
Herrmann, Judith ;
Almansour, Haidara ;
Benkert, Thomas ;
Weiland, Elisabeth ;
Hoelldobler, Thomas ;
Nikolaou, Konstantin ;
Gassenmaier, Sebastian .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2023, 104 (04) :178-184
[2]   Clinical feasibility of accelerated diffusion weighted imaging of the abdomen with deep learning reconstruction: Comparison with conventional diffusion weighted imaging [J].
Bae, Sung Hwan ;
Hwang, Jiyoung ;
Hong, Seong Sook ;
Lee, Eun Ji ;
Jeong, Jewon ;
Benkert, Thomas ;
Sung, JaeKon ;
Arberet, Simon .
EUROPEAN JOURNAL OF RADIOLOGY, 2022, 154
[3]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Intravoxel Incoherent Motion Magnetic Resonance Imaging in Skeletal Muscle: Review and Future Directions [J].
Englund, Erin K. ;
Reiter, David A. ;
Shahidi, Bahar ;
Sigmund, Eric E. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2022, 55 (04) :988-1012
[6]   Measuring Perfusion Intravoxel Incoherent Motion MR Imaging [J].
Federau, Christian .
MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2021, 29 (02) :233-242
[8]   Measuring brain perfusion with intravoxel incoherent motion (IVIM): Initial clinical experience [J].
Federau, Christian ;
O'Brien, Kieran ;
Meuli, Reto ;
Hagmann, Patric ;
Maeder, Philippe .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 39 (03) :624-632
[9]   Deep learning-based reconstruction of in vivo pelvis conductivity with a 3D patch-based convolutional neural network trained on simulated MR data [J].
Gavazzi, Soraya ;
den Berg, Cornelis A. T. van ;
Savenije, Mark H. F. ;
Kok, H. Petra ;
Boer, Peter de ;
Stalpers, Lukas J. A. ;
Lagendijk, Jan J. W. ;
Crezee, Hans ;
Lier, Astrid L. H. M. W. van .
MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (05) :2772-2787
[10]   Deep Learning Reconstruction at CT: Phantom Study of the Image Characteristics [J].
Higaki, Toru ;
Nakamura, Yuko ;
Zhou, Jian ;
Yu, Zhou ;
Nemoto, Takuya ;
Tatsugami, Fuminari ;
Awai, Kazuo .
ACADEMIC RADIOLOGY, 2020, 27 (01) :82-87