Calibration transfer for near-infrared (NIR) spectroscopy based on local preserving projection

被引:2
|
作者
Bin, Jun [1 ,2 ]
Yan, Xiufang [1 ]
Liu, Renxiang [1 ,2 ]
Kang, Chao [3 ]
Chen, Yi [4 ]
机构
[1] Guizhou Univ, Coll Tobacco Sci, Guiyang 550025, Peoples R China
[2] Guizhou Key Lab Tobacco Qual Res, Guiyang 550025, Peoples R China
[3] Guizhou Univ, Sch Chem & Chem Engn, Guiyang 550025, Peoples R China
[4] Yunnan Acad Tobacco Agr Sci, Kunming 650021, Peoples R China
基金
芬兰科学院;
关键词
NIR spectroscopy; Calibration transfer; Local preserving projection (LPP); Manifold learning; Linear feature extraction; MODEL TRANSFER; STANDARDIZATION; REGRESSION; SPECTRA; COMPONENTS;
D O I
10.1016/j.vibspec.2023.103538
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Calibration transfer is a mathematical or statistical technique to solve the universality of near-infrared (NIR) spectroscopy, which is particularly important in practical applications. Therefore, a calibration transfer method based on local preserving projection (CTLPP) to correct spectral differences was proposed in this paper. Local preserving projection (LPP) is a linear feature extraction algorithm, which is a linear approximation of Laplacian Eigenmap manifold learning. It can not only overcome the shortcomings of linear feature extraction algorithms which are difficult to maintain the nonlinear manifold of the original data, but also solve the problem that nonlinear manifold learning methods are difficult to directly map new samples, which is conducive to building the conversion relationship between the source spectra and the target spectra. The transfer performance of the proposed method was evaluated in three datasets, and compared with five approaches, namely, piecewise direct standardization (PDS), calibration transfer based on independent component analysis (CTICA), spectral space transformation (SST), principal components canonical correlation analysis (PC-CCA) and calibration transfer based on neighborhood preserving embedding (CTNPE). Experimental results indicated that this proposed method can successfully correct the detection spectra among different apparatus. In addition, it can obtain relatively low prediction root mean square error (RMSEP). Therefore, the comprehensive research conducted in this work demonstrated that the proposed method is a promising calibration transfer method in NIR applications, which can provide a reference when sharing the NIR spectral models with more and more instruments.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Calibration transfer of near-infrared spectra based on compression of wavelet coefficients
    Yoon, JG
    Lee, BW
    Han, CH
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2002, 64 (01) : 1 - 14
  • [22] Robust Calibration Design in the Pharmaceutical Quantitative Measurements with Near-Infrared (NIR) Spectroscopy: Avoiding the Chemometric Pitfalls
    Xiang, Dong
    Berry, Joseph
    Buntz, Susan
    Gargiulo, Paul
    Cheney, James
    Joshi, Yatindra
    Wabuyele, Busolo
    Wu, Huiquan
    Hamed, Mazen
    Hussain, Ajaz S.
    Khan, Mansoor A.
    JOURNAL OF PHARMACEUTICAL SCIENCES, 2009, 98 (03) : 1155 - 1166
  • [23] What!?: Another article about near-infrared (NIR) spectroscopy!?
    Psotka, J
    CEREAL FOODS WORLD, 1999, 44 (06) : 428 - 429
  • [24] THE SPREAD OF THE CALIBRATION SET IN NEAR-INFRARED REFLECTANCE SPECTROSCOPY
    WETHERILL, GZ
    MURRAY, I
    JOURNAL OF AGRICULTURAL SCIENCE, 1987, 109 : 539 - 544
  • [25] Calibration transfer of near-infrared spectroscopy by canonical correlation analysis coupled with wavelet transform
    Bin, Jun
    Li, Xin
    Fan, Wei
    Zhou, Ji-heng
    Wang, Cheng-wei
    ANALYST, 2017, 142 (12) : 2229 - 2238
  • [26] NEAR-INFRARED (NIR) OF CARBOHYDRATES
    MCDONALDLEWIS, C
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1994, 208 : 1 - CARB
  • [27] Calibration of a visible and near-infrared portable transfer radiometer
    Biggar, SF
    METROLOGIA, 1998, 35 (04) : 701 - 706
  • [28] CALIBRATION TRANSFER AND MEASUREMENT STABILITY OF NEAR-INFRARED SPECTROMETERS
    WANG, YD
    KOWALSKI, BR
    APPLIED SPECTROSCOPY, 1992, 46 (05) : 764 - 771
  • [29] CALIBRATION TRANSFER BETWEEN NEAR-INFRARED REFLECTANCE SPECTROPHOTOMETERS
    SHENK, JS
    WESTERHAUS, MO
    TEMPLETON, WC
    CROP SCIENCE, 1985, 25 (01) : 159 - 161
  • [30] Operationalizing Maintenance of Calibration Models Based on Near-Infrared Spectroscopy by Knowledge Integration
    Miyano, Takuya
    Nakagawa, Hiroshi
    Watanabe, Tomoyuki
    Minami, Hidemi
    Sugiyama, Hirokazu
    JOURNAL OF PHARMACEUTICAL INNOVATION, 2015, 10 (04) : 287 - 301