An integrated and fast imaging quality assurance phantom for a 0.35 T magnetic resonance imaging linear accelerator

被引:1
|
作者
Sohn, James J. [1 ]
Lim, Sara [1 ]
Das, Indra J. [1 ]
Yadav, Poonam [1 ]
机构
[1] Northwestern Univ, Northwestern Mem Hosp, Dept Radiat Oncol, Feinberg Sch Med, Chicago, IL 60611 USA
来源
PHYSICS & IMAGING IN RADIATION ONCOLOGY | 2023年 / 27卷
关键词
Magnetic resonance imaging linear accelerator; Imaging; Magnetic resonance imaging guided radiation; therapy; MRIdian; QA; Magphan; ACR alternatives; MR-LINAC; RADIATION; IMPLEMENTATION;
D O I
10.1016/j.phro.2023.100462
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Periodic imaging quality assurance (QA) of magnetic resonance imaging linear accelerator (MRL) is critical. The feasibility of a new MRL imaging phantom used for QA in the low field was evaluated with auto-mated image analysis of various parameters for accuracy and reproducibility. Methods and materials: The new MRL imaging phantom was scanned across every 30 degrees of the gantry, having the on/off state of the linac in a low-field MRL system using three magnetic resonance imaging sequences: true fast imaging with steady-state precession (TrueFISP), T1 weighted (T1W), and T2 weighted (T2W). The DICOM files were used to calculate the imaging parameters: geometric distortion, uniformity, resolution, signal-to-noise ratio (SNR), and laser alignment. The point spread function (PSF) and edge spread function (ESF) were also calculated for resolution analysis. Results: The phantom data showed a small standard deviation -and high consistency for each imaging parameter. The highest variability in data was observed with the true fast imaging sequence at the calibration angle, which was expected because of low resolution and short scan time (25 sec). The mean magnitude of the largest distortion measured within 200 mm diameter with TrueFISP was 0.31 & PLUSMN; 0.05 mm. The PSF, ESF, signal uni-formity, and SNR measurements remained consistent. Laser alignment traditional offsets and angular deviation remained consistent. Conclusions: The new MRL imaging phantom is reliable, reproducible, time effective, and easy to use for a 0.35 T MRL system. The results promise a more streamlined, time-saving, and error-free QA process for low-field MRL adapted in our clinical setting.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Accurate, repeatable, and geometrically precise diffusion-weighted imaging on a 0.35 T magnetic resonance imaging-guided linear accelerator
    Weygand, Joseph
    Armstrong, Tess
    Bryant, John Michael
    Andreozzi, Jacqueline M.
    Oraiqat, Ibrahim M.
    Nichols, Steven
    Liveringhouse, Casey L.
    Latifi, Kujtim
    Yamoah, Kosj
    Costello, James R.
    Frakes, Jessica M.
    Moros, Eduardo G.
    El Naqa, Issam M.
    Naghavi, Arash O.
    Rosenberg, Stephen A.
    Redler, Gage
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 28
  • [2] Quantitative analysis of diffusion weighted imaging in rectal cancer during radiotherapy using a magnetic resonance imaging integrated linear accelerator
    Ingle, Manasi
    Blackledge, Matthew
    White, Ingrid
    Wetscherek, Andreas
    Lalondrelle, Susan
    Hafeez, Shaista
    Bhide, Shreerang
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 23 : 32 - 37
  • [3] A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy
    Alzahrani, Meshal
    Broadbent, David A.
    Teh, Irvin
    Al-Qaisieh, Bashar
    Johnstone, Emily
    Speight, Richard
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2025, 33
  • [4] A quality assurance phantom for electronic portal imaging devices
    Das, Indra J.
    Cao, Minsong
    Cheng, Chee-Wai
    Misic, Vladimir
    Scheuring, Klaus
    Schuele, Edmund
    Johnstone, Peter A. S.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 391 - 403
  • [5] Quality assurance of linear accelerator: a comprehensive system using electronic portal imaging device
    Niyas, P.
    Abdullah, K. K.
    Noufal, M. P.
    Vysakh, R.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (02) : 138 - 149
  • [6] Planning feasibility of extremely hypofractionated prostate radiotherapy on a 1.5 T magnetic resonance imaging guided linear accelerator
    den Hartogh, Mariska D.
    de Boer, Hans C. J.
    De Groot-van Breugel, Eline N.
    van Zyp, Jochem R. N. van der Voort
    Hes, Jochem
    van der Heide, Uulke A.
    Pos, Floris
    Haustermans, Karin
    Depuydt, Tom
    Smeenk, Robert Jan
    Kunze-Busch, Martina
    Raaymakers, Bas W.
    Kerkmeijer, Linda G. W.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2019, 11 : 16 - 20
  • [7] The osteoarthritis initiative (OAI) magnetic resonance imaging quality assurance methods and results
    Schneider, E.
    NessAiver, M.
    White, D.
    Purdy, D.
    Martin, L.
    Fanella, L.
    Davis, D.
    Vignone, M.
    Wu, G.
    Gullapalli, R.
    OSTEOARTHRITIS AND CARTILAGE, 2008, 16 (09) : 994 - 1004
  • [8] Problems and Promises of Introducing the Magnetic Resonance Imaging Linear Accelerator Into Routine Care: The Case of Prostate Cancer
    Hehakaya, Charisma
    Van der Voort van Zyp, Jochem R.
    Lagendijk, Jan J. W.
    Grobbee, Diederick E.
    Verkooijen, Helena M.
    Moors, Ellen H. M.
    FRONTIERS IN ONCOLOGY, 2020, 10
  • [9] Reproducible phantom for quality assurance in abdominal MRI focussing kidney imaging
    Wolf, Marcos
    Kommer, Stefan
    Fembek, Sebastian
    Droeszler, Uwe
    Koerner, Tito
    Berg, Andreas
    Schmid, Albrecht Ingo
    Moser, Ewald
    Meyerspeer, Martin
    FRONTIERS IN PHYSICS, 2022, 10
  • [10] Validation of echo planar imaging based diffusion-weighted magnetic resonance imaging on a 0.35 T MR-Linac
    Wallimann, Philipp
    Piccirelli, Marco
    Nowakowska, Sylwia
    Armstrong, Tess
    Mayinger, Michael
    Boss, Andreas
    Bink, Andrea
    Guckenberger, Matthias
    Tanadini-Lang, Stephanie
    Andratschke, Nicolaus
    Pouymayou, Bertrand
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 30