Metallosis After Oxinium Total Knee Arthroplasty in a Patient With Rheumatoid Arthritis: A Case Report

被引:5
作者
Canizares, Stalin [1 ]
Barriga, Gabriela Carolina Carrera [2 ]
Jarrin, Fabiana Valencia [3 ]
Freire, Carlos Daniel Poveda [2 ]
机构
[1] Univ San Francisco Quito, Gen Surg, Quito, Ecuador
[2] Univ San Francisco Quito, Trauma & Orthopaed, Quito, Ecuador
[3] Univ San Fracisco Quito, Trauma & Orthopaed, Quito, Ecuador
关键词
case report; rheumatoid arthritis; oxinium; total knee arthroplasty; metallosis;
D O I
10.7759/cureus.34541
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Metallosis is a late uncommon complication of knee arthroplasties due to prosthetic loosening or component displacement. Oxinium prosthesis used to have components that attempted and proved to decrease prosthetic wear and consequent metallosis in the past. However, new studies showed that a combination of a shallow anterior tab snap-fit locking mechanism and thin dovetail lips make it susceptible to polyethylene dislocation and prosthesis loosening. The following case report show metallosis development in a 69-year-old female patient with a 20-year history of stage IV left gonarthrosis (Kellgren and Lawrence classification) who underwent a total knee arthroplasty (TKA) with a high-flex PS Genesis II prosthesis (Smith & Nephew, Hertfordshire, UK). We discuss the role of the material and her rheumatoid arthritis background in orthopedic mechanical failure. It is crucial that designers focus on the improvement of locking mechanisms and polyethylene properties.
引用
收藏
页数:9
相关论文
共 10 条
[1]   Spontaneous, recurrent dislodgment of the polyethylene tibial insert after total knee arthroplasty - A case report [J].
Anderson, John A. ;
MacDessi, Samuel J. ;
Valle, Alejandro Gonzalez Della .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (02) :404-407
[2]   Polyethylene liner dislocation of fixed-bearing medial oxinium unicompartmental arthroplasty with severe metallosis [J].
Greco, Nicholas ;
Berend, Keith .
KNEE, 2018, 25 (02) :341-345
[3]  
La Verde L, 2017, INT J SURG CASE REP, V30, P9, DOI 10.1016/j.ijscr.2016.11.023
[4]   Spontaneous Late Dissociation of the Tibial Insert After High-Flex Posterior-Stabilized Genesis II Total Knee Arthroplasty [J].
Lee, Dae-Hee ;
Lee, Tze Gin ;
Park, Sung-Joon ;
Han, Seung-Beom .
JOURNAL OF ARTHROPLASTY, 2013, 28 (02) :374.e13-374.e16
[5]   Ultrasonographic and radiographic findings of polyethylene component displacement with severe metallosis and metal-induced synovitis following total knee arthroplasty [J].
Mallon, Shane ;
Bussis, Kyle ;
Beswick, Zachary ;
North, W. Trevor ;
Soliman, Steven B. .
KNEE, 2019, 26 (04) :941-950
[6]   Total knee arthroplasty with an oxidised zirconium femoral component: a 5-year follow-up study [J].
Park, Derek H. ;
Leong, Julian ;
Palmer, Stephen J. .
JOURNAL OF ORTHOPAEDIC SURGERY, 2014, 22 (01) :75-79
[7]   Metallosis following a clip breakage in a total knee arthroplasty implant: A case report [J].
Saad, Ahmed Isam ;
Shahban, Shafiq Arif ;
Fernandes, Richard .
WORLD JOURNAL OF ORTHOPEDICS, 2018, 9 (12) :300-303
[8]  
Salem Khaled Hamed, 2020, J Clin Orthop Trauma, V11, P165, DOI 10.1016/j.jcot.2019.04.010
[9]  
Vecchini Eugenio, 2019, Acta Biomed, V90, P198, DOI 10.23750/abm.v90i1-S.8082
[10]  
Vivegananthan Bhairavi, 2014, BMJ Case Rep, V2014, DOI 10.1136/bcr-2013-202801