Commissioning and validation of a Monte Carlo algorithm for spine stereotactic radiosurgery

被引:0
作者
Knill, Cory [1 ,2 ]
Sandhu, Raminder [1 ]
Loughery, Brian [1 ]
Lin, Lifeng [1 ]
Halford, Robert [1 ]
Drake, Doug [1 ]
Snyder, Michael [1 ]
机构
[1] Corewell Hlth William Beaumont Univ Hosp, Dept Radiat Oncol, Royal Oak, MI USA
[2] Dearborn Hosp, Corewell Hlth, Dept Radiat Oncol, 18101 Oakwood, Dearborn, MI 48124 USA
关键词
brainlab spine SRS; Monte Carlo; versa HD; BODY RADIATION-THERAPY; BEAM DOSE CALCULATION; MEGAVOLTAGE PHOTON; RADIOTHERAPY; CONFORMITY; TRANSPORT; OUTCOMES; INDEX;
D O I
10.1002/acm2.14092
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeA 6FFF Monte Carlo (MC) dose calculation algorithm was commissioned for spine stereotactic radiosurgery (SRS). Model generation, validation, and ensuing model tuning are presented. MethodsThe model was generated using in-air and in-water commissioning measurements of field sizes between 10 and 400 mm(2). Commissioning measurements were compared to simulated water tank MC calculations to validate output factors, percent depth doses (PDDs), profile sizes and penumbras. Previously treated Spine SRS patients were re-optimized with the MC model to achieve clinically acceptable plans. Resulting plans were calculated on the StereoPHAN phantom and subsequently delivered to the microDiamond and SRSMapcheck to verify calculated dose accuracy. Model tuning was performed by adjusting the model's light field offset (LO) distance between physical and radiological positions of the MLCs, to improve field size and StereoPHAN calculation accuracy. Following tuning, plans were generated and delivered to an anthropomorphic 3D-printed spine phantom featuring realistic bone anatomy, to validate heterogeneity corrections. Finally, plans were validated using polymer gel (VIPAR based formulation) measurements. ResultsCompared to open field measurements, MC calculated output factors and PDDs were within 2%, profile penumbra widths were within 1 mm, and field sizes were within 0.5 mm. Calculated point dose measurements in the StereoPHAN were within 0.26% & PLUSMN; 0.93% and -0.10% & PLUSMN; 1.37% for targets and spinal canals, respectively. Average SRSMapcheck per-plan pass rates using a 2%/2 mm/10% threshold relative gamma analysis was 99.1% & PLUSMN; 0.89%. Adjusting LOs improved open field and patient-specific dosimetric agreement. Anthropomorphic phantom measurements were within -1.29% & PLUSMN; 1.00% and 0.27% & PLUSMN; 1.36% of MC calculated for the vertebral body (target) and spinal canal, respectively. VIPAR gel measurements confirmed good dosimetric agreement near the target-spine junction. ConclusionValidation of a MC algorithm for simple fields and complex SRS spine deliveries in homogeneous and heterogeneous phantoms has been performed. The MC algorithm has been released for clinical use.
引用
收藏
页数:11
相关论文
共 39 条
[1]   Surgical resection of epidural disease improves local control following postoperative spine stereotactic body radiotherapy [J].
Al-Omair, Ameen ;
Masucci, Laura ;
Masson-Cote, Laurence ;
Campbell, Mikki ;
Atenafu, Eshetu G. ;
Parent, Amy ;
Letourneau, Daniel ;
Yu, Eugene ;
Rampersaud, Raja ;
Massicotte, Eric ;
Lewis, Stephen ;
Yee, Albert ;
Thibault, Isabelle ;
Fehlings, Michael G. ;
Sahgal, Arjun .
NEURO-ONCOLOGY, 2013, 15 (10) :1413-1419
[2]   Stereotactic body radiation therapy: The report of AAPM Task Group 101 [J].
Benedict, Stanley H. ;
Yenice, Kamil M. ;
Followill, David ;
Galvin, James M. ;
Hinson, William ;
Kavanagh, Brian ;
Keall, Paul ;
Lovelock, Michael ;
Meeks, Sanford ;
Papiez, Lech ;
Purdie, Thomas ;
Sadagopan, Ramaswamy ;
Schell, Michael C. ;
Salter, Bill ;
Schlesinger, David J. ;
Shiu, Almon S. ;
Solberg, Timothy ;
Song, Danny Y. ;
Stieber, Volker ;
Timmerman, Robert ;
Tome, Wolfgang A. ;
Verellen, Dirk ;
Wang, Lu ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2010, 37 (08) :4078-4101
[3]   Outcomes for Spine Stereotactic Body Radiation Therapy and an Analysis of Predictors of Local Recurrence [J].
Bishop, Andrew J. ;
Tao, Randa ;
Rebueno, Neal C. ;
Christensen, Eva N. ;
Allen, Pamela K. ;
Wang, Xin A. ;
Amini, Behrang ;
Tannir, Nizar M. ;
Tatsui, Claudio E. ;
Rhines, Laurence D. ;
Li, Jing ;
Chang, Eric L. ;
Brown, Paul D. ;
Ghia, Amol J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 92 (05) :1016-1026
[4]   Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel [J].
Casar, Bozidar ;
Gershkevitsh, Eduard ;
Mendez, Ignasi ;
Jurkovic, Slaven ;
Huq, M. Saiful .
MEDICAL PHYSICS, 2020, 47 (01) :242-259
[5]   A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams [J].
Casar, Bozidar ;
Gershkevitsh, Eduard ;
Mendez, Ignasi ;
Jurkovic, Slaven ;
Huq, M. Saiful .
MEDICAL PHYSICS, 2019, 46 (02) :944-963
[6]   Phase I/II study of stereotactic body radiotherapy for spinal metastasis and its pattern of failure [J].
Chang, Eric L. ;
Shiu, Almon S. ;
Mendel, Ehud ;
Mathews, Leni A. ;
Mahajan, Anita ;
Allen, Pamela K. ;
Weinberg, Jeffrey S. ;
Brown, Barry W. ;
Wang, Xin Shelly ;
Woo, Shiao Y. ;
Cleeland, Charles ;
Maor, Moshe H. ;
Rhines, Laurence D. .
JOURNAL OF NEUROSURGERY-SPINE, 2007, 7 (02) :151-160
[7]   Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning [J].
Chetty, Indrin J. ;
Curran, Bruce ;
Cygler, Joanna E. ;
DeMarco, John J. ;
Ezzell, Gary ;
Faddegon, Bruce A. ;
Kawrakow, Iwan ;
Keall, Paul J. ;
Liu, Helen ;
Ma, C. -M. Charlie ;
Rogers, D. W. O. ;
Seuntjens, Jan ;
Sheikh-Bagheri, Daryoush ;
Siebers, Jeffrey V. .
MEDICAL PHYSICS, 2007, 34 (12) :4818-4853
[8]  
Chopra KL, 2018, Therapeutic Radiology and Oncology, V2, P28, DOI [10.21037/tro.2018.07.01, 10.21037/tro.2018.07.01, DOI 10.21037/TRO.2018.07.01]
[9]   Metastatic epidural spinal cord compression [J].
Cole, John S. ;
Patchell, Roy A. .
LANCET NEUROLOGY, 2008, 7 (05) :459-466
[10]   International Spine Radiosurgery Consortium Consensus Guidelines for Target Volume Definition in Spinal Stereotactic Radiosurgery [J].
Cox, Brett W. ;
Spratt, Daniel E. ;
Lovelock, Michael ;
Bilsky, Mark H. ;
Lis, Eric ;
Ryu, Samuel ;
Sheehan, Jason ;
Gerszten, Peter C. ;
Chang, Eric ;
Gibbs, Iris ;
Soltys, Scott ;
Sahgal, Arjun ;
Deasy, Joe ;
Flickinger, John ;
Quader, Mubina ;
Mindea, Stefan ;
Yamada, Yoshiya .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (05) :E597-E605