Validation of a novel automated system, Fluispotter®, for serial sampling of dried blood spots

被引:1
作者
Krogh, Jesper [1 ,2 ]
Plomgaard, Peter [2 ,3 ]
Frikke-Schmidt, Ruth [2 ,3 ]
Velschow, Sten [4 ]
Johannesen, Jesper [2 ,5 ]
Hilsted, Linda Maria [3 ]
Schroder, Malene [4 ]
Feldt-Rasmussen, Ulla [1 ,2 ]
机构
[1] Copenhagen Univ Hosp, Dept Endocrinol & Metab, Rigshosp, Copenhagen, Denmark
[2] Univ Copenhagen, Dept Clin Med, Copenhagen, Denmark
[3] Copenhagen Univ Hosp, Dept Clin Biochem, Rigshosp, Copenhagen, Denmark
[4] Fluisense ApS, Lillerod, Denmark
[5] Copenhagen Univ Hosp Herlev & Gentofte, Dept Pediat, Copenhagen, Denmark
关键词
dried blood spots; cortisol; diurnal; automated sampling; CORTISOL;
D O I
10.1530/EC-23-0087
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Repeated blood sampling is required in certain clinical and research settings, which is currently performed by drawing blood from venous catheters requiring manual handling of each sample at the time of collection. A novel body-worn device for repeated serial samples, Fluispotter & REG;, with automated extraction, collection, and storage of up to 20 venous dried blood spot samples over the course of 20 h may overcome problems with current methods for serial sampling. The purpose of this study was to assess the performance and safety of Fluispotter for the first time in healthy subjects. Fluispotter consists of a cartridge with tubing, a reservoir for flushing solution, pumps and filterpaper, and a multi-lumen catheter placed in the brachial vein. We recruited healthy subjects for testing in an in-hospital setting. Fluispotter was attached by an anesthesiologist to 22 healthy subjects of which 9/22 (40.9%) participants had all 20 samples taken, which was lower than the goal of complete sampling in 80% of the subjects (P = 0.02). The main reason for sample failure was clogging of blood flow which was observed in 11/22 (50%) of the participants. No serious adverse events occurred, and the participants rated the pain from the insertion and the removal of catheter as very low. A cortisol profile showed nadir values at midnight and highest values at 05:00 h. Although full sampling was not successful in all participants, the Fluispotter technology proved safe and highly acceptable to the participants producing the expected cortisol profile without the requirement of staff during sample collection.
引用
收藏
页数:9
相关论文
共 21 条
  • [1] Fluispotter, a novel automated and wearable device for accurate volume serial dried blood spot sampling
    Adhikari, Khem Bahadur
    Rohde, Morten
    Velschow, Sten
    Feldt-Rasmussen, Ulla
    Johannesen, Jesper
    Johnsen, Anders Holten
    [J]. BIOANALYSIS, 2020, 12 (10) : 665 - 681
  • [2] Automated 24-hours sampling of subcutaneous tissue free cortisol in humans
    Bhake, R.C.
    Leendertz, J.A.
    Linthorst, A.C.E.
    Lightman, S.L.
    [J]. Bhake, R.C. (mdzrb@bristol.ac.uk), 1600, Informa Healthcare (37): : 180 - 184
  • [3] Ethnic differences in diurnal cortisol profiles in healthy adults: A meta-analysis
    Boileau, Kayla
    Barbeau, Kheana
    Sharma, Rupali
    Bielajew, Catherine
    [J]. BRITISH JOURNAL OF HEALTH PSYCHOLOGY, 2019, 24 (04) : 806 - 827
  • [4] Diagnosis and Treatment of Primary Adrenal Insufficiency: An Endocrine Society Clinical Practice Guideline
    Bornstein, Stefan R.
    Allolio, Bruno
    Arlt, Wiebke
    Barthel, Andreas
    Don-Wauchope, Andrew
    Hammer, Gary D.
    Husebye, Eystein S.
    Merke, Deborah P.
    Murad, M. Hassan
    Stratakis, Constantine A.
    Torpy, David J.
    [J]. JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2016, 101 (02) : 364 - 389
  • [5] Laboratory-based testing for hepatitis C infection using dried blood spot samples: A systematic review and meta-analysis of diagnostic accuracy
    Carty, Paul G.
    McCarthy, Michael
    O'Neill, Sinead M.
    De Gascun, Cillian F.
    Harrington, Patricia
    O'Neill, Michelle
    Smith, Susan M.
    Teljeur, Conor
    Ryan, Mairin
    [J]. REVIEWS IN MEDICAL VIROLOGY, 2022, 32 (04)
  • [6] Analysis of the cortisol diurnal rhythmicity and cortisol reactivity in long-term breast cancer survivors
    Couture-Lalande, Marie-Eve
    Lebel, Sophie
    Bielajew, Catherine
    [J]. BREAST CANCER MANAGEMENT, 2014, 3 (06) : 465 - 476
  • [7] Clinical feasibility of dried blood spots: Analytics, validation, and applications
    Enderle, Yeliz
    Foerster, Kathrin
    Burhenne, Juergen
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 130 : 231 - 243
  • [8] Fluisense VS., 2013, BODY FLUID SAMPLING
  • [9] Fluisense VS., 2016, IMPROVED FLUID SAMPL
  • [10] Dried Blood Spot Technique Applied in Therapeutic Drug Monitoring of Anticancer Drugs: a Review on Conversion Methods to Correlate Plasma and Dried Blood Spot Concentrations
    Iacuzzi, Valentina
    Posocco, Bianca
    Zanchetta, Martina
    Gagno, Sara
    Poetto, Ariana Soledad
    Guardascione, Michela
    Toffoli, Giuseppe
    [J]. PHARMACEUTICAL RESEARCH, 2021, 38 (05) : 759 - 778