The accuracy and precision of acetabular implant measurements from CT imaging

被引:2
作者
Henckel, Johann [1 ]
Ramesh, Angelika [1 ,2 ]
Hothi, Harry [1 ,2 ]
Richards, Robin [3 ]
Di Laura, Anna [1 ,2 ]
Hart, Alister [3 ]
机构
[1] Royal Natl Orthopaed Hosp NHS Trust, London, England
[2] UCL, Dept Mech Engn, London, England
[3] UCL, Inst Orthopaed & Musculoskeletal Sci, London, England
关键词
3D-CT; inclination; version; acetabular component position; coordinate measuring arm; cup orientation; TOTAL HIP-ARTHROPLASTY; METAL ARTIFACT REDUCTION; COMPUTED-TOMOGRAPHY; ANTEVERSION; PLACEMENT; COMPONENT; POSITION; PATIENT;
D O I
10.3389/fbioe.2023.1150061
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The placement of acetabular implant components determines the short-and long-term outcomes of total hip replacement (THR) and a number of tools have been developed to assist the surgeon in achieving cup orientation to match the surgical plan. However, the accuracy and precision of 3D-CT for the measurement of acetabular component position and orientation is yet to be established. To investigate this, we compared measurements of cobalt chrome acetabular components implanted into 2 different bony pelvic models between a coordinate measuring Faro arm and 3 different low dose CT images, including 3D-CT, 2D anterior pelvic plane (APP) referenced CT and 2D scanner referenced (SR) CT. Intra-observer differences were assessed using the Intraclass correlation coefficient (ICC). The effect of imaging the pelvis positioned in 3 different orientations within the CT scanner was also assessed. The measured parameters were the angles of inclination and version. 3D-CT measurements were found to closely match the "true values" of the component position measurements, compared with the 2D-CT methods. ICC analysis also showed good agreement between the coordinate measuring arm (CMA) and 3D-CT but poor agreement between the 2D SR method, in the results from two observers. When using the coordinate system of the CT scanner, the measurements consistently produced the greatest error; this method yielded values up to 34 degrees different from the reference digitising arm. However, the difference between the true inclination and version angles and those measured from 3D APP CT was below half a degree in all cases. We concluded that low radiation dose 3D-CT is a validated reference standard for the measurement of acetabular cup orientation.
引用
收藏
页数:9
相关论文
共 27 条
[1]   Validating the use of 3D biplanar radiography versus CT when measuring femoral anteversion after total hip arthroplasty A COMPARATIVE STUDY [J].
Anderson, C. G. ;
Brilliant, Z. R. ;
Jang, S. J. ;
Sokrab, R. ;
Mayman, D. J. ;
Vigdorchik, J. M. ;
Sculco, P. K. ;
Jerabek, S. A. .
BONE & JOINT JOURNAL, 2022, 104B (11) :1196-1201
[2]   Accuracy of measuring acetabular cup position after total hip arthroplasty: comparison between a radiographic planning software and three-dimensional computed tomography [J].
Bayraktar, Vahdettin ;
Weber, Markus ;
von Kunow, Frederik ;
Zeman, Florian ;
Craiovan, Benjamin ;
Renkawitz, Tobias ;
Grifka, Joachim ;
Woerner, Michael .
INTERNATIONAL ORTHOPAEDICS, 2017, 41 (04) :731-738
[3]   Improved accuracy of CT based measurements for anterior prominence of acetabular prosthesis using a novel protocol based on anatomical landmarks [J].
Brownlie, Craig Andrew ;
Evans, Rob ;
Morrison, David ;
Hayes, Alex ;
Song, Swithin ;
Kuster, Markus Stefan .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2020, 106 (03) :563-568
[4]   2D measurements of cup orientation are less reliable than 3D measurements A retrospective study of 87 metal-on-metal hips [J].
Davda, Kinner ;
Smyth, Niall ;
Cobb, Justin P. ;
Hart, Alister J. .
ACTA ORTHOPAEDICA, 2015, 86 (04) :485-490
[5]  
Deep K, 2017, SICOT-J, V3, DOI 10.1051/sicotj/2017034
[6]   Precision of acetabular cup placement in robotic integrated total hip arthroplasty [J].
Elson, Leah ;
Dounchis, Jon ;
Illgen, Richard ;
Marchand, Robert C. ;
Padgett, Douglas E. ;
Bragdon, Charles R. ;
Malchau, Henrik .
HIP INTERNATIONAL, 2015, 25 (06) :531-536
[7]   Advances and innovations in total hip arthroplasty [J].
Fontalis, Andreas ;
Epinette, Jean-Alain ;
Thaler, Martin ;
Zagra, Luigi ;
Khanduja, Vikas ;
Haddad, Fares S. .
SICOT-J, 2021, 7
[8]  
Health Department, 2001, IONISING RAD MEDICAL
[9]   Structural properties of fourth-generation composite femurs and tibias [J].
Heiner, Anneliese D. .
JOURNAL OF BIOMECHANICS, 2008, 41 (15) :3282-3284
[10]   3D-printed Patient-specific Guides for Hip Arthroplasty [J].
Henckel, Johann ;
Holme, Thomas J. ;
Radford, Warwick ;
Skinner, John A. ;
Hart, Alister J. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2018, 26 (16) :E342-E348