Impact of the automatic tube current modulation (ATCM) system on virtual monoenergetic image quality for dual-source CT: A phantom study

被引:4
|
作者
Dabli, D. [1 ,2 ,3 ]
Durand, Quentin [1 ,2 ]
Frandon, J. [1 ]
de Oliveira, Fabien [1 ]
Pastor, Maxime [1 ]
Beregi, J. P. [1 ]
Greffier, J. [1 ,2 ]
机构
[1] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, IMAGINE UR UM 103, F-30029 Nimes, France
[2] Nimes Univ Hosp, Dept Med Phys, F-30029 Nimes 9, France
[3] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, CHU Nimes,IMAGINE UR UM 103, Bd Prof Robert Debre, F-30029 Nimes 9, France
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2023年 / 109卷
关键词
Dual-source CT scan; Dual-energy CT; Virtual monochromatic images; Automatic tube current modulation; Detectability; ITERATIVE RECONSTRUCTION ALGORITHM; ENERGY COMPUTED-TOMOGRAPHY; LIVER-LESIONS; IODINE QUANTIFICATION; MAGNETIC-RESONANCE; EXPOSURE CONTROL; DOSE REDUCTION; OPTIMIZATION; PERFORMANCE; ACCURACY;
D O I
10.1016/j.ejmp.2023.102574
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the impact of the automatic tube current modulation (ATCM) on virtual monoenergetic images (VMIs) quality in dual-source CT(DSCT).Materials and methods: Acquisitions were performed on DSCT using the Mercury phantom. The acquisition pa-rameters for an abdomen-pelvic examination with single-energy CT(SECT) and dual-energy CT(DECT) imaging were used. Acquisitions were performed for each imaging mode using fixed mAs and ATCM. The mAs value was set to obtain a volume CT dose index of 11 mGy in fixed mAs acquisitions. This value was used as the reference mAs in ATCM acquisitions. The noise power spectrum and task-based transfer function at 40,50,60 and 70 keV levels were computed on VMIs and SECT images. The detectability index (d ') was calculated for a lesion with an iodine concentration of 10 mg/mL.Results: The noise magnitude on VMIs was higher with the ATCM system than with fixed mAs for all energy levels and section diameters of 21,26 and 31 cm. The noise texture and spatial resolution were similar between the fixed mAs and ATCM acquisitions for both imaging modes. The d' values were lower for all energy levels with ATCM than with fixed mAs acquisitions for 21 and 26 cm diameters by-39.82 +/- 9.32%, similar at 31 cm diameter-4.13 +/- 0.24% and higher at 36 cm diameter 10.40 +/- 6.69%. It was higher on VMIs at all energy levels compared to SECT images.Conclusions: The ATCM system could be used with DECT imaging to optimize patient exposure without changing the noise texture and spatial resolution of VMIs compared to fixed mAs and SECT.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT
    Agostini, Andrea
    Mari, Alberto
    Lanza, Cecilia
    Schicchi, Nicolo
    Borgheresi, Alessandra
    Maggi, Stefania
    Giovagnoni, Andrea
    RADIOLOGIA MEDICA, 2019, 124 (08): : 745 - 752
  • [42] Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT
    Yu, Lifeng
    Primak, Andrew N.
    Liu, Xin
    McCollough, Cynthia H.
    MEDICAL PHYSICS, 2009, 36 (03) : 1019 - 1024
  • [43] Performance of iodine quantification through high-pitch dual-source photon-counting CT: a phantom study
    Liu, Peng
    Zhou, Shanshui
    Dong, Haipeng
    Li, Jiqiang
    Xu, Zhihan
    Lin, Shushen
    Yang, Wenjie
    Yan, Fuhua
    Qin, Le
    JAPANESE JOURNAL OF RADIOLOGY, 2025, 43 (02) : 309 - 318
  • [44] Radiation dose and physical image quality in 128-section dual-source computed tomographic coronary angiography: a phantom study
    Matsubara, Kosuke
    Koshida, Haruka
    Sakuta, Keita
    Takata, Tadanori
    Horii, Junsei
    Iida, Hiroji
    Koshida, Kichiro
    Ichikawa, Katsuhiro
    Matsui, Osamu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (05): : 252 - 261
  • [45] The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter
    Tsalafoutas, Ioannis A.
    AlKhazzam, Shady
    Kharita, Mohammed Hassan
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (06):
  • [46] Technical Note: Evaluating automatic tube current modulation in CT using the standard CTDI dosimetry phantom
    Papadakis, Antonios E.
    Damilakis, John
    MEDICAL PHYSICS, 2021, 48 (02) : 659 - 666
  • [47] Dual-source photon counting detector CT with a tin filter: a phantom study on iodine quantification performance
    Tao, Ashley
    Huang, Richard
    Tao, Shengzhen
    Michalak, Gregory J.
    McCollough, Cynthia H.
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11)
  • [48] Virtual Monoenergetic Images (VMI plus ) in Dual-Source Dual-Energy CT Venography (DSDE-CTV) of the Lower Extremity Prior to Coronary Artery Bypass Graft (CABG): A Feasibility Study
    Schmoee, Jonas
    Dirrichs, Timm
    Fehrenbacher, Kerstin
    Tietz, Eric
    Hardt, Felix
    Schmid, Michael
    Gohmann, Robin Fabian
    Reinartz, Sebastian Daniel
    ACADEMIC RADIOLOGY, 2020, 27 (09) : 1249 - 1254
  • [49] Quantitative assessment of image noise and streak artifact on CT image: Comparison of z-axis automatic tube current modulation technique with fixed tube current technique
    Imai, Kuniharu
    Ikeda, Mitsuru
    Enchi, Yukihiro
    Niimi, Takanaga
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2009, 33 (05) : 353 - 358
  • [50] The impact of vertical off-centering on image noise and breast dose in chest CT with organ-based tube current modulation: A phantom study
    Makela, Teemu
    Kortesniemi, Mika
    Kaasalainen, Touko
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 100 : 153 - 163