Impact of the automatic tube current modulation (ATCM) system on virtual monoenergetic image quality for dual-source CT: A phantom study

被引:4
|
作者
Dabli, D. [1 ,2 ,3 ]
Durand, Quentin [1 ,2 ]
Frandon, J. [1 ]
de Oliveira, Fabien [1 ]
Pastor, Maxime [1 ]
Beregi, J. P. [1 ]
Greffier, J. [1 ,2 ]
机构
[1] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, IMAGINE UR UM 103, F-30029 Nimes, France
[2] Nimes Univ Hosp, Dept Med Phys, F-30029 Nimes 9, France
[3] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, CHU Nimes,IMAGINE UR UM 103, Bd Prof Robert Debre, F-30029 Nimes 9, France
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2023年 / 109卷
关键词
Dual-source CT scan; Dual-energy CT; Virtual monochromatic images; Automatic tube current modulation; Detectability; ITERATIVE RECONSTRUCTION ALGORITHM; ENERGY COMPUTED-TOMOGRAPHY; LIVER-LESIONS; IODINE QUANTIFICATION; MAGNETIC-RESONANCE; EXPOSURE CONTROL; DOSE REDUCTION; OPTIMIZATION; PERFORMANCE; ACCURACY;
D O I
10.1016/j.ejmp.2023.102574
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the impact of the automatic tube current modulation (ATCM) on virtual monoenergetic images (VMIs) quality in dual-source CT(DSCT).Materials and methods: Acquisitions were performed on DSCT using the Mercury phantom. The acquisition pa-rameters for an abdomen-pelvic examination with single-energy CT(SECT) and dual-energy CT(DECT) imaging were used. Acquisitions were performed for each imaging mode using fixed mAs and ATCM. The mAs value was set to obtain a volume CT dose index of 11 mGy in fixed mAs acquisitions. This value was used as the reference mAs in ATCM acquisitions. The noise power spectrum and task-based transfer function at 40,50,60 and 70 keV levels were computed on VMIs and SECT images. The detectability index (d ') was calculated for a lesion with an iodine concentration of 10 mg/mL.Results: The noise magnitude on VMIs was higher with the ATCM system than with fixed mAs for all energy levels and section diameters of 21,26 and 31 cm. The noise texture and spatial resolution were similar between the fixed mAs and ATCM acquisitions for both imaging modes. The d' values were lower for all energy levels with ATCM than with fixed mAs acquisitions for 21 and 26 cm diameters by-39.82 +/- 9.32%, similar at 31 cm diameter-4.13 +/- 0.24% and higher at 36 cm diameter 10.40 +/- 6.69%. It was higher on VMIs at all energy levels compared to SECT images.Conclusions: The ATCM system could be used with DECT imaging to optimize patient exposure without changing the noise texture and spatial resolution of VMIs compared to fixed mAs and SECT.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] A comparison of breast and lung doses from chest CT scans using organ-based tube current modulation (OBTCM) vs. Automatic tube current modulation (ATCM)
    Layman, Rick R.
    Hardy, Anthony J.
    Kim, Hyun J.
    Chou, Ei Ne
    Bostani, Maryam
    Cagnon, Chris
    Cody, Dianna
    McNitt-Gray, Michael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (05): : 97 - 109
  • [32] A Phantom Evaluation of the Use of Ct Automatic Tube Current Modulation with Low Tube Potentials for Iodinated Contrast Studies
    Sookpeng, Supawitoo
    Martin, Colin J.
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (03) : 188 - 195
  • [33] Role of Virtual Monoenergetic Images in the Assessment of Vessel Enhancement in Segmental Level in Third-Generation Dual-Source Dual-Energy CT Pulmonary Angiography-A Prospective Study
    Ethiraju, Vijayakumaran
    Varatharajaperumal, Rajeshkumar
    Arunachalam, Venkatesh Kasi
    KuruVambath, Abdulla
    Punniyakotti, Rajesh Shanmugam
    Rajasekaran, Sriman
    Mehta, Pankaj
    Cherian, Mathew
    INDIAN JOURNAL OF RADIOLOGY AND IMAGING, 2024,
  • [34] Liver virtual non-enhanced CT with dual-source, dual-energy CT: a preliminary study
    Long-Jiang Zhang
    Jin Peng
    Sheng-Yong Wu
    Z. Jane Wang
    Xin-Sheng Wu
    Chang-Sheng Zhou
    Xue-Man Ji
    Guang-Ming Lu
    European Radiology, 2010, 20 : 2257 - 2264
  • [35] Liver virtual non-enhanced CT with dual-source, dual-energy CT: a preliminary study
    Zhang, Long-Jiang
    Peng, Jin
    Wu, Sheng-Yong
    Wang, Z. Jane
    Wu, Xin-Sheng
    Zhou, Chang-Sheng
    Ji, Xue-Man
    Lu, Guang-Ming
    EUROPEAN RADIOLOGY, 2010, 20 (09) : 2257 - 2264
  • [36] Effect of High-Pitch Dual-Source CT to Compensate Motion Artifacts: A Phantom Study
    Farshad-Amacker, Nadia A.
    Alkadhi, Hatem
    Leschka, Sebastian
    Frauenfelder, Thomas
    ACADEMIC RADIOLOGY, 2013, 20 (10) : 1234 - 1239
  • [37] Comparison of image quality and radiation dose between combined automatic tube current modulation and fixed tube current technique in CT of abdomen and pelvis
    Lee, Sanghee
    Yoon, Sang-Wook
    Yoo, Seung-Min
    Ji, Young Geon
    Kim, Kyoung Ah
    Kim, Sang Heum
    Lee, Jong Tae
    ACTA RADIOLOGICA, 2011, 52 (10) : 1101 - 1106
  • [38] Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT
    Andrea Agostini
    Alberto Mari
    Cecilia Lanza
    Nicolo’ Schicchi
    Alessandra Borgheresi
    Stefania Maggi
    Andrea Giovagnoni
    La radiologia medica, 2019, 124 : 745 - 752
  • [39] Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging
    Sudarski, Sonja
    Apfaltrer, Paul
    Nance, John W., Jr.
    Schneider, David
    Meyer, Mathias
    Schoenberg, Stefan O.
    Fink, Christian
    Henzler, Thomas
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (10) : E574 - E581
  • [40] Dual-Source Versus Single-Source Cardiac CT Angiography: Comparison of Diagnostic Image Quality
    Donnino, Robert
    Jacobs, Jill E.
    Doshi, Jay V.
    Hecht, Elizabeth M.
    Kim, Danny C.
    Babb, James S.
    Srichai, Monvadi B.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (04) : 1051 - 1056