The Influence of Patello-Femoral Overstuffing After Modular Unlinked Bicompartmental Knee Arthroplasty (BiKA) for Medial Tibio-Femoral and Patello-Femoral Osteoarthritis of the Knee

被引:1
作者
Baba, Rikiya [1 ]
Ohkoshi, Yasumitsu [1 ]
Maeda, Tatsunori [1 ]
Suzuki, Ko [1 ]
Iwadate, Akane [1 ]
Iwasaki, Koji [2 ,3 ]
Onodera, Tomohiro [2 ,3 ]
Kondo, Eiji [4 ]
Iwasaki, Norimasa [2 ,3 ]
机构
[1] Hakodate Orthoped Clin, Dept Orthoped Surg, Hakodate, Hokkaido, Japan
[2] Hokkaido Univ, Fac Med, Dept Orthopaed Surg, Sapporo, Hokkaido, Japan
[3] Hokkaido Univ, Grad Sch Med, Sapporo, Hokkaido, Japan
[4] Hokkaido Univ Hosp, Ctr Sports Med, Sapporo, Hokkaido, Japan
关键词
bicompartmental knee arthroplasty; patello-femoral overstuffing; patello-femoral arthroplasty; unicompartmental knee arthroplasty; patient-reported outcome measures; ALIGNMENT; OUTCOMES; KINEMATICS;
D O I
10.1016/j.arth.2023.08.069
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The factors affecting results after bicompartmental knee arthroplasty (BiKA) have not been fully elucidated. This major ligament-preserving procedure may be more susceptible to overstuffing of the patello-femoral (PF) joint than the major ligament-sacrificing total knee arthroplasty. Currently, we investigated the effect of PF overstuffing after BiKA on its clinical outcome. Methods: There were 71 patients (74 knees) who underwent modular unlinked BiKA at our clinic who had a follow-up of 5 to 9 years. Final follow-up results were assessed by evaluating knee range of motion, the 2011 Knee Society Score (2011KSS), Japanese Knee Osteoarthritis Measure, and radiological findings. The degree of postoperative PF overstuffing was evaluated by computed tomography and magnetic resonance images for 55 knees, and the correlation between the degree of overstuffing and postoperative clinical results were examined. Results: Overall clinical results improved significantly after surgery without any revision cases. The X-ray measurements showed the improved coronal alignments and the appropriate implant installation angles. Higher degree of postoperative PF overstuffing caused by insufficient amount of osteotomy on the anterior surface of the femur correlated with worse postoperative total 2011KSS at 2 years after surgery (Spearman's rank correlation coefficient (rs) = -0.387, P = .004), as opposed to no correlation at the time of the final follow-up (Spearman's rank correlation coefficient = 0.068, P = .623). Conclusion: Modular unlinked BiKA provided patients with a high level of satisfaction and functional improvement over 5 to 9 years postoperatively. However, because PF overstuffing affects initial patient satisfaction, the amount of osteotomy should be determined carefully during the surgery. (c) 2023 Elsevier Inc. All rights reserved.
引用
收藏
页码:638 / 644
页数:7
相关论文
共 32 条
[1]  
Akai M, 2005, J RHEUMATOL, V32, P1524
[2]   Journey-Deuce bicompartmental knee arthroplasty with the addition of computer navigation achieves good clinical outcomes and implant survival at 10 years [J].
Aujla, Randeep S. ;
Woodhouse, Jennifer ;
Ebert, Jay R. ;
Finsterwald, Michael ;
Jones, Christopher W. ;
Yates, Piers ;
D'Alessandro, Peter ;
Wood, David J. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2022, 30 (09) :3168-3175
[3]   Total Knee Arthroplasty for Osteoarthritis in Patients Less Than Fifty-Five Years of Age: A Systematic Review [J].
Aujla, Randeep S. ;
Esler, Colin N. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (08) :2598-+
[4]   Patella tracking and patella contact pressure in modular patellofemoral arthroplasty: a biomechanical in vitro analysis [J].
Calliess, Tilman ;
Ettinger, Max ;
Schado, Ssuheib ;
Becher, Christoph ;
Hurschler, Christof ;
Ostermeier, Sven .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2016, 136 (06) :849-855
[5]   Are 20% of Patients Actually Dissatisfied Following Total Knee Arthroplasty? A Systematic Review of the Literature [J].
DeFrance, Michael J. ;
Scuderi, Giles R. .
JOURNAL OF ARTHROPLASTY, 2023, 38 (03) :594-599
[6]  
EWALD FC, 1989, CLIN ORTHOP RELAT R, P9
[7]   Medial bicompartmental arthroplasty patients display more normal gait and improved satisfaction, compared to matched total knee arthroplasty patients [J].
Garner, Amy J. ;
Dandridge, Oliver W. ;
van Arkel, Richard J. ;
Cobb, Justin P. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2023, 31 (03) :830-838
[8]   The effect of overstuffing the patellofemoral joint on the extensor retinaculum of the knee [J].
Ghosh, K. M. ;
Merican, A. M. ;
Iranpour, F. ;
Deehan, D. J. ;
Amis, Andrew A. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2009, 17 (10) :1211-1216
[9]   Ten year outcomes for the prospective randomised trial comparing unlinked, modular bicompartmental knee arthroplasty and total knee arthroplasty [J].
Goh, Jeremy Keng Meng ;
Chen, Jerry Yongqiang ;
Yeo, Nicholas Eng Meng ;
Liow, Ming Han Lincoln ;
Chia, Shi-Lu ;
Yeo, Seng Jin .
KNEE, 2020, 27 (06) :1914-1922
[10]   THE OXFORD KNEE FOR UNICOMPARTMENTAL OSTEO-ARTHRITIS - THE 1ST 103 CASES [J].
GOODFELLOW, JW ;
KERSHAW, CJ ;
BENSON, MKD ;
OCONNOR, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (05) :692-701