Brain-implanted conductors amplify radiofrequency fields in rodents: Advantages and risks

被引:2
作者
Voeroeslakos, Mihaly [1 ]
Yaghmazadeh, Omid [1 ]
Alon, Leeor [2 ]
Sodickson, Daniel K. [1 ,2 ,4 ]
Buzsaki, Gyoergy [1 ,3 ,4 ]
机构
[1] NYU, Langone Med Ctr, Neurosci Inst, Dept Neurosci & Physiol, New York, NY USA
[2] NYU, Grossman Sch Med, Dept Radiol, New York, NY USA
[3] NYU, Dept Neurol, Grossman Sch Med, New York, NY USA
[4] NYU, Neurosci Inst, Langone Med Ctr, New York, NY 10016 USA
基金
美国国家卫生研究院;
关键词
brain implants; brain stimulation; implant safety; radiofrequency stimulation; MAGNETIC-RESONANCE; STIMULATION ELECTRODES; MRI; SAFETY; SYSTEMS; TEMPERATURE; CATHETERS; RESPONSES; PATTERNS; EXPOSURE;
D O I
10.1002/bem.22489
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Over the past few decades, daily exposure to radiofrequency (RF) fields has been increasing due to the rapid development of wireless and medical imaging technologies. Under extreme circumstances, exposure to very strong RF energy can lead to heating of body tissue, even resulting in tissue injury. The presence of implanted devices, moreover, can amplify RF effects on surrounding tissue. Therefore, it is important to understand the interactions of RF fields with tissue in the presence of implants, in order to establish appropriate wireless safety protocols, and also to extend the benefits of medical imaging to increasing numbers of people with implanted medical devices. This study explored the neurological effects of RF exposure in rodents implanted with neuronal recording electrodes. We exposed freely moving and anesthetized rats and mice to 950 MHz RF energy while monitoring their brain activity, temperature, and behavior. We found that RF exposure could induce fast onset firing of single neurons without heat injury. In addition, brain implants enhanced the effect of RF stimulation resulting in reversible behavioral changes. Using an optical temperature measurement system, we found greater than tenfold increase in brain temperature in the vicinity of the implant. On the one hand, our results underline the importance of careful safety assessment for brain-implanted devices, but on the other hand, we also show that metal implants may be used for neurostimulation if brain temperature can be kept within safe limits.
引用
收藏
页码:139 / 155
页数:17
相关论文
共 71 条
[1]   Vibrational Resonances in biological systems at microwave frequencies [J].
Adair, RK .
BIOPHYSICAL JOURNAL, 2002, 82 (03) :1147-1152
[2]   Heat equation inversion framework for average SAR calculation from magnetic resonance thermal imaging [J].
Alon, Leeor ;
Sodickson, Daniel K. ;
Deniz, Cem M. .
BIOELECTROMAGNETICS, 2016, 37 (07) :493-503
[3]  
[Anonymous], 1996, FCC96326
[4]  
[Anonymous], 2018, Assessment of the Safety of Magnetic Resonance Imaging for Patients with an Active Implantable Medical Device
[5]  
ASTM F2503-20, 2000, F250320 ASTM US DEP
[6]  
Babouri A, 2007, P ANN INT IEEE EMBS, P5684
[7]   Implantable Devices: Issues and Challenges [J].
Bazaka, Kateryna ;
Jacob, Mohan V. .
ELECTRONICS, 2013, 2 (01) :1-34
[8]   Responses of neurons to an amplitude modulated microwave stimulus [J].
Beason, RC ;
Semm, P .
NEUROSCIENCE LETTERS, 2002, 333 (03) :175-178
[9]   Improving Safety of MRI in Patients with Deep Brain Stimulation Devices [J].
Boutet, Alexandre ;
Chow, Clement T. ;
Narang, Keshav ;
Elias, Gavin J. B. ;
Neudorfer, Clemens ;
Germann, Juergen ;
Ranjan, Manish ;
Loh, Aaron ;
Martin, Alastair J. ;
Kucharczyk, Walter ;
Steele, Christopher J. ;
Hancu, Ileana ;
Rezai, Ali R. ;
Lozano, Andres M. .
RADIOLOGY, 2020, 296 (02) :250-262
[10]   3-Tesla MRI of deep brain stimulation patients: safety assessment of coils and pulse sequences [J].
Boutet, Alexandre ;
Hancu, Ileana ;
Saha, Utpal ;
Crawley, Adrian ;
Xu, David S. ;
Ranjan, Manish ;
Hlasny, Eugen ;
Chen, Robert ;
Foltz, Warren ;
Sammartino, Francesco ;
Coblentz, Ailish ;
Kucharczyk, Walter ;
Lozano, Andres M. .
JOURNAL OF NEUROSURGERY, 2020, 132 (02) :586-594