Implantable cardioverter defibrillator lead performance: A systematic review and individual patient data Meta-analysis

被引:3
作者
Giacopelli, Daniele [1 ,2 ,8 ]
Azzolina, Danila [3 ]
Comoretto, Rosanna Irene [4 ]
Quartieri, Fabio [5 ]
Rovaris, Giovanni [6 ]
Schillaci, Vincenzo [7 ]
Gargaro, Alessio [2 ]
Gregori, Dario [1 ]
机构
[1] Univ Padua, Dept Cardiac Thorac Vasc Sci & Publ Hlth, Padua, Italy
[2] Biotronik Italia, Clin Unit, Milan, Italy
[3] Univ Ferrara, Dept Environm & Prevent Sci, Ferrara, Italy
[4] Univ Torino, Dept Publ Hlth & Pediat, Turin, Italy
[5] Arcispedale Santa Maria Nuova, Dept Cardiol, Reggio Emilia, Italy
[6] Osped San Gerardo, Dept Cardiol, Monza, Italy
[7] Clin Montevergine, Electrophysiol Unit, Mercogliano, Italy
[8] Univ Padua, Dept Cardiac Thorac Vasc Sci & Publ Hlth, Via Giustiniani 2, I-35128 Padua, Italy
关键词
Implantable cardioverter-defibrillator; Post market surveillance; ICD leads; Failure; meta; -analysis; LONG-TERM PERFORMANCE; SPRINT FIDELIS; FAILURE RATES; BIOTRONIK LINOX; RIATA; RISK; SURVIVAL; SURVEILLANCE; MULTICENTER; DYSFUNCTION;
D O I
10.1016/j.ijcard.2022.11.048
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Reliable post-approval surveillance of implantable cardioverter-defibrillator (ICD) lead performance remains a challenge. In the past, two ICD leads were recalled due to a high frequency of failures. In this meta -analysis, we sought to provide a combined estimate of failure-free rate for ICD leads by reconstructing individual patient data from published Kaplan-Meier (KM) curves and to investigate whether estimates could be influenced by the characteristics of the study.Methods: Observational studies assessing failure-free estimates of transvenous ICD leads with KM method, were identified through a systematic search up to November 2021.Results: Forty-four studies were eligible that included 41,870 (63.1%) non-recalled leads and 24,493 (36.9%) recalled leads. The 8-year cumulative failure-free rate was 94.1% (CI, 93.6% -94.6%) for contemporary non-recalled leads and 81.2% (80.3% -82.0%) for recalled leads (hazard ratio [HR], 3.15 [2.85-3.47], p < 0.001). Failure-free rate was lower in single-center studies in both the non-recalled (HR, 0.28 [0.15-0.51], p < 0.001) and recalled (HR, 0.54 [0.33-0.88], p = 0.014) group compared with multicenter studies. Similarly, estimates were significantly lower in small (i.e. extracted KM curve with <312 leads) versus large studies (HR non-recalled group, 0.54 [CI, 0.33-0.89], p = 0.015; HR recalled group, 0.62 [CI, 0.43-0.89], p = 0.009). Conclusions: In this meta-analysis including >66,000 leads, we provide pooled survival curves that may play a role in generating evidence-based standards for assessing clinically acceptable failure rates for ICD leads. Lead performance was underestimated with single-center and small-sized studies; multicenter studies remain the main tool to reliably conduct post-market surveillance of ICD leads.
引用
收藏
页码:57 / 63
页数:7
相关论文
共 50 条
[21]   Effectiveness of cognitive behavioral therapy on mood symptoms in patients with implantable cardioverter defibrillator: A systematic review and meta-analysis [J].
Li, Zimeng ;
Liu, Ying ;
Wang, Jiayao ;
Zhang, Chunmei ;
Liu, Yanhui .
COMPLEMENTARY THERAPIES IN CLINICAL PRACTICE, 2022, 47
[22]   Performance of Lead Integrity Alert to Assist in the Clinical Diagnosis of Implantable Cardioverter Defibrillator Lead Failures Analysis of Different Implantable Cardioverter Defibrillator Leads [J].
Ellenbogen, Kenneth A. ;
Gunderson, Bruce D. ;
Stromberg, Kurt D. ;
Swerdlow, Charles D. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2013, 6 (06) :1169-1177
[23]   Association between fine particulate matter and atrial fibrillation in implantable cardioverter defibrillator patients: a systematic review and meta-analysis [J].
Yue, Chao ;
Yang, Fan ;
Wang, Luyi ;
Li, Fengwei ;
Chen, Yingtai .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2020, 59 (03) :595-601
[24]   Atrial Fibrillation Is Associated With Higher Overall Mortality in Patients With Implantable Cardioverter-Defibrillator: A Systematic Review and Meta-Analysis [J].
Mustafa, Usman ;
Dherange, Parinita ;
Reddy, Rohit ;
DeVillier, Joseph ;
Chong, Jessica ;
Ihsan, Alarozia ;
Jones, Ryan ;
Duddyala, Narendra ;
Reddy, Pratap ;
Dominic, Paari .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2018, 7 (22)
[25]   Association between fine particulate matter and atrial fibrillation in implantable cardioverter defibrillator patients: a systematic review and meta-analysis [J].
Chao Yue ;
Fan Yang ;
Luyi Wang ;
Fengwei Li ;
Yingtai Chen .
Journal of Interventional Cardiac Electrophysiology, 2020, 59 :595-601
[26]   Implantable cardioverter defibrillator therapy in chronic kidney disease: a meta-analysis [J].
Korantzopoulos, Panagiotis ;
Liu, Tong ;
Li, Lijian ;
Goudevenos, John A. ;
Li, Guangping .
EUROPACE, 2009, 11 (11) :1469-1475
[27]   The subcutaneous implantable cardioverter defibrillator-review of the recent data [J].
Westerman, Stacy B. ;
El-Chami, Mikhael .
JOURNAL OF GERIATRIC CARDIOLOGY, 2018, 15 (03) :222-228
[28]   Predictors of Mortality in Patients With an Implantable Cardiac Defibrillator: A Systematic Review and Meta-analysis [J].
Alba, Ana C. ;
Braga, Juarez ;
Gewarges, Mena ;
Walter, Stephen D. ;
Guyatt, Gordon H. ;
Ross, Heather J. .
CANADIAN JOURNAL OF CARDIOLOGY, 2013, 29 (12) :1729-1740
[29]   Implantable cardioverter-defibrillators in adults with congenital heart disease: a systematic review and meta-analysis [J].
Vehmeijer, Jim T. ;
Brouwer, Tom F. ;
Limpens, Jacqueline ;
Knops, Reinoud E. ;
Bouma, Berto J. ;
Mulder, Barbara J. M. ;
de Groot, Joris R. .
EUROPEAN HEART JOURNAL, 2016, 37 (18) :1439-1448
[30]   Importance of Implantable Cardioverter-Defibrillator Back-Up in Cardiac Resynchronization Therapy Recipients: A Systematic Review and Meta-Analysis [J].
Barra, Sergio ;
Providencia, Rui ;
Tang, Anthony ;
Heck, Patrick ;
Virdee, Munmohan ;
Agarwal, Sharad .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2015, 4 (11)