Effect of inferior vena cava filter placement position on device complications

被引:7
|
作者
Grubman, Scott [1 ,5 ]
Kostiuk, Valentyna [1 ]
Brahmandam, Anand [2 ]
Tonnessen, Britt [2 ]
Mojibian, Hamid [3 ]
Schneider, Eric [4 ]
Guzman, Raul J. [2 ]
Chaar, Cassius Iyad Ochoa [2 ]
机构
[1] Yale Univ, Dept Surg, Sch Med, New Haven, CT USA
[2] Yale Univ, Div Vasc Surg & Endovasc Therapy, Dept Surg, Sch Med, New Haven, CT USA
[3] Yale Univ, Dept Radiol & Biomed Imaging, Sch Med, New Haven, CT USA
[4] Yale Univ, Ctr Hlth Serv & Outcomes Res, Dept Surg, Sch Med, New Haven, CT USA
[5] Yale Univ, Sch Med, Dept Surg, 333 Cedar St, New Haven, CT 06510 USA
关键词
Inferior vena cava; Postoperative complications; Thrombosis; Venous thromboembolism; RENAL-VEIN INFLOW; UPDATE; CELECT; TRENDS; IMPACT; RATES;
D O I
10.1016/j.jvsv.2023.05.020
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Indwelling inferior vena cava (IVC) filters can cause complications, including penetration into surrounding structures, migration, and thrombosis of the vena cava. Computational fluid dynamics suggests juxtarenal placement of IVC filters decreases the risk of thrombosis; however, this has not been explored clinically. The present study examines the effect of filter placement position on long-term device complications with an emphasis on IVC thrombosis. We hypothesized that IVC filters placed further caudal to the renal veins were more likely to develop long-term thrombosis.Methods: A retrospective review of the medical records of patients receiving IVC filters at a single tertiary center between 2008 and 2016 was performed. Patients missing follow-up or procedural imaging data were excluded. The placement procedure venograms were reviewed, and the distance from the filter apex to the more inferior renal vein was measured using reported IVC filter lengths for calibration. The patients were divided into three groups according to the tip position relative to the more inferior renal vein: at or superior (group A), 1 to 20 mm inferior (group B), and >20 mm inferior (group C). The patient and procedural characteristics and outcomes were compared between the three groups. The primary end points were IVC thrombosis and device-related mortality.Results: Of 1497 eligible patients, 267 (17.8%) were excluded. The most common placement position was group B (64.0%). The mean age was lowest in group C, followed by groups A and B (age, 59.5 years, 64.6 years, and 62.2 years, respectively; P = .003). No statistically significant differences were found in the distribution of sex or the measured comorbidities. Group C was the most likely to receive jugular access (group C, 71.7%; group A, 48.3%; group B, 62.4%; P < .001) and received more first-generation filters (group C, 58.5%; group A, 46.6%; group B, 52.5%; P = .045). The short-term (<30-day) and long-term (>= 30-day) outcomes, including access site hematoma, deep vein thrombosis, and pulmonary embolism, were uncommon, with no differences between the groups. Cases of symptomatic filter penetration, migration, and fracture were rare (one, one, and three cases, respectively). Although a pattern of increasing thrombosis with more inferior placement was found, the difference between groups was not statistically significant (group A, 1.5%; group B, 1.8%; group C, 2.5%; P = .638). No cases of device-related mortality occurred. All-cause mortality after a mean follow-up of 2.6 +/- 2.3 years was 41.3% and did not vary significantly between the groups (P = .051). Multivariate logistic regression revealed that placement position did not predict for short- or long-term deep vein thrombosis, pulmonary embolism, IVC thrombosis, or all-cause mortality after adjustment for the baseline patient characteristics.Conclusions: IVC filters have low rates of short- and long-term complications, including IVC thrombosis. The placement position did not affect the occurrence of device complications in this study.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Retrospective analysis of outcomes following inferior vena cava (IVC) filter placement in a managed care population
    Everhart, Damian
    Vaccaro, Jamieson
    Worley, Karen
    Rogstad, Teresa L.
    Seleznick, Mitchel
    JOURNAL OF THROMBOSIS AND THROMBOLYSIS, 2017, 44 (02) : 179 - 189
  • [42] Racial and ethnic disparities in inferior vena cava filter placement for deep vein thrombosis in the United States
    Juarez, Jordan J.
    Khalid, Muhammad U.
    Ulloa, Bianca A.
    Romero, Carlos M.
    Maruthi, Rohit
    Shah, Devrat
    Chang, Eric
    Shafi, Irfan
    Lakhter, Vladimir
    Zhao, Huaqing
    Rodriquez, Erik J.
    Perez-Stable, Eliseo J.
    Bashir, Riyaz
    JOURNAL OF VASCULAR SURGERY-VENOUS AND LYMPHATIC DISORDERS, 2024, 12 (01)
  • [43] Difficult withdrawal of an inferior vena cava filter: technical considerations and associated variables
    Serrano, E.
    Vila-Trias, E.
    Zarco, F.
    Martinez, C. Zamora
    Moises, J.
    Gomez, F. M.
    Lopez-Rueda, A.
    RADIOLOGIA, 2023, 65 (03): : 230 - 238
  • [44] Anticoagulation Is Associated with Decreased Inferior Vena Cava Filter-Related Complications in Patients with Metastatic Carcinoma
    Mahmood, Syed S.
    Abtahian, Farhad
    Fogerty, Annemarie E.
    Cefalo, Philip
    MacKay, Cheryl
    Jaff, Michael R.
    Weinberg, Ido
    AMERICAN JOURNAL OF MEDICINE, 2017, 130 (01) : 77 - +
  • [45] Successful retrieval of a Ionglasting temporary inferior vena cava filter
    Couturier, M.
    Annweiler, T.
    Bertoletti, L.
    Barral, F. -G.
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2016, 97 (04) : 481 - 482
  • [46] Symptomatic caval penetration by a Celect inferior vena cava filter
    Conor O. Bogue
    Philip R. John
    Bairbre L. Connolly
    David J. Rea
    Joao G. Amaral
    Pediatric Radiology, 2009, 39 : 1110 - 1113
  • [47] Symptomatic caval penetration by a Celect inferior vena cava filter
    Bogue, Conor O.
    John, Philip R.
    Connolly, Bairbre L.
    Rea, David J.
    Amaral, Joao G.
    PEDIATRIC RADIOLOGY, 2009, 39 (10) : 1110 - 1113
  • [48] Effect of Delayed Inferior Vena Cava Filter Retrieval After Early Initiation of Anticoagulation
    Weinberg, Ido
    Abtahian, Farhad
    DeBiasi, Ralph
    Cefalo, Philip
    MacKay, Cheryl
    Hawkins, Beau M.
    Jaff, Michael R.
    AMERICAN JOURNAL OF CARDIOLOGY, 2014, 113 (02) : 389 - 394
  • [49] Usefulness of a Computerized Reminder System to Improve Inferior Vena Cava Filter Retrieval and Complications
    Mikhael, Bassem
    Albaghdadi, Mazen
    Abtahian, Farhad
    MacKay, Cheryl
    Secemsky, Eric
    Jaff, Michael R.
    Weinberg, Ido
    AMERICAN JOURNAL OF CARDIOLOGY, 2019, 123 (02) : 348 - 353
  • [50] Bilateral inferior vena cava filter insertion in a patient with duplication of the infrarenal vena cava
    S. Leong
    F. Oisin
    J. E. Barry
    M. M. Maher
    C. O. Bogue
    Irish Journal of Medical Science, 2012, 181 : 389 - 391