Continuous flow of reagents interferes with the kinetics and equilibrium of antibody-antigen binding in microfluidic heterogeneous immunoassays

被引:4
作者
Barbosa, Ana I. [1 ]
Reis, Nuno M. [2 ,3 ]
机构
[1] Loughborough Univ, Dept Chem Engn, Loughborough LE11 3TU, England
[2] Univ Bath, Dept Chem Engn, Bath BA2 7AY, England
[3] Univ Bath, Ctr Biosensors Bioelect & Biodevices C3Bio, Claverton, Bath BA2 7AY, England
关键词
Heterogenous immunoassay; Reagent flow; Microfluidic; Fluoropolymer; Antibody-antigen binding; Kinetics; HIGH-SENSITIVITY; QUANTITATION; BIOMARKERS; CHIP; ADSORPTION; DIFFUSION; DEVICES; SYSTEM; VOLUME;
D O I
10.1016/j.snb.2023.134454
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Though fundamentally important for the design and development of quantitative point -of -care diagnostics devices, the effect of fluid flow in heterogeneous immunoassays remains unclear. We present the first experimental study using actual microfluidic strips showing both kinetics and equilibrium of antibody-analyte binding are dependent on flow rate, Q of reagents/sample. We have passively adsorbed an antibody in the inner surface of 212 +/- 12 mu m, 10 -bore microcapillary film and pumped continuously a second (antigen) molecule in the range Q = 10-1000 mu L/min. Sections of the film with 40 mm in length were then sacrificed over time enabling ELISA quantitation of the extent of antibody -antigen binding on each strip; other microfluidic technologies would require sacrificing the whole device, making it laborious and expensive. Absorbance data fitted to a kinetic model enabled determining Kon, Koff and affinity constant Keq = Kon/Koff, confirming the impact of Q. With limiting antigen, a 4 -fold reduction in Kon and 3 -fold in Keq was observed by shifting antigen incubation from stagnant to continuous Q = 10 mu L/min. Flow caused up to 2 -fold increment in the limit of detection (LoD) and assay time, suggesting a direct physical interference of fluid flow on the molecular binding and non-specific (background) binding, presumably through the effect of shear flow, which could include re -orientation of biomolecules. Although the biophysics behind such flow effect is yet to be understood, this work will help speeding up the development of protocols for development and implementation of microfluidic immunoassay technology for high-performance point -of -care (POC) testing.
引用
收藏
页数:8
相关论文
共 33 条
[1]   Sensitive optical detection of clinically relevant biomarkers in affordable microfluidic devices: Overcoming substrate diffusion limitations [J].
Barbosa, Ana I. ;
Castanheira, Ana P. ;
Reis, Nuno M. .
SENSORS AND ACTUATORS B-CHEMICAL, 2018, 258 :313-320
[2]   A critical insight into the development pipeline of microfluidic immunoassay devices for the sensitive quantitation of protein biomarkers at the point of care [J].
Barbosa, Ana I. ;
Reis, Nuno M. .
ANALYST, 2017, 142 (06) :858-882
[3]   Portable smartphone quantitation of prostate specific antigen (PSA) in a fluoropolymer microfluidic device [J].
Barbosa, Ana I. ;
Gehlot, Poonam ;
Sidapra, Kalpita ;
Edwards, Alexander D. ;
Reis, Nuno M. .
BIOSENSORS & BIOELECTRONICS, 2015, 70 :5-14
[4]   A lab-in-a-briefcase for rapid prostate specific antigen (PSA) screening from whole blood [J].
Barbosa, Ana I. ;
Castanheira, Ana P. ;
Edwards, Alexander D. ;
Reis, Nuno M. .
LAB ON A CHIP, 2014, 14 (16) :2918-2928
[5]   Transparent, Hydrophobic Fluorinated Ethylene Propylene Offers Rapid, Robust, and Irreversible Passive Adsorption of Diagnostic Antibodies for Sensitive Optical Biosensing [J].
Barbosa, Ana Isabel ;
Barreto, Augusto Sampaio ;
Reis, Nuno Miguel .
ACS APPLIED BIO MATERIALS, 2019, 2 (07) :2780-2790
[6]   Cytokine profiles as markers of disease severity in sepsis: a multiplex analysis [J].
Bozza, Fernando A. ;
Salluh, Jorge I. ;
Japiassu, Andre M. ;
Soares, Marcio ;
Assis, Edson F. ;
Gomes, Rachel N. ;
Bozza, Marcelo T. ;
Castro-Faria-Neto, Hugo C. ;
Bozza, Patricia T. .
CRITICAL CARE, 2007, 11 (02)
[7]   Multiplexed femtomolar quantitation of human cytokines in a fluoropolymer microcapillary film [J].
Castanheira, Ana P. ;
Barbosa, Ana I. ;
Edwardsa, Alexander D. ;
Reis, Nuno M. .
ANALYST, 2015, 140 (16) :5609-5618
[8]   High-sensitivity miniaturized immunoassays for tumor necrosis factor a using microfluidic systems [J].
Cesaro-Tadic, S ;
Dernick, G ;
Juncker, D ;
Buurman, G ;
Kropshofer, H ;
Michel, B ;
Fattinger, C ;
Delamarche, E .
LAB ON A CHIP, 2004, 4 (06) :563-569
[9]   Commercialization of microfluidic point-of-care diagnostic devices [J].
Chin, Curtis D. ;
Linder, Vincent ;
Sia, Samuel K. .
LAB ON A CHIP, 2012, 12 (12) :2118-2134
[10]   Simulation and Theory of Antibody Binding to Crowded Antigen-Covered Surfaces [J].
De Michele, Cristiano ;
De Los Rios, Paolo ;
Foffi, Giuseppe ;
Piazza, Francesco .
PLOS COMPUTATIONAL BIOLOGY, 2016, 12 (03)