Optimising image quality in intravenous cerebral cone beam computed tomography

被引:2
作者
Broadley, Lisa [1 ]
Erskine, Brendan [1 ]
Marshall, Elissa [1 ]
Ewert, Kyle [1 ]
Picker, Bjoern [1 ]
机构
[1] Alfred Hosp, Melbourne, Vic 3004, Australia
关键词
angiography; head and neck; vascular interventional; CONTRAST-MEDIUM; CT ANGIOGRAPHY; DOSE REDUCTION; TUBE VOLTAGE; FOLLOW-UP; RADIATION; INJECTION; OPTION;
D O I
10.1002/jmrs.735
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
IntroductionThe efficacy of intravenous cerebral Cone Beam Computed Tomography (IV CBCT) is well established; however, image quality has only ever been authenticated by subjective evaluation. The aim of this study was to quantify the factors pertinent to achieving consistent and optimal image quality when performing IV CBCT.MethodsBetween 1 March 2021 and 30 October 2022, 79 patients received IV CBCT. These candidates were divided into three main acquisition field size categories (22/32, 42 and 48 cm) according to the clinical indication. The images were analysed using both a quantitative assessment and a subjective evaluation. Here, a comparison of Hounsfield units (HUs), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and noise index was conducted for each study and compared relative to the acquisition field size. The subjective analysis was performed in a non-blinded fashion where the diagnostic value (DV) of the exam was determined according to a graded scale. A phantom analysis for each of the acquisition field sizes was conducted and modulation transfer function (MTF) graphed.ResultsSignificantly higher HU, SNR, CNR and lower noise indices were achieved with the 42-cm protocol than the 22/32 and 48-cm protocols. Here a greater DV was also reported. The MTF demonstrates marginally improved spatial resolution for the 22-cm protocol, but this is near equivocal for the 32-, 42 and 48-cm protocols.ConclusionThe use of larger acquisition field sizes provides improved image quality when performing IV CBCT as an alternative to intra-arterial (IA) CBCT. The efficacy of intravenous (IV) Cone Beam Computed Tomography (CBCT) is well established; however the image quality has only ever been authenticated by subjective evaluation. The aim of this study was to quantify the factors pertinent to achieving consistent and optimal image quality when performing IV CBCT.image
引用
收藏
页码:26 / 34
页数:9
相关论文
共 27 条
[11]   Cone-Beam CT Compared to Multi-Slice CT for the Diagnostic Analysis of Conductive Hearing Loss: A Feasibility Study [J].
Kemp, Pieter ;
Van Stralen, Jiska ;
De Graaf, Pim ;
Berkhout, Erwin ;
Van Horssen, Pepijn ;
Merkus, Paul .
JOURNAL OF INTERNATIONAL ADVANCED OTOLOGY, 2020, 16 (02) :222-+
[12]   A technical note on intra-arterial cone-beam computed tomography for the evaluation of flow-diverter stents: Image quality differences between diluted and non-diluted contrast medium [J].
Kocer, Naci ;
Kandemirli, Sedat G. ;
Ruijters, Daniel ;
Mantatzis, Michalis ;
Kizilkilic, Osman ;
Islak, Civan .
INTERVENTIONAL NEURORADIOLOGY, 2020, 26 (02) :164-169
[13]   Dose reduction in cone-beam CT scanning for intracranial stent deployment before coil embolization of intracranial wide-neck aneurysms [J].
Kuriyama, Takumi ;
Sakai, Nobuyuki ;
Niida, Norimitsu ;
Sueoka, Masaki ;
Beppu, Mikiya ;
Dahmani, Chihebeddine ;
Kojima, Iwao ;
Sakai, Chiaki ;
Imamura, Hirotoshi ;
Masago, Katsuhiro ;
Katakami, Nobuyuki .
INTERVENTIONAL NEURORADIOLOGY, 2016, 22 (04) :420-425
[14]   Tube Potential and CT Radiation Dose Optimization [J].
Lira, Diego ;
Padole, Atul ;
Kalra, Mannudeep K. ;
Singh, Sarabjeet .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (01) :W4-W10
[15]   Influence of field of view size on image quality: ultra-high-resolution CT vs. conventional high-resolution CT [J].
Miyata, Tomo ;
Yanagawa, Masahiro ;
Hata, Akinori ;
Honda, Osamu ;
Yoshida, Yuriko ;
Kikuchi, Noriko ;
Tsubamoto, Mitsuko ;
Tsukagoshi, Shinsuke ;
Uranishi, Ayumi ;
Tomiyama, Noriyuki .
EUROPEAN RADIOLOGY, 2020, 30 (06) :3324-3333
[16]   Effect of Tube Voltage on Image Quality in 64-Section Multidetector 3D CT Angiography: Evaluation with a Vascular Phantom with Superimposed Bone Skull Structures [J].
Murakami, Y. ;
Kakeda, S. ;
Kamada, K. ;
Ohnari, N. ;
Nishimura, J. ;
Ogawa, M. ;
Otsubo, K. ;
Morishita, Y. ;
Korogi, Y. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (04) :620-625
[17]   FACTORS AFFECTING X-RAY-SPECTRA [J].
NICKOLOFF, EL ;
BERMAN, HL .
RADIOGRAPHICS, 1993, 13 (06) :1337-1348
[18]   Effects of different tube potentials and iodine concentrations on image enhancement, contrast-to-noise ratio and noise in micro-CT images: a phantom study [J].
Razak, Hairil Rashmizal Abdul ;
Rahmat, Said Mohd Shaffiq Said ;
Saad, Wan Mazlina Md .
QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2013, 3 (05) :256-261
[19]  
Riabroi Kittipong, 2018, Neurointervention, V13, P110, DOI 10.5469/neuroint.2018.00983
[20]   Value of Flat-detector Computed Tomography Angiography with Intravenous Contrast Media Injection in the Evaluation and Treatment of Acutely Ruptured Aneurysms of the AcomA complex: ASingle Center Experience in 15 Cases [J].
Roesch, Julie ;
Lang, Stefan ;
Goelitz, Philipp ;
Kallmuenzer, Bernd ;
Roessler, Karl ;
Doerfler, Arnd ;
Struffert, Tobias .
CLINICAL NEURORADIOLOGY, 2018, 28 (04) :545-551