Optimising image quality in intravenous cerebral cone beam computed tomography

被引:1
作者
Broadley, Lisa [1 ]
Erskine, Brendan [1 ]
Marshall, Elissa [1 ]
Ewert, Kyle [1 ]
Picker, Bjoern [1 ]
机构
[1] Alfred Hosp, Melbourne, Vic 3004, Australia
关键词
angiography; head and neck; vascular interventional; CONTRAST-MEDIUM; CT ANGIOGRAPHY; DOSE REDUCTION; TUBE VOLTAGE; FOLLOW-UP; RADIATION; INJECTION; OPTION;
D O I
10.1002/jmrs.735
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
IntroductionThe efficacy of intravenous cerebral Cone Beam Computed Tomography (IV CBCT) is well established; however, image quality has only ever been authenticated by subjective evaluation. The aim of this study was to quantify the factors pertinent to achieving consistent and optimal image quality when performing IV CBCT.MethodsBetween 1 March 2021 and 30 October 2022, 79 patients received IV CBCT. These candidates were divided into three main acquisition field size categories (22/32, 42 and 48 cm) according to the clinical indication. The images were analysed using both a quantitative assessment and a subjective evaluation. Here, a comparison of Hounsfield units (HUs), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and noise index was conducted for each study and compared relative to the acquisition field size. The subjective analysis was performed in a non-blinded fashion where the diagnostic value (DV) of the exam was determined according to a graded scale. A phantom analysis for each of the acquisition field sizes was conducted and modulation transfer function (MTF) graphed.ResultsSignificantly higher HU, SNR, CNR and lower noise indices were achieved with the 42-cm protocol than the 22/32 and 48-cm protocols. Here a greater DV was also reported. The MTF demonstrates marginally improved spatial resolution for the 22-cm protocol, but this is near equivocal for the 32-, 42 and 48-cm protocols.ConclusionThe use of larger acquisition field sizes provides improved image quality when performing IV CBCT as an alternative to intra-arterial (IA) CBCT. The efficacy of intravenous (IV) Cone Beam Computed Tomography (CBCT) is well established; however the image quality has only ever been authenticated by subjective evaluation. The aim of this study was to quantify the factors pertinent to achieving consistent and optimal image quality when performing IV CBCT.image
引用
收藏
页码:26 / 34
页数:9
相关论文
共 27 条
  • [1] [Anonymous], 2005, RAD PROTECTION SERIE, V8, P13
  • [2] Comparison of effective doses obtained from dose-area product and air kerma measurements in interventional radiology
    Bor, D
    Sancak, T
    Olgar, T
    Elcim, Y
    Adanali, A
    Sanlidilek, U
    Akyar, S
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916) : 315 - 322
  • [3] Effective Dose during Needle Interventions: Cone-beam CT Guidance Compared with Conventional CT Guidance
    Braak, Sicco J.
    van Strijen, Marco J. L.
    van Es, Hendrik W.
    Nievelstein, Rutger A. J.
    van Heesewijk, Johannes P. M.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (04) : 455 - 461
  • [4] Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know
    Brenner, DJ
    Doll, R
    Goodhead, DT
    Hall, EJ
    Land, CE
    Little, JB
    Lubin, JH
    Preston, DL
    Preston, RJ
    Puskin, JS
    Ron, E
    Sachs, RK
    Samet, JM
    Setlow, RB
    Zaider, M
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2003, 100 (24) : 13761 - 13766
  • [5] Angiographic CT with intravenous administration of contrast medium is a noninvasive option for follow-up after intracranial stenting
    Buhk, Jan-Hendrik
    Lingor, Paul
    Knauth, Michael
    [J]. NEURORADIOLOGY, 2008, 50 (04) : 349 - 354
  • [6] Future Directions of Flow Diverter Therapy
    Chiu, Albert Ho Yuen
    Phillips, Timothy John
    [J]. NEUROSURGERY, 2020, 86 : S106 - S116
  • [7] Local diagnostic reference levels for angiographic and fluoroscopic procedures: Australian practice
    Erskine, Brendan J.
    Brady, Zoe
    Marshall, Elissa M.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (01) : 75 - 82
  • [8] A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise-power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantom
    Friedman, Saul N.
    Fung, George S. K.
    Siewerdsen, Jeffrey H.
    Tsui, Benjamin M. W.
    [J]. MEDICAL PHYSICS, 2013, 40 (05)
  • [9] Optimized Angiographic CT Using Intravenous Contrast Injection: A Noninvasive Imaging Option for the Follow-Up of Coiled Aneurysms?
    Goelitz, P.
    Struffert, T.
    Kaschka, I.
    Roessler, K.
    Knossalla, F.
    Doerfler, A.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (12) : 2341 - 2347
  • [10] Angiographic CT with Intravenous Contrast Injection Compared with Conventional Rotational Angiography in the Diagnostic Work-Up of Cerebral Aneurysms
    Goelitz, P.
    Struffert, T.
    Knossalla, F.
    Saake, M.
    Ott, S.
    Ganslandt, O.
    Doerfler, A.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (05) : 982 - 987