Background: The renin-angiotensin system produces a series of biologically active angiotensin (Ang) peptides. These Ang peptides are the major regulators of blood pressure and Na homeostasis, and play a critical role in maintaining cardiovascular and fluid homeostasis. The concentration of Ang peptides in the body is at trace levels, making their detection and quantification a challenge. In this study, a rapid and sensitive analytical method using mass spectrometry coupled with ultrahigh-performance liquid chromatography (UHPLC/MS) was developed to simultaneously quantify 14 Ang peptides.Methods: UHPLC/MS was employed to quantify 14 Ang peptides in mouse and human plasma. An HSS T3 column (2.1 x 100 mm, 1.8 mu m) with an HSS T3 precolumn and triple-quadrupole mass spectrometer combined with an electrospray ionization source were utilized. Sample pretreatment involved a one-step protein precipitation using methanol. The total analysis time was within 7.5 min and the target peptides were detected in positive ion mode and quantified by selected reaction monitoring mode.Results: The method was validated for linearity, detection and quantification limits, precision, stability, recovery and matrix effect. The limits of detection of Ang II, Ang III, Ang-(1-7), Ang-(2-7), Ang-(3-7), Ang-(1-9), bradykinin, Asn(1) and Val(5)-Ang II are all less than 1 pg mL(-1), indicating high sensitivity. The intra-day and inter-day precision was within 15%, and the accuracy was between 85% and 115%. Meanwhile, the sample and reference solution were stable within 48 h, and the recovery and matrix effect met the quantitative requirements.Conclusions: The method is currently reported to allow the largest number of Ang peptide species to be detected at one time. In addition, the proposed method offers a fast and reliable approach for comprehensive analysis of Ang metabolism in biological samples, facilitating research on the physiological and pathological states of cardiovascular, kidney and respiratory diseases.