Impact of Phantom Size on Low-Energy Virtual Monoenergetic Images of Three Dual-Energy CT Platforms

被引:5
|
作者
Greffier, Joel [1 ]
Van Ngoc Ty, Claire [2 ]
Fitton, Isabelle [2 ]
Frandon, Julien [1 ]
Beregi, Jean-Paul [1 ]
Dabli, Djamel [1 ]
机构
[1] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, IMAGINE UR UM 103, F-30029 Nimes, France
[2] Univ Paris, Hop Europeen Georges Pompidou, AP HP, Dept Radiol, F-75015 Paris, France
关键词
dual-energy; multidetector computed tomography; task-based image quality assessment; split-filter; RECONSTRUCTION;
D O I
10.3390/diagnostics13193039
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to compare the quality of low-energy virtual monoenergetic images (VMIs) obtained with three Dual-Energy CT (DECT) platforms according to the phantom diameter. Three sections of the Mercury Phantom 4.0 were scanned on two generations of split-filter CTs (SFCT-1st and SFCT-2nd) and on one Dual-source CT (DSCT). The noise power spectrum (NPS), task-based transfer function (TTF), and detectability index (d') were assessed on VMIs from 40 to 70 keV. The highest noise magnitude values were found with SFCT-1st and noise magnitude was higher with DSCT than with SFCT-2nd for 26 cm (10.2% +/- 1.3%) and 31 cm (7.0% +/- 2.5%), and the opposite for 36 cm (-4.2% +/- 2.5%). The highest average NPS spatial frequencies and TTF values at 50% (f50) values were found with DSCT. For all energy levels, the f50 values were higher with SFCT-2nd than SFCT-1st for 26 cm (3.2% +/- 0.4%) and the opposite for 31 cm (-6.9% +/- 0.5%) and 36 cm (-5.6% +/- 0.7%). The lowest d' values were found with SFCT-1st. For all energy levels, the d' values were lower with DSCT than with SFCT-2nd for 26 cm (-6.2% +/- 0.7%), similar for 31 cm (-0.3% +/- 1.9%) and higher for 36 cm (5.4% +/- 2.7%). In conclusion, compared to SFCT-1st, SFCT-2nd exhibited a lower noise magnitude and higher detectability values. Compared with DSCT, SFCT-2nd had a lower noise magnitude and higher detectability for the 26 cm, but the opposite was true for the 36 cm.
引用
收藏
页数:13
相关论文
共 50 条
  • [41] Detection of fatty liver using virtual non-contrast dual-energy CT
    Pengcheng Peter Zhang
    Hailey H. Choi
    Michael A. Ohliger
    Abdominal Radiology, 2022, 47 : 2046 - 2056
  • [42] Detection of fatty liver using virtual non-contrast dual-energy CT
    Zhang, Pengcheng Peter
    Choi, Hailey H.
    Ohliger, Michael A.
    ABDOMINAL RADIOLOGY, 2022, 47 (06) : 2046 - 2056
  • [43] Monoenergetic Imaging of Dual-energy CT Reduces Artifacts from Implanted Metal Orthopedic Devices in Patients with Factures
    Zhou, Changsheng
    Zhao, Yan E.
    Luo, Song
    Shi, Hongyuan
    Ii, Lin
    Zheng, Ling
    Zhang, Long Jiang
    Lu, Guangming
    ACADEMIC RADIOLOGY, 2011, 18 (10) : 1252 - 1257
  • [44] The synthesis of high-energy CT images from low-energy CT images using an improved cycle generative adversarial network
    Zhou, Haojie
    Liu, Xinfeng
    Wang, Haiyan
    Chen, Qihang
    Wang, Rongpin
    Pang, Zhi-Feng
    Zhang, Yong
    Hu, Zhanli
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (01) : 28 - 42
  • [45] Quantification of spinal bone marrow fat fraction using three-material decomposition technique on dual-energy CT: A phantom study
    Li, Baojun
    Hua, Ning
    Li, Janelle
    Andreu-Arasa, V. Carlota
    LeBedis, Christina
    Anderson, Stephan W.
    MEDICAL PHYSICS, 2023, 50 (07) : 4105 - 4111
  • [46] Accuracy of iodine removal using dual-energy CT with or without a tin filter: an experimental phantom study
    Kawai, Tatsuya
    Takeuchi, Mitsuru
    Hara, Masaki
    Ohashi, Kazuya
    Suzuki, Hirochika
    Yamada, Kiyotaka
    Sugimura, Yuya
    Shibamoto, Yuta
    ACTA RADIOLOGICA, 2013, 54 (08) : 954 - 960
  • [47] Robustness of radiomics among photon-counting detector CT and dual-energy CT systems: a texture phantom study
    Zhu, Lan
    Dong, Haipeng
    Sun, Jing
    Wang, Lingyun
    Xing, Yue
    Hu, Yangfan
    Lu, Junjie
    Yang, Jiarui
    Chu, Jingshen
    Yan, Chao
    Yuan, Fei
    Zhong, Jingyu
    EUROPEAN RADIOLOGY, 2025, 35 (02) : 871 - 884
  • [48] Determination of optimal virtual monochromatic energy level for target delineation of brain metastases in radiosurgery using dual-energy CT
    Karino, Tsukasa
    Ohira, Shingo
    Kanayama, Naoyuki
    Wada, Kentaro
    Ikawa, Toshiki
    Nitta, Yuya
    Washio, Hayate
    Miyazaki, Masayoshi
    Teshima, Teruki
    BRITISH JOURNAL OF RADIOLOGY, 2020, 93 (1106)
  • [49] Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT
    Nakhostin, Dominik
    Sartoretti, Thomas
    Eberhard, Matthias
    Krauss, Bernhard
    Mueller, Daniel
    Alkadhi, Hatem
    Euler, Andre
    ABDOMINAL RADIOLOGY, 2021, 46 (05) : 2079 - 2089
  • [50] Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT
    Dominik Nakhostin
    Thomas Sartoretti
    Matthias Eberhard
    Bernhard Krauss
    Daniel Müller
    Hatem Alkadhi
    André Euler
    Abdominal Radiology, 2021, 46 : 2079 - 2089