Silent active device tracking for MR-guided interventional procedures

被引:1
|
作者
Stehning, Christian [1 ,5 ]
Krueger, Sascha [2 ]
Weiss, Steffen [2 ]
Smink, Jouke [3 ]
Koken, Peter [2 ]
Hindricks, Gerhard [4 ]
Jahnke, Cosima [4 ]
Paetsch, Ingo [4 ]
机构
[1] Philips Clin Sci, Hamburg, Germany
[2] Philips Res Labs, Hamburg, Germany
[3] Philips Healthcare, Best, Netherlands
[4] HELIOS Heart Ctr Leipzig, Dept Electrophysiol, Leipzig, Germany
[5] Philips Clin Sci, Roentgenstr 24-26, D-22335 Hamburg, Germany
关键词
catheter tracking; interventional MRI; noise reduction; positioning; CATHETER TRACKING; ACOUSTIC NOISE; ABLATION;
D O I
10.1002/mrm.29567
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo evaluate a silent MR active catheter tracking sequence that allows conducting catheter interventions with low acoustic noise levels. MethodsTo reduce the acoustic noise associated with MR catheter tracking, we implemented a technique previously used in conventional MRI. The gradient waveforms are modified to reduce the sound pressure level (SPL) and avoid acoustic resonances of the MRI system. The efficacy of the noise reduction was assessed by software-predicted SPL and verified by measurements. Furthermore, the quality of the catheter tracking signal was assessed in a phantom experiment and during interventional cardiovascular MRI sessions targeted at isthmus-related flutter ablation. ResultsThe maximum measured SPL in the scanner room was 104 dB(A) for real-time imaging, and 88 dB(A) and 69 dB(A) for conventional and silent tracking, respectively. The SPL measured at different positions in the MR suite using silent tracking were 65-69 dB(A), and thus within the range of a normal conversation. Equivalent signal quality and tracking accuracy were obtained using the silent variant of the catheter tracking sequence. ConclusionOur results indicate that silent MR catheter tracking capabilities are identical to conventional catheter tracking. The achieved acoustic noise reduction comes at no penalty in terms of tracking quality or temporal resolution, improves comfort and safety, and can overcome the need for MR-compatible communication equipment and background noise suppression during the actual interventional procedure.
引用
收藏
页码:2005 / 2013
页数:9
相关论文
共 50 条
  • [1] MR-guided interventional procedures:: A review
    Sequeiros, RB
    Ojala, R
    Kariniemi, J
    Perälä, J
    Niinimäki, J
    Reinikainen, H
    Tervonen, O
    ACTA RADIOLOGICA, 2005, 46 (06) : 576 - 586
  • [2] MR-guided device tracking with projection reconstruction
    Flask, C
    Lee, K
    Wong, EY
    Zhang, Q
    Shankaranarayanan, A
    Wendt, M
    RADIOLOGY, 2000, 217 : 348 - 348
  • [3] MR-guided interventional breast procedures considering vacuum biopsy in particular
    Viehweg, P
    Heinig, A
    Amaya, B
    Alberich, T
    Laniado, M
    Heywang-Köbrunner, SH
    EUROPEAN JOURNAL OF RADIOLOGY, 2002, 42 (01) : 32 - 39
  • [5] Fast and accurate automatic registration for MR-guided procedures using active microcoils
    Krueger, Sascha
    Wolff, Stephan
    Schmitgen, Arno
    Timinger, Holger
    Bublat, Martin
    Schaeffter, Tobias
    Nabavi, Arya
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2007, 26 (03) : 385 - 392
  • [6] Interventional MR-guided excisional biopsy of breast lesions
    Gould, SWT
    Lamb, G
    Lomax, D
    Gedroyc, W
    Darzi, A
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01): : 26 - 30
  • [7] MR-Guided intravascular procedures:: Real-time parameter control and automated slice positioning with active tracking coils
    Bock, M
    Volz, S
    Zühlsdorff, S
    Umathum, R
    Fink, C
    Hallscheidt, P
    Semmler, W
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 19 (05) : 580 - 589
  • [8] A novel MR-guided interventional device for 3D circumferential access to breast tissue
    Smith, Matthew
    Zhai, Xu
    Harter, Ray
    Sisney, Gale
    Elezaby, Mai
    Fain, Sean
    MEDICAL PHYSICS, 2008, 35 (08) : 3779 - 3786
  • [9] Real-Time Active MR-Tracking of Metallic Stylets in MR-Guided Radiation Therapy
    Wang, Wei
    Dumoulin, Charles L.
    Viswanathan, Akila N.
    Tse, Zion T. H.
    Mehrtash, Alireza
    Loew, Wolfgang
    Norton, Isaiah
    Tokuda, Junichi
    Seethamraju, Ravi T.
    Kapur, Tina
    Damato, Antonio L.
    Cormack, Robert A.
    Schmidt, Ehud J.
    MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (05) : 1803 - 1811
  • [10] MR-guided percutaneous interventional surgery robot system: A review
    Zhang, Weixi
    Yu, Jiaxing
    Lu, Xikun
    Xu, Yanjie
    INTERNATIONAL JOURNAL OF ADVANCED ROBOTIC SYSTEMS, 2025, 22 (01):