Is a portable pressure plate an alternative to force plates for measuring postural stability and interlimb coordination of quiet standing balance control?

被引:2
|
作者
Schroder, Jonas [1 ,2 ,10 ]
Hallemans, Ann [1 ]
Saeys, Wim [1 ,3 ]
Yperzeele, Laetitia [4 ,5 ,6 ]
Kwakkel, Gert [7 ,8 ,9 ]
Truijen, Steven [1 ]
机构
[1] Univ Antwerp, Dept Rehabil Sci & Physiotherapy REVAKI, Res Grp MOVANT, Antwerp, Belgium
[2] Hasselt Univ, Fac Rehabil Sci, REVAL Rehabil Res Ctr, Hasselt, Belgium
[3] RevArte Rehabil Hosp, Dept Neurorehabil, Edegem, Belgium
[4] Antwerp Univ Hosp, Neurovasc Ctr Antwerp, Dept Neurol, Edegem, Belgium
[5] Antwerp Univ Hosp, Dept Neurol, Stroke Unit, Edegem, Belgium
[6] Univ Antwerp, Res Grp Translat Neurosci, Antwerp, Belgium
[7] Vrije Univ Amsterdam, Dept Rehabil Med, Amsterdam Movement Sci, Amsterdam Neurosci,Amsterdam UMC, Amsterdam, Netherlands
[8] Northwestern Univ, Dept Phys Therapy & Human Movement Sci, Chicago, IL USA
[9] Amsterdam Rehabil Res Ctr Reade, Amsterdam, Netherlands
[10] Hasselt Univ, Fac Rehabil Sci, REVAL Rehabil Res Ctr, Campus Diepenbeek,Wetenschapspk 7, B-3590 Hasselt, Belgium
关键词
Reliability and validity; postural balance; posturography; biomechanics; force plate; center of pressure; RELIABILITY; REHABILITATION; TRAJECTORIES; PARAMETERS; RECOVERY; SCALE;
D O I
10.1177/20556683241234858
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction: Center-of-pressure (COP) synchronization and symmetry can inform adaptations in balance control following one-sided sensorimotor impairments (e.g., stroke). As established force plates are impossible to transport, we aimed to criterion validate a portable pressure plate for obtaining reliable COP synchronization and symmetry measures, next to conventional postural stability measures. Methods: Twenty healthy adults participated. In a single session, three 40-s eyes-open and eyes-closed quiet stance trials were performed per plate-type, randomly ordered. Individual-limb COPs were measured to calculate between-limb synchronization (BLS) and dynamic control asymmetry (DCA). Net COP (i.e., limbs combined) area, amplitude, and velocity were used to describe anteroposterior (AP) and mediolateral (ML) postural stability. Criterion validity was evaluated using Spearman correlations (r) and Bland-Altman plots. Test-retest reliability was tested using intraclass correlation coefficients (ICC). Results: Strong correlations (r > 0.75) and acceptable reliability (ICC > 0.80) were found regarding individual-limb COP velocity and DCA, net COP ML amplitude and AP and ML velocities. Bland-Altman plots yielded possible proportional bias; the pressure plate systematically underestimated COP scores by force plates and a larger error associated with a larger measurement. Conclusions: Despite correlations between instruments and sufficient reliability for measuring postural stability and DCA, this technical note strongly suggests, due to a systematic deviation, using the same plate-type to accurately assess performance change within subjects longitudinally over time.
引用
收藏
页数:10
相关论文
共 2 条
  • [1] Force Plate Assessment of Quiet Standing Balance Control: Perspectives on Clinical Application Within Stroke Rehabilitation
    Mansfield, Avril
    Inness, Elizabeth L.
    REHABILITATION PROCESS AND OUTCOME, 2015, 4 : 7 - 15
  • [2] Evaluation of sitting and standing postural balance in cerebral palsy by center-of-pressure measurement using force plates: comparison with clinical measurements
    Shim, Dain
    Park, Dongho
    Yoo, Beomki
    Choi, Joong-on
    Hong, Juntaek
    Choi, Tae Young
    Park, Eun Sook
    Rha, Dong-wook
    GAIT & POSTURE, 2022, 92 : 110 - 115